ABSTRACT |
|---|
|
|
|---|
After tracheal extubation, upper and total airway resistances may
frequently be increased resulting in an increase in inspiratory effort to breathe. We tested whether breathing a helium-oxygen mixture (HeO2) would reduce inspiratory effort in the period after extubation. Eighteen consecutive patients with no chronic obstructive pulmonary disease who had received mechanical ventilation (> 48 h) were successively studied immediately after extubation (N2O2), 15 min after breathing HeO2, and after return to
N2O2. Effort to breathe, assessed by the transdiaphragmatic pressure swings (
Pdi) and the pressure-time index of the diaphragm
(PTI), comfort, and gas exchange, were the main end points. The
mean reduction of the transdiaphragmatic pressure under HeO2
was 19 ± 5%. All but three patients presented a decrease in transdiaphragmatic pressure under HeO2, ranging from
4 to
55%,
and a significant reduction in
Pdi was observed between HeO2
and N2O2 (10.2 ± 0.7 versus 8.6 ± 1.1 versus 10.0 ± 0.8 cm H2O
for the three consecutive periods; p < 0.05). PTI also differed significantly between HeO2 and N2O2 (197 ± 19 versus 166 ± 22 versus 201 ± 23 cm H2O/s/min for the three periods; p < 0.05).
Breathing HeO2 significantly improved comfort, whereas gas exchange was not modified. We conclude that the use of HeO2 in the
immediate postextubation period decreases inspiratory effort and
improves comfort.
Keywords: Helium-oxygen; exutbation; weaning; work of breathing
| |
INTRODUCTION |
|---|
|
|
|---|
After endotracheal extubation, upper and total airway resistances may frequently be increased, resulting in high inspiratory effort to breathe. A few patients, ranging from 5 to 16%, develop postextubation airway obstruction and frank respiratory distress (1). In addition, a substantial number of patients develop inspiratory distress after extubation, leading to reintubation (4, 5). In these patients an increase in upper airway and total inspiratory resistance may also participate to this respiratory distress.
Helium-oxygen (HeO2) mixture has a low density and a high kinematic viscosity, allowing for a reduction in airway resistance. Some studies showed that it could have beneficial effects in the treatment of upper airway obstruction (6). Previous studies have demonstrated that breathing a helium-oxygen mixture reduces dyspnea and improves gas exchange in nonintubated patients with severe exacerbation of chronic obstructive lung disease (7) or asthma (8). Recently, helium-oxygen breathing was shown to improve pH, PaCO2 and inspiratory effort in patients with acute exacerbation of chronic obstructive pulmonary disease (COPD), and these beneficial effects were enhanced by combining noninvasive pressure support ventilation with helium-oxygen (9). The use of helium-oxygen in the postextubation period where the upper and total airway resistances may be increased, could also decrease patient's effort to breathe in this situation.
The aim of this study was to evaluate the short-term effects of HeO2 inhalation on gas exchange, comfort, and respiratory effort in the postextubation period. Inspiratory effort and gas exchange were also evaluated before extubation when patients were spontaneously breathing through their endotracheal tube. The effect of helium-oxygen (HeO2) was compared with the measurements obtained with the usual air-oxygen mixture (N2O2) Because we previously assessed the effect of HeO2 in patients with COPD (9), these patients were not included in the present study.
| |
METHODS |
|---|
|
|
|---|
The institutional review board for human subjects of Henri Mondor Hospital (Comité Consultatif de Protection des Personnes dans la Recherche Biomédicale) approved the experimental protocol. Written informed consent was obtained from each patient.
Patients
Eighteen consecutive patients who had been intubated for at least 48 h in the unit, were studied immediately after endotracheal extubation. Patients with COPD were not included. The study was performed over 5 mo in 1998.
Protocol
Patients were fully conscious and sitting in their bed at 60 degrees from horizontal. Their position was kept constant throughout the study. The esophageal-gastric catheter for measurements of effort was inserted before the beginning of the spontaneous breathing trial. During the spontaneous breathing trial, the endotracheal tube was connected to a T-piece providing humidification and oxygen. The first measurements were obtained at the end of the spontaneous breathing trial.
Two gas mixtures were tested after extubation: N2O2 and HeO2. The same FIO2 (25 to 40%) was adjusted in each patient to maintain SaO2 > 92% and was kept constant with the two gas mixtures. N2O2 was studied twice: N2O2-1 was the period immediately after extubation; the following period used HeO2, and a new period of N2O2-2 ensued. Each period lasted for 15 min.
