Published ahead of print on June 12, 2008, doi:10.1164/rccm.200801-062OC
© 2008 American Thoracic Society doi: 10.1164/rccm.200801-062OC
Clinically Important Factors Influencing the Diagnostic Measurement of Pleural Fluid pH and Glucose1 Oxford Centre for Respiratory Medicine and University of Oxford, Oxford, United Kingdom; 2 Norfolk and Norwich University Hospital, Norwich, United Kingdom; and 3 Centre for Respiratory Research, University College London, London, United Kingdom Correspondence and requests for reprints should be addressed to Dr. Y. C. Gary Lee, M.B.Ch.B., Ph.D., F.R.A.C.P., Consultant Chest Physician and Senior Lecturer, Oxford Centre for Respiratory Medicine, Churchill Hospital, Oxford, OX3 7LJ, UK. E-mail: ycgarylee{at}hotmail.com
Rationale: Accurate pleural fluid pH and glucose measurement is a key component in the diagnosis and management of patients with pleural effusion. Standardized methods of pleural fluid collection have not been defined. Objectives: To assess the effect of common clinical factors that may distort measurement accuracy of pleural fluid pH and glucose. Methods: Ninety-two exudative pleural aspirates were collected in commercially available blood gas syringes. Measurements and Main Results: Samples were analyzed immediately using a blood gas analyzer. The effects of residual air, lidocaine, heparin, and delay in analysis (24 h) on pH and glucose measurement accuracy were assessed. Pleural fluid pH was significantly increased by residual air (mean ± SD, 0.08 ± 0.07; 95% confidence interval [CI], 0.06 to 0.09; P < 0.001) and significantly decreased by residual lidocaine (0.2 ml; mean change in pH, –0.15 ± 0.09; 95% CI, –0.13 to –0.18; P < 0.001) and residual heparin (mean change in pH, –0.02 ± 0.05; 95% CI, –0.01 to –0.04; P = 0.027). Pleural fluid pH was stable at room temperature for 1 hour and significantly increased at 4 (mean ± SD, 0.03 ± 0.07; 95% CI, 0.01 to 0.04; P = 0.003) and 24 hours (0.05 ± 0.12; 95% CI, 0.03 to 0.08; P < 0.001). Pleural fluid glucose concentration was not clinically significantly altered by residual air, lidocaine (up to 0.4 ml), or 24-hour analysis delay. Conclusions: Accuracy of measured pleural pH is critically dependent on sample collection method. Residual air, lidocaine, and analysis delay significantly alter pH and may impact on clinical management. Pleural fluid glucose concentration is not significantly influenced by these factors. Protocols defining appropriate sampling and analysis methods are needed.
Key Words: pleural fluid pH glucose measurement
Approximately 300,000 pleural effusions require diagnostic assessment each year in the United States and United Kingdom (1, 2). Near patient testing of pleural fluid pH is an important component of this assessment for the diagnosis of pleural infection (including tuberculosis), rheumatoid pleural effusion, and esophageal rupture (3), and the prognostic assessment of malignant effusion (1, 4–7). Measurement of pleural fluid pH is advocated in the literature and in management guidelines from professional societies (including the American and British Thoracic Societies [3, 8–11]) as part of first-line pleural fluid analysis. A low pleural fluid pH implies a high metabolic activity in the pleural space. A reduced pH (typically <7.20 [12]) has been shown to be a robust marker for instituting intercostal tube drainage in patients with suspected pleural infection (13, 14), and is used as an entry criterion for clinical studies of pleural infection (15, 16). In malignant pleural effusion, low pleural fluid pH correlates with a poor prognosis (7) and predicts pleurodesis failure (6). To maximize diagnostic accuracy in individual patients requires that any measurement inaccuracies that cancel out in larger samples be minimized. Previous studies advocate pH measurement using a blood gas analyzer (17), within 1–2 hours (18, 19) of collection, but beyond this there have been no systematic studies to define the sampling algorithm that best reduces variation during measurement, and there is likely to be substantial variation in collection methods (20). There have been no comparisons of the effects of such sampling strategies on pleural fluid glucose, which closely reflects pleural fluid pH (21) and which is advocated as a suitable substitute for pH measurement (12). We hypothesized that pleural fluid pH and/or glucose may be significantly altered by commonly observed variations in collecting practice. Admixture of fluid with air, residual lidocaine or heparin (which are of acidic pH), or analysis delay could produce clinically important changes in measured pH, with implications for patient care. Equally, omission of heparin could allow fibrin clot formation, risking blood gas analyzer damage, or distorted analysis results. This study was designed to describe the effects of these factors.
