INTRODUCTION |
|---|
|
|
|---|
One of the most frequent reasons for instituting mechanical ventilation is to decrease a patient's work of breathing. To achieve this goal, neuromuscular blockade was initially used to eliminate a patient's respiratory efforts, resulting in controlled ventilation. Neuromuscular blocking agents, however, predispose to numerous hazards, and, today, some means of assisted ventilation is usually employed. Surprisingly, we do not know the ideal level of respiratory muscle unloading during mechanical ventilation. To consider this issue, we need a means of measuring a patient's respiratory effort. Inspiratory pressure-time product (PTP), which quantifies swings in intrathoracic pressure, is closely related to oxygen cost of breathing and has several advantages over measurements of mechanical work of breathing (1). In 17 patients failing a trial of weaning from mechanical ventilation, their average PTP was 388 cm H2O · s/min (2); the average value in healthy subjects is 94 cm H2O · s/min. With carefully selected settings, a patient's inspiratory PTP can be reduced into the normal range with assisted ventilation (3). For the most effective unloading of the inspiratory muscles, the ventilator should cycle in synchrony with the activity of a patient's own respiratory rhythm. The interplay between these two pumps is complex, and problems can arise at several points in the respiratory cycle: the onset of ventilator triggering, the rest of inspiration after triggering, the switch from inspiration to expiration, and the end of expiration.
| |
TRIGGERING OF THE VENTILATOR |
|---|
|
|
|---|
Ventilators provide positive pressure assistance whenever a patient's inspiratory effort decreases pressure in the ventilator circuit by 1 to 2 cm H2O. The latter value, selected by clinicians, is termed the set sensitivity (4). Patients reach the set sensitivity by activating their inspiratory muscles, but they cannot switch off respiratory motor output at the point of triggering (5). As a result, considerable effort is commonly expended over the period of mechanical inflation that follows the trigger phase. The level of patient effort during this post-trigger phase is closely related to a patient's respiratory drive at the point of triggering (r = 0.78) (3). As such, measures that decrease respiratory drive may enhance respiratory muscle rest during mechanical ventilation. If sedatives are used for this purpose, however, the dosage and method of administration need to be carefully selected (6).
If respiratory drive at the point of triggering is important, one might intuitively expect that effort during the time of triggering would determine patient effort during the remainder of inspiration. Such an association was suggested by Giuliani and coworkers (7). Leung and coworkers (3) applied graded levels of pressure support in 11 critically ill patients and achieved a fourfold reduction in overall patient effort but observed no change in patient effort during the time of triggering. The constancy of effort during the trigger phase was probably due to different factors becoming operational as the level of ventilator assistance was varied (Figure 1). When patients were receiving a low level of pressure support, their respiratory drive was high and the time to trigger the machine was short; as a result, they generated large swings in intrathoracic pressure over a short time. When patients received a high level of pressure support, their respiratory drive was lower but more time was needed to trigger the machine; as a result, they generated much smaller swings in intrathoracic pressure, but over a longer time. Thus, increases in the level of ventilator assistance do not substantially decrease patient effort during the time of triggering.
|
Opening a demand valve during pressure triggering can impose substantial effort (4). In an attempt to overcome this problem, flow triggering was introduced. With this approach, the ventilator is triggered when a patient's effort creates a difference between the inspiratory and expiratory base flow in the circuit. Aslanian and coworkers (8) found that the time required for triggering was 43% less with flow triggering than with pressure triggering, and effort during the time of triggering was decreased by 62%. Patient effort during the post-triggering phase, however, was equivalent for pressure and flow triggering. As a result, the overall effect on total patient effort was small, and the clinical benefit of flow triggering appears to be much less than commonly stated.
| |
INEFFECTIVE TRIGGERING |
|---|
|
|
|---|
The increased use of bedside displays of pressure and flow tracings has led to a growing awareness of the frequency with which patients fail to trigger a ventilator (3, 9, 10). When receiving high levels of pressure support or assist-control ventilation, a quarter to a third of a patient's inspiratory efforts may fail to trigger the machine (3). The number of ineffective triggering attempts increases in direct proportion to the level of ventilator assistance (3).