To make sure that the only alteration was due to the gas mixture, the three conditions after endotracheal extubation were delivered with the same experimental apparatus, and the two studied gas mixtures (N2O2 and HeO2) were delivered in a similar fashion in terms of patient's equipment. The patients were not informed about the nature of the inhaled gas mixture. However, some patients were able to detect HeO2 if they happened to speak ("Donald Duck" voice).
Experimental Apparatus
Medical gas administration. The medical gases were administered through a nonrebreathing face mask, in order to reduce the amount of external air entrained on inspiration, which could result in diluting helium concentration. The HeO2 mixture composed of 22% oxygen and 78% helium was provided by a high-pressure source (cylinder of 4 bar-compressed gas; Air Liquide Medical, Bonneuil S/Marne, France). The gas was depressurized through a specific pressure regulator (HBS 315-8 No. 20 305) which was connected by plastic tubing to a tight-fitting face mask (HeO2 flow between 6 and 12 L/min). A second part of T connector was connected to the wall oxygen supply, permitting O2 enrichment of the gas mixture and the desired FIO2 as described below (oxygen flow between 1 and 5 L/min). To ensure that the reservoir bag of the rebreathing mask had similar degree of filling when the patient was breathing N2O2, and to maintain the same FIO2, the plastic tubing was connected to the wall air supply and the airflow was adjusted to provide the desired FIO2 as described below.
The FIO2 was maintained constant throughout the study and below 40% (beyond such a Heproportion, the physical properties of the HeO2 mixture become comparable to an N2O2 mixture).
Measurements. Measurements were performed before extubation at the end of a 2-h T-piece trial, 10 to 15 min after extubation when patients breathed N2O2 (N2O2-1), after 15 min of HeO2 breathing, and after 15 min of N2O2 (N2O2-2). Data were recorded during a 2-min period in each studied condition once a stable breathing pattern was observed.
Esophageal (Pes) and gastric (Pga) pressures were measured using a double balloon catheter (Marquat, Boissy Saint Léger, France). This catheter was inserted through the nose after the application of topical anesthesia and advanced until the distal balloon was in the stomach and the proximal balloon was in the middle portion of the esophagus. Each balloon, filled with 1 ml of air, was connected to a differential pressure transducer (MP45, ± 100 cm H2O; Validyne, Northridge, CA). Appropriate placement of the esophageal balloon was assessed with the usual method (10). Adequate placement of the gastric balloon was ascertained by gentle manual pressure on the patient's abdomen to observe fluctuations in Pga, as well as by asking the patient to swallow and verifying that the sharp increase in Pes caused by esophageal contraction was not observed on the Pga tracing (11). Transdiaphragmatic pressure (Pdi) was obtained by subtracting the Pes signal from the Pga signal. Pressure signals were digitized at 128 Hz and sampled using an analog-to-digital converter system (MP100; Biopac Systems, Santa Barbara, CA). Flow measurement was not performed because using a pneumotachograph at the mouth would have considerably modified the breathing pattern of the patients by adding equipment dead space, inspiratory resistance, and altering the route of breathing. Because we were expecting moderate changes in effort to breathe and because we wanted to be as close as possible to the clinical situation, we decided against having any measurement of flow.
The FIO2 was measured with an oxygen analyzer (Miniox I; MSA, Pittsburg, PA).
Dyspnea. Changes in comfort with each gas mixture were evaluated by a simple questionnaire when going from N2O2-1 to HeO2, then when switching from HeO2 to N2O2-2, as follows: +2: large improvement, +1: moderate improvement, 0: no change,
1: moderate worsening,
2: major worsening.
A catheter was inserted in a radial artery to allow arterial blood gas analysis at the end of each tested period. Blood gases were measured using an ABL 520 analyzer (Radiometer, Copenhagen, Denmark). Standard three-lead monitoring electrodes continuously monitored heart rate (HR) and rhythm. Systolic (SBP) and diastolic (DBP) arterial blood pressure were continuously monitored through the radial artery catheter signal. Oxygen saturation (SaO2) was continuously monitored by pulse oxymetry.