Inclusion Criteria Consecutive patients presenting to a tertiary referral respiratory center with a pleural effusion requiring thoracentesis, chest drain insertion, or thoracoscopy were included between October 2006 and September 2007. To increase representation of samples from patients with suspected pleural infection (in whom the pleural fluid pH is particularly important), patients were specifically recruited with suspected parapneumonic pleural effusions for a 2-month period (February to April 2008, 12 further cases added).
Exclusion Criteria
Pleural Fluid Collection
Fluid was immediately introduced into pre-prepared syringes (Portex Arterial Blood Sampling System, CE 98/79/EC; Smiths Medical International, Watford, UK), which have a maximum volume of 3 ml and contain preloaded heparin sodium (
Pleural fluid samples were analyzed within 10 minutes of collection by a blood gas machine (ABL 700 series; Radiometer, Copenhagen, Denmark). During each analysis, 195 µl of fluid were withdrawn for measurements of pleural fluid pH, partial pressure of O2 and CO2, and glucose concentration. The machine conducted an autocalibration protocol on a 4-hourly basis and in addition was manually calibrated by a technician three times a day according to the manufacturer's instructions.
Stability of pleural fluid pH.
Pleural fluid glucose analysis. All pleural fluid samples were sent to the laboratory for lactate dehydrogenase (LDH) and total protein quantification. Additional tests (e.g., cytology, microbiology) were performed as clinically indicated. To assess the contribution of cellular contents in pleural fluid to any observed changes in pH (e.g., release of cell metabolites), supernatant of pleural fluid was obtained by centrifugation of the fluid at 3,000 rpm for 10 minutes (5702R Centrifuge; Eppendorf, Cambridge, UK), or by passing the pleural fluid through a 0.2-µM filter—both methods serving to remove cellular material from the sample. The supernatants and paired untreated pleural fluid samples (i.e., not centrifuged or filtered) were placed in identical containers with all air expelled, and their pH at Time 0 (just after centrifugation period) and at 24 hours was measured as described above in a subset of samples.
Statistical Methods
A total of 92 pleural fluid samples from 81 patients were obtained. All samples were used to assess pleural fluid pH and glucose stability over time (see below). In a subset of 52 samples, the effects of exposure to air, residual lidocaine or heparin on pH, and glucose analyses were assessed. Baseline characteristics of the 92 samples are presented in Table 1. Mean pleural fluid pH at baseline was 7.27 ± 0.17.
Presence of Air on pH The presence of air significantly increased pleural fluid pH by 0.08 ± 0.07 (95% confidence interval [CI], 0.06 to 0.09; P < 0.001) (Figure 1). In 71% of samples, presence of air in the collection syringe resulted in a clinically significant change in pleural fluid pH of greater than 0.05 (Table 2). The change in pH was not related to the baseline pH (data not shown). Change in CO2 and change in pH were correlated in each air-exposed sample, giving a Pearson's correlation coefficient of –0.42 (P = 0.001).
Presence of Residual Lidocaine on pH The pH of lidocaine used was 5.40 ± 0.05 (average of three independent measurements). A significant reduction in pleural fluid pH was observed in the presence of even minimal volumes (0.2 ml) of residual lidocaine (mean change of pH, –0.15 ± 0.09; 95% CI, –0.13 to –0.18; P < 0.001). A dose-dependent reduction of pH with lidocaine was observed with 0.4 ml lidocaine (mean change of pH, –0.24 ± 0.11; 95% CI, –0.21 to –0.27; P < 0.001) and with 1.0 ml lidocaine (mean change of pH, –0.43 ± 0.14; 95% CI, –0.39 to –0.47; P < 0.001) (Figure 2).
The presence of residual lidocaine induced a clinical significant reduction in pH in nearly all samples: 94, 96, and 100% of samples in the presence of 0.2, 0.4, and 1.0 ml of lidocaine, respectively (Table 2). The change in pH in cases of suspected pleural infection were not different from those with other diagnoses (Table 3).