In a study of factors contributing to ineffective triggering, a
decrease in the magnitude of inspiratory effort at a given level
of assistance was not the cause
indeed, effort was 38%
higher during nontriggering attempts than during the triggering phase of attempts that successfully opened the inspiratory
valve (3). Significant differences, however, were noted in the
characteristics of the breaths before the triggering and nontriggering attempts. Breaths before non-triggering attempts had
a higher tidal volume than did the breaths before triggering attempts, 486 ± 19 and 444 ± 16 ml, respectively, and a shorter
expiratory time, 1.02 ± 0.04 and 1.24 ± 0.03 s, respectively. An
abbreviated expiratory time does not allow the lung to return
to its relaxation volume, leading to an increase in elastic recoil
pressure. Indeed, intrinsic positive end-expiratory pressure
(PEEPi) was higher at the onset of nontriggering attempts
than at triggering attempts: 4.22 ± 0.26 and 3.25 ± 0.23 cm
H2O, respectively. Thus, nontriggering results from premature
inspiratory efforts that are not sufficient to overcome the increased elastic recoil associated with dynamic hyperinflation (3).
An elevated PEEPi may result from an increase in elastic
recoil pressure or expiratory muscle activity. The relative contributions of these two factors to ineffective triggering were
investigated in healthy subjects receiving pressure support and
in whom airflow limitation was induced with a Starling resistor
(11). PEEPi was more than 30% higher at the onset of nontriggering attempts than before triggering attempts
similar
to the situation in critically ill patients (3). The magnitude of
expiratory effort, quantified as the expiratory increase in gastric pressure, did not differ before triggering and nontriggering
attempts. In contrast, elastic recoil, estimated as gross PEEPi minus the increase in gastric pressure, was higher before nontriggering attempts than before triggering attempts: for example,
14.1 and 4.9 cm H2O, respectively, at pressure support of 20 cm H2O. That nontriggering was caused by the elastic recoil
fraction of PEEPi, not that resulting from expiratory effort,
suggests that external PEEP might be clinically useful in reducing ineffective triggering.
Although the magnitude of expiratory effort does not appear to influence the success of triggering attempts, the time
that expiratory efforts commence in relation to the cycling of
the ventilator is an important factor. Investigators have quantified the relationship between the onset of expiratory muscle
activity, measured with a wire electrode in the subject's transversus abdominis, and the termination of mechanical inflation
by the ventilator (11). This relationship was quantified in terms
of phase angle: expiratory muscle activity commencing before
the termination of mechanical inflation produces a negative
phase angle, and activity commencing after the end of inflation
gives a positive phase angle. The phase angle between neural
and mechanical events was more negative before nontriggering
attempts than before triggering attempts: for example,
50°
and
19°, respectively, at pressure support of 20 cm H2O (11).
The more negative phase angle means that mechanical inflation continued for a longer time into neural expiration in the
breaths preceding nontriggering attempts. Continuation of mechanical inflation into neural expiration counters expiratory flow, and also decreases the time available for unopposed exhalation. Consequently, elastic recoil increases. In turn, a greater
inspiratory effort will be needed to achieve effective triggering.
In this way, the time that a patient commences an expiratory
effort (in relation to cycling-off of mechanical inflation) partly
determines the success of the ensuing inspiratory effort in triggering the ventilator. Algorithms that achieve better coordination between the end of mechanical inflation and the onset of a
patient's expiration should lessen this form of patient-ventilator asynchrony, although such algorithms have yet to be rigorously evaluated in patients (12).
| |
SETTING OF INSPIRATORY FLOW |
|---|
|
|
|---|
When a patient is first connected to a ventilator, inspiratory flow is set at some default value, such as 60 L/min. Many critically ill patients have an elevated respiratory drive and the initial setting of flow may be insufficient to meet flow demands. As a result, patients will struggle against their own respiratory impedance and that of the ventilator, with consequent increase in the work of breathing. To minimize this likelihood, clinicians commonly employ a much higher flow, e.g., 80 to 100 L/min. Recent studies, however, suggest that a high flow setting may be counterproductive in this situation.