Data Analysis and Assessment of Patient's Effort
The patient's inspiratory effort was evaluated by pressure indexes: the
esophageal (Pes) and transdiaphragmatic pressures swings (
Pdi), and
the transdiaphragmatic pressure-time product (PTP). These calculations are independent of flow and volume signals. The pressure-time
product is calculated as the integral of pressure over time during the
phase of inspiratory muscular effort. The pressure-time product per
breath for the diaphragm (PTPdi/breath) was obtained by measuring
the area under the Pdi signal from the onset of its positive deflection
to its return baseline (12). After elimination of artefactual cycles corresponding to cough, and esophageal spasms, 10 to 20 successive
breaths were used to compute average values. The average PTPdi/
breath (cm H2O/s) was multiplied by respiratory frequency to obtain
the PTPdi/min (cm H2O/s/min).
Statistical Analysis
Data are given as mean ± SD. The measurements obtained during the four different conditions of the study were compared, the patients serving as their own controls. Two-way repeated measure analysis of variance (ANOVA) was used to test whether the different conditions had significant effects on the different variables, with Fisher's least statistical difference test being used for two-by-two comparisons when appropriate. A paired chi-square test was used to compare the comfort changes with each gas mixture. A p value below 0.05 was considered as the limit of significance.
| |
RESULTS |
|---|
|
|
|---|
Patients
The characteristics of the eighteen patients enrolled in this study are listed in Table 1 (13). All patients were studied in the immediate postextubation period. All 18 patients tolerated well HeO2 breathing. The mean FIO2 was 33 ± 2 %. Two patients (Patients 4 and 17) were reintubated in the following 48 h.
|
Respiratory Rate, Gas Exchange, and Hemodynamics
Respiratory rate and blood gas and hemodynamic parameter values are given in Table 2. All patients but one had a pH value above 7.37, and all had a PaCO2 value below 46 mm Hg. Respiratory rate and blood gas and hemodynamic parameters values did not significantly differ among the four study conditions.
|
Effort to Breathe
The transdiaphragmatic pressure swings and the pressure time product were similar before and after extubation (p = NS).
After extubation, all indexes of respiratory effort decreased
with HeO2 as compared with N2O2, whereas no difference existed between N2O2-1 and N2O2-2. The significant decrease in
the transdiaphragmatic pressure-time product observed with
HeO2 (197 ± 19 for N2O2-1 versus 166 ± 22 for HeO2 versus
201 ± 23 cm H2O/s/min for N2O2-2; p < 0.05) is illustrated in
Figure 1. Individual values for
Pdi (in cm H2O) are presented in Figure 2. The mean reduction of the transdiaphragmatic pressure under HeO2 was 1.5 ± 0.6 cm H2O or 19 ± 5%.
All but three patients presented a decrease in
Pdi with HeO2, ranging from
0.2 to
3.9 cm H2O or
4 to
55%,
and a significant reduction in
Pdi was observed between
HeO2 and N2O2 (10.2 ± 0.7 for N2O2-1 versus 8.6 ± 1.1 for
HeO2 versus 10.0 ± 0.8 cm H2O for N2O2-2; p < 0.05). An experimental record from one patient (Patient 5) illustrates that
swings in Pes, Pga, and Pdi were smaller in HeO2 than in N2O2
(Figure 3).
|
|
|
Comfort to Breathe
Individual values for comfort scale obtained when patient were switched from N2O2-1 to HeO2 and from HeO2 to N2O2-2 are presented in Figure 4 for 16 subjects (in two subjects the information could not be obtained). Breathing helium-oxygen mixture improved comfort to breathe (p < 0.001).
|
| |
DISCUSSION |
|---|
|
|
|---|
Abnormalities of the upper airway and including the glottis could participate to contribute a high degree of inspiratory mechanical loads experienced by some patients in the period after extubation, and contribute to postextubation respiratory distress. The present study is the first evaluation of the effects of a helium-oxygen mixture (HeO2) on inspiratory effort, blood gases, and comfort to breathe in the immediate postextubation period, in non-COPD patients who had been intubated at least for 48 h. The results obtained demonstrate that using HeO2 instead of N2O2 can reduce the inspiratory effort and improve comfort after tracheal extubation.