Presence of Heparin on pH Retaining preloaded heparin within the blood gas syringe resulted in a small but statistically significant decrease in pH (mean change of pH, –0.02 ± 0.05; CI, –0.01 to –0.04; P = 0.027) (Figure 3). The pH of the heparin in the blood gas syringes was 6.49 ± 0.16 (based on three independent measurements).
Effect of Delayed Analysis on pH Pleural fluid pH measurements changed significantly from baseline over time, and the direction of change (increase or decrease) was variable among samples, although an overall rise in pH over time was observed. Statistically, no significant mean difference was observed in pleural fluid pH measured at baseline and at 1 hour (mean change in pH from baseline, 0.01 ± 0.05; 95% CI, –0.00 to 0.02; P = 0.07). The changes in pH became significant at 4 and 24 hours (4 h: mean change in pH, 0.03 ± 0.07; 95% CI, 0.01 to 0.04; P = 0.003; 24 h: 0.05 ± 0.12; 95% CI, 0.03 to 0.08; P < 0.001) (Figure 4). At 1 hour, 13% of samples were altered beyond the clinically significant limit (>0.05), compared with 26% at 4 hours and 68% at 24 hours (Table 2). Change in pH was plotted against baseline pH, and no significant correlation was demonstrated.
The PO2 level in samples showed a consistent increase in samples over time (PO2 rise after 1 h: 3.20 ± 2.75 kPa [24.0 ± 20.6 mm Hg]; 4 h: 3.98 ± 3.76 kPa [29.9 ± 28.2 mm Hg]; 24 h: 5.25 ± 5.55 kPa [39.4 ± 41.6 mm Hg]).
pH and PCO2 To assess the contribution of cellular content (e.g., via metabolism or release of metabolic products) to changes in pleural fluid pH, pleural fluid samples were centrifuged or filtered to remove cells, and were compared with a paired untreated sample at both 0 and 24 hours. The sample centrifugation or filtration process resulted in a significant change in pleural fluid pH compared with untreated samples (mean difference for both = 0.06), and precluded further analysis.
Glucose Measurement by Blood Gas Analyzer Some of the minor variation observed between the two measurements is likely to be due to rounding up of results below 1.0 mmol/L to 1.0 in the lab sample, which did not occur with the blood gas machine.
Presence of Air, Lidocaine, and Heparin on Pleural Fluid Glucose Concentration
The addition of small volumes of lidocaine (0.2 ml) caused a statistically but not clinically significant decrease in the pleural fluid glucose concentration (mean difference, –0.41 mmol/L [10.8 mg/dl]; SD, 0.60 mmol/L; 95% CI, –0.24 to –0.60; P < 0.001) (Figure 5B). More pronounced reductions in glucose concentration were seen with larger volumes of lidocaine, and were likely due to dilution effect (0.4 ml lidocaine: mean difference, –0.51 mmol/L [9.18 mg/dl]; SD, 0.49 mmol/L; 95% CI, –0.33 to –0.68; P < 0.001; 1.0 ml lidocaine: mean difference, –1.19 mmol/L [21.4 mg/dl]; SD, 0.99 mmol/L; 95% CI, –0.83 to –1.56; P < 0.001). The addition of heparin caused a statistically but not clinically significant decrease in pleural fluid glucose concentration (mean difference, –0.66 mmol/L [11.88 mg/dl]; SD, 0.52 mmol/L; 95% CI, –0.47 to –0.85; P < 0.001) (Figure 5C). Comparable reductions in glucose concentrations with similar volumes of heparin and lidocaine were observed, suggesting that the observed reduction in glucose concentration was the result of dilution.
Effect of Delayed Analysis on Glucose Measurements Measurements of pleural fluid glucose stored in oxalate tubes for 24 hours were highly consistent with those analyzed immediately (mean difference, –0.03 mmol/L [0.54 mg/dl]; SD, 0.16 mmol/L; 95% CI, –0.08 to 0.03; P = 0.33).