In healthy subjects, Puddy and Younes (13) found that an
increase in inspiratory flow from 30 to 90 L/min caused respiratory frequency to increase from 8.8 to 14.1 breaths/min. The
increase in frequency was nearly complete within the first two
breaths and did not change thereafter despite the development of respiratory alkalosis. This flow-associated tachypnea
has been shown to be independent of breathing route, temperature of inspired gas, and delivered volume; it also persists after anesthesia of the airway, denervation of receptors in the
lungs and chest wall, and during sleep (14). In addition to
the effect of delivered flow on frequency, investigators have
shown that frequency is influenced by delivered tidal volume
an effect seen under isocapnic and hypocapnic conditions, and
during wakefulness and sleep (18, 19).
When studying the effects of a change in a ventilator's flow
or tidal volume, the results may be influenced unwittingly by simultaneous changes in ventilator inspiratory time. When inspiratory flow is increased and tidal volume kept constant,
ventilator inspiratory time must decrease. When tidal volume
is increased and inspiratory flow kept constant, ventilator inspiratory time must increase. A series of experiments has been undertaken in healthy volunteers to delineate the separate influences
of these three settings
flow, tidal volume, ventilator inspiratory time
in mediating flow-associated tachypnea (20).
For inspiratory flow of 30, 60, and 90 L/min, the respective
frequencies were 12.9, 15.5, and 18.2 breaths/min. Because tidal volume was kept constant, an increase in flow was accompanied by a proportional decrease in ventilator inspiratory time.
The increase in frequency was proportional to the decrease in
ventilator inspiratory time (r =
0.69). When flow and tidal
volume were held constant and ventilator inspiratory pauses
of as much as 2 s were imposed, frequency again changed as a
function of ventilator inspiratory time (r =
0.86). When flow
was increased from 30 to 60 L/min, and tidal volume adjusted
to maintain a constant ventilator inspiratory time, frequency
decreased from 17.9 to 16.0 breaths/min. This series of observations shows that imposed inspiratory time can determine
frequency independently of delivered inspiratory flow and
tidal volume (20). The findings have clinical implications. Clinicians often increase tidal volume to lower carbon dioxide
tension. An increase in tidal volume without change in inspiratory flow, however, must be accompanied by a longer inspiratory time, which is likely to decrease frequency. The consequent change in minute ventilation and carbon dioxide will
fall short of that intended. (The opposite scenario holds when
a lower tidal volume is selected.)
The effects of delivered flow, tidal volume, and inspiratory
time on frequency can be interpreted according to current understanding of the Hering-Breuer reflex. First, for a fixed inspiratory time, an increase in inspiratory flow will necessarily
increase end-inspiratory lung volume. An increase in delivered volume causes a progressive increase in the activity of the
slowly adapting receptors (stretch receptors). This activity is
transmitted by the vagus to the brainstem where it inhibits inspiration (negative feedback). Any decrease in neural inspiratory time is also likely to be accompanied by a decrease in
neural expiratory time, because the two phases are closely
linked (21). Accordingly, decreases in neural inspiratory time
and expiratory time secondary to lung inflation are likely to be
major factors in explaining flow-associated tachypnea. Second, a decrease in neural inspiratory time fosters the extension of mechanical inflation into neural expiratory time, which, in
turn, causes prolongation of neural expiratory time (22, 23). A mechanical expiratory time that exceeds neural inspiration
as can occur with imposed inspiratory pauses, higher tidal volume, or slower flow rates
is thus an important modulator of
frequency. In summary, the response of frequency to a change
in ventilator settings is complex, because it depends on the relative dominance of physiologic mechanisms acting in different
directions. High flows are often used with the intent of lowering PEEPi by achieving a shorter inspiratory time and, thus,
allowing more time for exhalation. The associated tachypnea
could cause expiratory time to shorten, and thus increase
PEEPi, or to lengthen, secondary to continuation of mechanical inflation into neural expiration; which of these opposing
scenarios prevails has yet to be determined.