The use of HeO2 mixture in adults (14) and in children (18) with upper airway obstruction has been reported in several anecdotal series and a few studies, and it has become one of the more accepted indications for HeO2 use (6). Although its effect was shown mainly through observational studies, the fact that the immediate improvement obtained with HeO2 breathing was reversed when it was discontinued even briefly, suggested an independent beneficial effect of the gas related to its physical properties. Barach (14) first reported the use of HeO2 mixture in the treatment of both upper and lower airway obstruction. Skrinskas and colleagues (17) reported their experience with HeO2 treatment in a series of 10 patients with acute upper airway obstruction. Four developed respiratory distress after the removal of an endotracheal tube, three had an obstruction associated with radiation therapy, and in three patients the cause was a primary tumor. Except in a case of intermittent complete blockade of the airway, the use of HeO2 improved the patients, allowing the avoidance of intubation and/or creating time to prepare the patients for a surgical procedure. Duncan (18) reported a pediatric experience in seven patients with upper airway obstruction caused by viral or postintubation severe croup, in whom artificial airways were avoided by the use of HeO2. In a double-blind, randomized, controlled, crossover trial, Kemper and coworkers (21) reported a 38% reduction in respiratory distress score with HeO2 breathing compared with N2O2 in 13 children with burns and other trauma with postextubation stridor. These results are comparable to those of the study of Rodeberg and coworkers (22) who reported the successful use of HeO2 in the treatment of postextubation stridor in six of eight pediatric patients with burns, who were unresponsive to racemic epinephrine.
In the present study, both the effort to breathe and the
comfort were improved with HeO2. All but three patients (Patients 2, 3, and 7) presented a decrease in transdiaphragmatic
pressure measured under HeO2 from a range of
4 to
55%.
Patient 3, who presented an increase in transdiaphragmatic
pressure with helium-oxygen (16.6 to 19.5 cm H2O), had an esotracheal fistula diagnosed subsequently, and the modification
of the gas mixture did not modify the comfort scale. The two
other patients (Patients 2, 7) presented either a moderate improvement in comfort with HeO2 or no change after return to
N2O2. The two patients (Patients 4 and 17) who were reintubated in the following 48 h, presented a 31 and 17% reduction in transdiaphragmatic pressure with helium-oxygen, and moderate and large improvement in comfort with HeO2.
Our study has presented some limitations. First, some patients could discern which gas they were receiving, which could bias the patients' assessment of their degree of dyspnea. Second, the investigators were not blinded to the gas being delivered.
Some studies (23, 24) reported an increase in the work of
breathing after extubation, and the investigators suggested a
role of the supraglottic airway in explaining the increase of work of breathing noticed after extubation. Straus and coworkers (25) reported no significant difference in inspiratory
effort before and after removal of the endotracheal tube in
successfully extubated patients, and these results are confirmed by the present study. Straus and coworkers also recorded the longitudinal area profile of the upper airway after
extubation with the acoustic reflection technique used by
Louis and coworkers (26) and found that in all but one patient, the narrowest part of the airway was the glottis. The
glottis area was found to be smaller than normal, supporting
the idea that mild inflammation of the glottis is probably a frequent phenomenon in this situation. This phenomenon may
explain in part the beneficial effect of helium-oxygen breathing mixture in the present study. The density of the tested
HeO2 mixture used (FIO2 = 0.33) was about two times smaller
than the density of air. This reduced density (
), together with
the almost negligible increase in the dynamic viscosity (µ) of
HeO2 as compared with air (
10%), results in an increase in
kinematic viscosity
(= µ/
) and in a decrease in the Reynolds number (Re). This reduction in Reynolds number is responsible for less turbulent and more laminar flow, and less
airway resistance. These improvements lead to the decrease in
the inspiratory effort.
No patient with COPD was included in the present study because the effect of HeO2 in these patients has been previously studied (9). The observed improvement of inspiratory effort and comfort obtained in the studied patients could have been strongly reinforced by including patients with COPD who do have a reduction in breathing effort with HeO2. In a previous study assessing the effect of HeO2 in patients with COPD (9), three of the 10 patients were studied during an episode of mild-to-moderate respiratory distress after extubation. The use of HeO2 in these three patients, similar to that in the other patients with COPD and respiratory distress, permitted a significant reduction in inspiratory effort and an improvement in gas exchange, with either HeO2 alone in spontaneous ventilation or when combining noninvasive ventilation with HeO2.
In the present study, flow measurements were not performed in order to evaluate inspiratory effort in real conditions, without additional work load induced by the pneumotachograph and equipment, and without altering the route of breathing. Several studies (7, 9, 27), however, have already demonstrated that substituting HeO2 for N2O2 does not significantly modify breathing pattern parameters. The lack of significant alterations in respiratory rate and gas exchange observed in the study is in accordance with these results.