This study has demonstrated that the accuracy of measured pleural pH is critically dependent on the method of sample collection. There is currently no standardized protocol for pleural fluid collection for pH/glucose measurement and it is therefore likely that the factors studied here are common influences on measurement in day-to-day clinical practice. These data demonstrate that presence of air or residual lidocaine in the collection syringe results in clinically significant changes in pH. Heparin produces a statistically detectable pH change that is of little clinical importance. In contrast, pleural fluid glucose measurements, analyzed by either blood gas machine or conventional laboratory assays, are less susceptible to variations in collection practice or delay in analysis. These data are the first to show that even small amounts of air in the pleural fluid collection syringe result in a significant increase in pleural fluid pH. This finding applies to all samples, regardless of the baseline pH, and shows that air in pleural fluid is capable of artificially elevating the pH sufficiently to change clinical management—for example, by deferring the drainage of an acidic parapneumonic effusion. The mechanism of the increase in pH with air has not been explored. We hypothesize that the gradient between the partial pressures of CO2 in pleural fluid (mean PCO2: 7.12 kPa, 53.4 mm Hg, in our samples) and atmospheric air results in rapid CO2 diffusion through the fluid–air interface, raising the pH. Lidocaine, administered as a local anesthetic before a pleural procedure, appears clinically important in its potential effect on pH measurement. Even the presence of a minute amount of residual lidocaine (0.2 ml), compatible with the dead space of fine-bore needles, causes a clinically significant drop in pH in 94% of samples, and this effect becomes more marked in a dose-dependent manner. This helps explain previous data showing reduction in pleural fluid pH in samples collected after infiltration of intercostal tissue with a local anesthetic (24). Heparin in the sample produces a statistically significant, but clinically insignificant change in pH. This finding is in contrast with one previous smaller study (25). Fibrinous clots often develop in pleural fluid samples and addition of a small volume of anticoagulant (e.g., heparin) to thoracentesis specimens, to avoid clotting before laboratory analysis, has been advocated, especially for leukocyte differential counts (18). Commercially available blood gas syringes, which are commonly used for pleural fluid analysis, are often preheparinized. The presented results confirm that heparin is acidic (pH 6.49) and the preloaded heparin in commercially prepared blood gas syringes produces a statistically significant reduction in pH. Reassuringly, this change was not clinically significant. Delay after sample collection does not seem to substantially alter measured pH over 1 hour, although delays of 4 hours or longer produce clinically significant alterations. A delay in analysis of 24 hours produced clinically significant inaccuracies in pH readings in over two-thirds of samples. A close correlation was observed between increase in pH and decrease in pleural fluid PCO2 over 24 hours. This may suggest that the mechanism of pH change is related to CO2 diffusing out of pleural fluid with time. The PO2 level consistently increases in these samples over 24 hours, supporting the suggestion that constant gas exchange occurs between atmospheric air and pleural fluid, even in capped containers. Changes in pH over time may also be a result of metabolism of cells or bacteria (in infected samples). The pH and glucose levels in pleural fluids are strongly correlated (21). The presented data show that pleural fluid glucose measurements are less susceptible to inaccuracy induced by the presence of air, lidocaine, or heparin, except where the volume of additive is sufficient to cause a dilutional effect. This study highlights the need for a standardized protocol for pleural pH sampling and analysis, both for clinical practice and for future research studies. Most existing literature on pleural pH did not specify details of its measurement, which may explain some of the conflicting data in this field. Clinicians must realize that, although existing literature/guidelines often use definitive thresholds of pH to guide management (e.g., a cutoff pH of 7.20 determines need for chest tube drainage of parapneumonic effusions), measurements of pH are vulnerable to variations in processing protocols. This adds to recent observations that pH may vary in different locules within a septated pleural effusion (26). Decisions on clinical management should not be governed by pleural pH alone, and must be made in conjunction with information from other investigations. There are limitations to this study. No data exist on what is the "gold standard" of pleural pH measurement, and we compared samples with an assumed "standard" collection (samples analyzed immediately with blood gas machine in the absence of air and other additives). In conclusion, pleural fluid pH measurements are subject to substantial variability during collection, which could be minimized by the systematic application of a simple collection protocol. Pleural fluid should be collected avoiding inclusion of air or other additives with the sample—particularly lidocaine—and analyzed in less than 1 hour using a blood gas analyzer. Pleural fluid glucose is an alternative to pH and its measurements are less vulnerable to changes in collection method.
Supported by a Medical Research Council Training Fellowship (N.M.R.), the British Lung Foundation (H.E.D.), and the Medical Research Council (Y.C.G.L.). Originally Published in Press as DOI: 10.1164/rccm.200801-062OC on June 12, 2008 Conflict of Interest Statement: None of the authors has a financial relationship with a commercial entity that has an interest in the subject of this manuscript. Received in original form January 10, 2008; accepted in final form June 10, 2008