| |
MODE-SPECIFIC EFFECTS OF INSPIRATORY UNLOADING |
|---|
|
|
|---|
Intermittent mandatory ventilation (IMV) was the first mode designed to provide graded levels of assistance. This mode was introduced as a means of resting the respiratory muscles during machine-supported breaths, while avoiding atrophy and fostering reconditioning by allowing the patient to take some spontaneous breaths (24). More than 15 yr elapsed after the introduction of IMV before it was recognized that patients have difficulty in adapting to the intermittent nature of the assistance (25).
Leung and coworkers (3) undertook a head-to-head comparison of the relative effectiveness of graded levels of IMV and pressure support in unloading the inspiratory muscles. Over the entire span of assistance, the rate of change in effort per minute did not differ between the modes (analyzed in terms of slopes). The efficacy of unloading, however, did differ according to the stages of assistance. From zero assistance up to a medium level of assistance (60% of maximum), the decrease in patient effort per minute was greater with pressure support than with IMV; the converse was observed as ventilator assistance was increased further to a maximal level (Figure 2). The changes in effort per breath and frequency help with understanding the different responses (3). Effort per breath decreased linearly as the rate of IMV was increased. Effort per breath also decreased linearly as pressure support was increased to a medium level, but little thereafter, presumably because pressure support provides assistance only if patients makes some effort. Frequency decreased linearly as pressure support was increased; with IMV, frequency changed little until a high level of assistance was provided. As IMV is increased from a medium to a high level, all of a patient's ventilatory requirements will be met and frequency decreases dramatically; thus, a greater rate of respiratory unloading occurred between medium and high levels of assistance with IMV than with pressure support. Because these two modes are used primarily to provide partial, as opposed to complete, assistance, the greater decrease in patient effort per minute with pressure support at 20 to 60% of assistance makes it a clinically more useful mode than IMV.
|
Pressure support and IMV are commonly combined in a given patient. In an international survey of mechanical ventilation (26), this combination tied with assist-control ventilation as the most commonly used mode of ventilation in North America (34% for each). The rationale for combining the modes is unclear, but presumably clinicians use pressure support to overcome the work imposed by the endotracheal tube and demand valve during the nonmandatory breaths. Examining the response of the respiratory centers to this combination of modes provides useful insight into patient-ventilator interaction. A decrease in the number of mandatory breaths produces a decrease in the average tidal volume (3), with inevitable increase in the dead space-to-tidal volume ratio. To avoid a decrease in alveolar ventilation, the patients increased respiratory drive, inspiratory effort, and frequency. Adding pressure support of 10 cm H2O caused a decrease in effort at any given IMV rate. Interestingly, the decrease in effort during the mandatory ventilator breaths was related to the decrease in respiratory drive during the intervening breaths (r = 0.67) (3). In other words, the reduction in drive during the intervening breaths achieved by adding pressure support was carried over to the mandatory breaths, facilitating greater unloading. Combining IMV and pressure support provides a sometimes useful means of achieving a high level of assistance; the combination has a clinical advantage when it is difficult to achieve a high inspiratory flow in the assist-control mode, as with the Siemens 900C ventilator (Siemens Corporation, New York, NY).
| |
INSPIRATION-EXPIRATION SWITCHING |
|---|
|
|
|---|
Patients are commonly ventilated with a volume-cycled mode,
such as assist-control or IMV. "Cycling-off" of mechanical inflation, however, may be based only indirectly on volume. Instead, inspiratory flow is commonly preset and the ventilator
adjusts inspiratory time to achieve a given tidal volume. This
system is more precisely termed time-cycled ventilation. Inflation time is constant with a time-cycled machine, but patients
invariably display considerable breath-to-breath variability in
inspiratory time (27). Accordingly, a patient's neural inspiratory time may be shorter or longer than the inflation time of
the machine. If the machine delivers the set tidal volume before the end of a patient's neural inspiratory time, ventilator
assistance will cease while the patient continues to make an inspiratory effort
with double triggering (two ventilator breaths
for a single effort) a likely consequence (28).