Using HeO2 mixture alone or in association with noninvasive ventilation could be proposed in patients who develop acute respiratory failure in the postextubation period, and especially if upper airway obstruction is part of the pathophysiology of the decompensation. Although the majority of patients who develop postextubation stridor do not require intubation and there is not yet a randomized control trial showing that noninvasive ventilation is benefited after extubation, the present results suggest that, in some cases, HeO2 could be an interesting additional therapy, allowing clinical improvement and giving time to usual therapy to work such as epinephrine inhalation and corticosteroid. Helium is an inert gas, almost insoluble in tissues and nonreactive with biologic membranes, and not associated with any toxic effects. This unloading effect could facilitate the period after extubation in selected patients who develop respiratory distress shortly after extubation and may avoid reintubation in some cases. Further studies are needed to better define its therapeutic role.
| |
Footnotes |
|---|
Correspondence and requests for reprints should be addressed to Pr L. Brochard, Réanimation Médicale, Hôpital Henri Mondor, 94000 Créteil, France. E-mail: laurent. brochard{at}hmn.ap-hop-paris.fr
(Received in original form August 4, 2000 and in revised form May 11, 2001).
Presented in part at the 2000 International conference of the American Thoracic Society, May 5-10, Toronto, Canada.
Acknowledgments:
Supported by Grants from Air Liquide Santé and INSERM
| |
References |
|---|
|
|
|---|
1. Anene O, Meert KL, Uy H, Simpson P, Sarnaik AP. Dexamethasone for the prevention of postextubation airway obstruction: a prospective, randomized, double-blind, placebo-controlled trial. Crit Care Med 1996; 24: 1966-1969 .
2. Darmon JY, Rauss A, Dreyfuss D, Bleichner G, Elkharrat D, Schlemmer B, Tenaillon A, Brun-Buisson C, Huet Y. Evaluation of risk factors for laryngeal edema after tracheal extubation in adults and its prevention by dexamethasone: a placebo-controlled, double-blind, multicenter study. Anesthesiology 1992; 77: 245-251 [Medline].
3. Ho LI, Harn HJ, Lien TC, Hu PY, Wang JH. Postextubation laryngeal edema in adults: risk factor evaluation and prevention by hydrocortisone. Intensive Care Med 1996; 22: 933-936 [Medline].
4.
Epstein SK,
Ciubotaru RL,
Wong JB.
Effect of failed extubation on the
outcome of mechanical ventilation.
Chest
1997;
112:
186-192
5.
Epstein SK,
Ciubotaru RL.
Independent effects of etiology of failure
and time to reintubation on outcome for patients failing extubation.
Am J Respir Crit Care Med
1998;
158:
489-493
6. Manthous CA, Morgan S, Pohlman A, Hall JB. Heliox in the treatment of airflow obstruction: a critical review of the literature. Respir Care 1997; 42: 1034-1042 .
7.
Swidwa DM,
Montenegro HB,
Goldman MD,
Lutchen KR,
Saidel GM.
Helium-oxygen breathing in severe chronic obstructive pulmonary
disease.
Chest
1985;
87:
790-795
8. Manthous CA, Hall JB, Caputo MA, Walter JM, Klochsieben G, Schmidt A, Wood LD. Heliox improves pulsus paradoxus and peak expiratory flow in nonintubated patients with severe asthma. Am J Respir Crit Care Med 1995; 151: 310-314 [Abstract].
9.
Jaber S,
Fodil R,
Carlucci A,
Boussarsar M,
Pigeot J,
Lemaire F,
Harf A,
Lofaso F,
Isabey D,
Brochard L.
Noninvasive ventilation with helium-oxygen in acute exacerbations of chronic obstructive pulmonary disease.
Am J Respir Crit Care Med
2000;
161:
1191-1200
10. Baydur A, Behrakis PK, Zin WA, Jaeger MJ, Milic-Emili J. A simple method for assessing the validity of the esophageal balloon technique. Am Rev Respir Dis 1982; 126: 788-791 [Medline].
11.
Diehl JL,
Lofaso F,
Deleuze P,
Similowski T,
Lemaire F,
Brochard L.
Clinically relevant diaphragmatic dysfunction after cardiac operations.
J Thorac Cardiovasc Surg
1994;
107:
487-498
12. Barnard PA, Levine S. Critique on application of diaphragmatic time tension index to spontaneously breathing humans. J Appl Physiol 1986; 54: 1067-1072 .