The algorithm for "cycling-off" of mechanical inflation during pressure support varies among brands, but most manufacturers employ some fall in inspiratory flow (29). With the Siemens 900C ventilator, inflation ceases when inspiratory flow decreases to 25% of the peak value; with the Puritan-Bennett 7200 ventilator (Mallincrodt, Inc., St. Louis, MO), a decrease in flow to 5 L/min terminates inflation. Such algorithms are problematic in patients with chronic obstructive pulmonary disease, because increases in resistance and compliance produce a slow time-constant of the respiratory system. The longer time needed for flow to fall to the threshold value can cause mechanical inflation to persist into neural expiration. In 12 patients with chronic obstructive pulmonary disease receiving pressure support of 20 cm H2O, five recruited their expiratory muscles while the machine was still inflating the thorax (30). Interestingly, the patients who recruited their expiratory muscles during mechanical inflation had an average time constant of 0.54 s, compared with an average of 0.38 s in the patients who did not exhibit expiratory muscle activity. The persistence of mechanical inflation into neural expiration is very uncomfortable, as is well recognized with use of inverse-ratio ventilation.
| |
FURTHER IMPLICATIONS |
|---|
|
|
|---|
The respiratory work performed by patients receiving partial
ventilator assistance can be considerable. For example, when patients receive an IMV of 10 breaths/min or pressure support of 7 cm H2O
common settings in clinical practice
inspiratory PTP can exceed 200 cm H2O · s/min (3). Such inspiratory
effort may be sufficient to induce respiratory muscle fatigue or
prevent its recovery. Unfortunately, physical examination and
most monitoring modalities do not reliably detect such elevated respiratory work. With carefully selected settings, mechanical ventilation decreases respiratory work from intolerable levels to the normal range. Lowering patient effort into the
normal range must be balanced against the greater likelihood
of nontriggering and complications at high levels of assistance.
Patients who are successfully weaned from mechanical ventilation have an elevated inspiratory PTP, averaging about 200 cm H2O · s/min (2). Thus, the right target for patient effort
during assisted ventilation is not known. It is also not known
if decreasing patient effort to the normal range, or lower, predisposes to respiratory muscle deconditioning. Circumstantial evidence suggests that assisted ventilation
as opposed to controlled ventilation
is unlikely to produce respiratory muscle
atrophy. Surprisingly little spontaneous activity is sufficient to
prevent atrophy of a limb muscle (31). Moreover, a strategy
requiring assist-control ventilation for at least 22 h a day has
been shown to achieve more rapid weaning from mechanical
ventilation than strategies employing graded levels of ventilator assistance (32, 33).
Future studies of patient-ventilator interaction need to address the mechanisms that determine synchronization of machine cycling with a patient's intrinsic respiratory rhythm (34). For example, the extent of synchronization may differ considerably during sleep versus wakefulness (35). Also, we need studies that define the level of unloading that avoids muscle fatigue while not fostering respiratory muscle atrophy. Lastly, better and simpler tools for monitoring patient effort are required.
| |
Footnotes |
|---|
Correspondence and requests for reprints should be addressed to Martin J. Tobin, M.D., Division of Pulmonary and Critical Care Medicine, Edward Hines Jr., Veterans Affairs Hospital, Route 111N, Hines, IL 60141. E-mail: mtobin2{at}luc.edu
(Received in original form May 26, 2000 and in revised form January 17, 2001).
Acknowledgments:
Supported by a Merit Review grant from the Veterans Affairs Research Service.
| |
References |
|---|
|
|
|---|
1. Tobin MJ. Monitoring respiratory mechanics in spontaneously breathing patients. In: Tobin MJ, editor. Principles and practice of intensive care monitoring. New York: McGraw-Hill; 1998. p. 617-653.
2. Jubran A, Tobin MJ. Pathophysiologic basis of acute respiratory distress in patients who fail a trial of weaning from mechanical ventilation. Am J Respir Crit Care Med 1997; 155: 906-915 [Abstract].
3. Leung P, Jubran A, Tobin MJ. Comparison of assisted ventilator modes on triggering, patients' effort, and dyspnea. Am J Respir Crit Care Med 1997; 155: 1940-1948 [Abstract].
4. Sassoon CSH, Gruer SE. Characteristic of the ventilator pressure- and flow-trigger variables. Intensive Care Med 1995; 21: 159-168 [Medline].
5.
Marini JJ,
Capps JS,
Culver BH.
The inspiratory work of breathing during assisted mechanical ventilation.
Chest
1985;
87:
612-618
6.
Kress JP,
Pohlman AS,
O'Connor MF,
Hall JB.
Daily interruption of
sedative infusions in critically ill patients undergoing mechanical ventilation.
N Engl J Med
2000;
342:
1471-1477
7. Giuliani R, Mascia L, Recchia F, Caracciolo A, Fiore T, Ranieri VM. Patient-ventilator interaction during synchronized intermittent mandatory ventilation: effects of flow triggering. Am J Respir Crit Care Med 1995; 151: 1-9 [Abstract].
8. Aslanian P, El Atrous S, Isabey D, Valente E, Corsi D, Harf A, Lemaire F, Brochard L. Effects of flow triggering on breathing effort during partial ventilatory support. Am J Respir Crit Care Med 1998;157:135-143.
9.
Giannouli E,
Webster K,
Roberts D,
Younes M.
Response of ventilator-dependent patients to different levels of pressure support and proportional assist.
Am J Respir Crit Care Med
1999;
159:
1716-1725
10. Sinderby C, Navalesi P, Beck J, Skrobik Y, Comtois N, Friberg S, Gottfried SB, Lindstrom L. Neural control of mechanical ventilation in respiratory failure. Nature Med 1999; 5: 1433-1436 [Medline].
11.
Parthasarathy S,
Jubran A,
Tobin MJ.
Cycling of inspiratory and expiratory muscle groups with the ventilator in airflow limitation.
Am J
Respir Crit Care Med
1998;
158:
1471-1478
12.
Yamada Y,
Du H.
Analysis of the mechanisms of expiratory asynchrony
in pressure support ventilation: a mathematical approach.
J Appl
Physiol
2000;
88:
2143-2150
13. Puddy A, Younes M. Effect of inspiratory flow rate on respiratory output in normal subjects. Am Rev Respir Dis 1992; 146: 787-789 [Medline].
14. Georgopoulos D, Mitrouska I, Bshouty Z, Webster K, Anthonisen NR, Younes M. Effects of breathing route, temperature and volume of inspired gas, and airway anesthesia on the response of respiratory output to varying inspiratory flow. Am J Respir Crit Care Med 1996; 153: 168-175 [Abstract].
15.
Corne S,
Gillespie D,
Roberts D,
Younes M.
Effect of inspiratory flow
rate on respiratory rate in intubated ventilated patients.
Am J Respir
Crit Care Med
1997;
156:
304-308
16. Mitrouska I, Bshouty Z, Younes M, Georgopoulos D. Effects of pulmonary and intercostal denervation on the response of breathing frequency to varying inspiratory flow. Eur Respir J 1998; 11: 895-900 [Abstract].
17. Georgopoulos D, Mitrouska I, Bshouty Z, Anthonisen NR, Younes M. Effects of non-REM sleep on the response of respiratory output to varying inspiratory flow. Am J Respir Crit Care Med 1996; 153: 1624-1630 [Abstract].