13. Le Gall JR, Lemeshow S, Saulnier F. A new simplified acute physiology score (SAPS II) based on a European/North American multicenter study. JAMA 1993; 270: 2957-2963 [Abstract].
14. Barach A. The use of helium in the treatment of asthma and obstructive lesions in the larynx and trachea. Ann Intern Med 1935; 9: 739-765 .
15. Barach A. The thearpeutic use of helium. JAMA 1936; 107: 1273-1275 .
16. Boorstein J, Boorstein SM, Humphries GN, Johnston CC. Using helium-oxygen mixtures in the emergency management of upper airway obstruction. Ann Emerg Med 1989; 18: 688-690 [Medline].
17. Skrinskas G, Hyland R, Hutcheon M. Using helium-oxygen mixtures in the management of upper airway obstruction. Can Med Assoc J 1983; 128: 555-558 [Medline].
18. Duncan P. Efficacy of helium-oxygen mixtures in the management of severe viral and post-intubation croup. Can Anaesth Soc J 1979; 26: 206-212 [Medline].
19. Nelson D, McClellan L. Helium-oxygen mixtures as adjunctive support for refractory viral croup. Ohio State Med J 1982; 26: 206-212 .
20. Kemper KJ, Izenberg S, Marvin JA, Heimbach DM. Treatment of postextubation stridor in a pediatric patient with burns: the role of heliox. J Burn Care Rehabil 1990; 11: 337-339 [Medline].
21. Kemper KJ, Ritz RH, Benson MS, Bishop MS. Helium-oxygen mixture in the treatment of postextubation stridor in pediatric trauma patients. Crit Care Med 1991; 19: 356-359 [Medline].
22. Rodeberg DA, Easter AJ, Washam MA, Housinger TA, Greenhalgh DG, Warden GD. Use of a helium-oxygen mixture in the treatment of postextubation stridor in pediatric patients with burns. J Burn Care Rehabil 1995; 16: 476-480 [Medline].
23.
Nathan SD,
Ishaaya AM,
Koerner SK,
Belman MJ.
Prediction of minimal pressure support during weaning from mechanical ventilation.
Chest
1993;
103:
1215-1219
24.
Ishaaya AM,
Nathan SD,
Belman MJ.
Work of breathing after extubation.
Chest
1995;
107:
204-209
25.
Straus C,
Louis B,
Isabey D,
Lemaire F,
Harf A,
Brochard L.
Contribution of the endotracheal tube and the upper airway to breathing work
load.
Am J Respir Crit Care Med
1998;
157:
23-30
26.
Louis B,
Glass GM,
Fredberg JJ.
Pulmonary airway area by two-microphone acoustic reflection method.
J Appl Physiol
1994;
76:
2234-2240
27. Jolliet P, Tassaux JM, Thouret DP, Chevrolet JC. Beneficial effects of helium-oxygen vs. air-oxygen noninvasive pressure support in decompensated COPD patients. Crit Care Med 1999; 27: 2422-2429 [Medline].
This article has been cited by other articles:
![]() |
M. Brouns, S. T. Jayaraju, C. Lacor, J. De Mey, M. Noppen, W. Vincken, and S. Verbanck Tracheal stenosis: a flow dynamics study J Appl Physiol, March 1, 2007; 102(3): 1178 - 1184. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. A. Wedzicha Heliox in chronic obstructive pulmonary disease: lightening the airflow. Am. J. Respir. Crit. Care Med., April 15, 2006; 173(8): 825 - 826. [Full Text] [PDF] |
||||
![]() |
E. A. Laude, N. C. Duffy, C. Baveystock, B. Dougill, M. J. Campbell, R. Lawson, P. W. Jones, and P. M. Calverley The Effect of Helium and Oxygen on Exercise Performance in Chronic Obstructive Pulmonary Disease: A Randomized Crossover Trial Am. J. Respir. Crit. Care Med., April 15, 2006; 173(8): 865 - 870. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. W. Berkenbosch, R. E. Grueber, G. R. Graff, and J. D. Tobias Patterns of Helium-Oxygen (Heliox) Usage in the Critical Care Environment J Intensive Care Med, November 1, 2004; 19(6): 335 - 344. [Abstract] [PDF] |
||||
![]() |
M. J. TOBIN Critical Care Medicine in AJRCCM 2001 Am. J. Respir. Crit. Care Med., March 1, 2002; 165(5): 565 - 583. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Proc. Am. Thorac. Soc. | Am. J. Respir. Cell Mol. Biol. |