18. Tobert DG, Simon PM, Stroetz RW, Hubmayr RD. The determinants of respiratory rate during mechanical ventilation. Am J Respir Crit Care Med 1997; 155: 485-492 [Abstract].
19.
Puddy A,
Patrick W,
Webster K,
Younes M.
Respiratory control during
volume-cycled ventilation in normal humans.
J Appl Physiol
1996;
80:
1749-1758
20.
Laghi F,
Karamchandani K,
Tobin MJ.
Influence of ventilator settings in
determining respiratory frequency during mechanical ventilation.
Am
J Respir Crit Care Med
1999;
160:
1766-1770
21.
Clark FJ,
von Euler C.
On the regulation of depth and rate of breathing.
J Physiol (London)
1972;
222:
267-295
22.
Younes M,
Vaillancourt P,
Milic-Emili J.
Interaction between chemical
factors and duration of apnea following lung inflation.
J Appl Physiol
1974;
36:
190-201
23.
Zuperku EJ,
Hopp FA,
Kampine JP.
Central integration of pulmonary
stretch receptor input in the control of expiration.
J Appl Physiol
1982;
52:
1296-1315
24. Sassoon CSH. Intermittent mandatory ventilation. In: Tobin MJ, editor. Principles and practice of mechanical ventilation. New York: McGraw-Hill; 1994. p. 221-237.
25. Marini JJ, Smith TC, Lamb VJ. External work output and force generation during synchronized intermittent mechanical ventilation: effect of machine assistance on breathing effort. Am Rev Respir Dis 1988; 138: 1169-1179 [Medline].
26.
Esteban A,
Anzueto A,
Alia I,
Gordo F,
Apezteguia C,
Palizas F,
Cide D,
Goldwaser R,
Soto L,
Bugedo G,
Rodrigo C,
Pimentel J,
Raimondi G,
Tobin MJ.
How is mechanical ventilation employed in the intensive care unit? An international utilization review.
Am J Respir Crit
Care Med
2000;
161:
1450-1458
27. Tobin MJ, Yang KL, Jubran A, Lodato RF. Interrelationship of breath components in neighboring breaths of normal eupneic subjects. Am J Respir Crit Care Med 1995; 152: 1967-1976 [Abstract].
28. Younes M. Interactions between patients and ventilators. In: Roussos C, editor. The thorax, 2nd ed. New York: Marcel Dekker; 1995. p. 2367- 2420.
29. Brochard L. Pressure support ventilation. In: Tobin MJ, editor. Principles and practice of mechanical ventilation. New York: McGraw-Hill; 1994. p. 239-257.
30. Jubran A, Van de Graaff WB, Tobin MJ. Variability of patient-ventilator interaction with pressure-support ventilation in patients with COPD. Am J Respir Crit Care Med 1995; 152: 129-136 [Abstract].
31. Muller EA. Influence of training and of inactivity on muscle strength. Arch Phys Med Rehabil 1970; 51: 449-462 [Medline].
32.
Esteban A,
Frutos F,
Tobin MJ,
Alia I,
Solsona JF,
Valverdu I,
Fernandez R,
de la Cal MA,
Benito S,
Tomas R.
A comparison of four methods of weaning patients from mechanical ventilation. Spanish Lung
Failure Collaborative Group.
N Engl J Med
1995;
332:
345-350
33.
Ely EW,
Baker AM,
Dunagan DP,
Burke HL,
Smith AC,
Kelly PT,
Johnson MM,
Browder RW,
Bowton DL,
Haponik EF.
Effect on the
duration of mechanical ventilation of identifying patients capable of
breathing spontaneously.
N Engl J Med
1996;
335:
1864-1869
34. Parthasarathy S, Jubran A, Tobin MJ. Assessment of neural inspiratory time in ventilator-supported patients. Am J Respir Crit Care Med 2000; 160: 546-552 .
35.
Simon PM,
Zurob AS,
Wies WM,
Leiter JC,
Hubmayr RD.
Entrainment
of respiration in humans by periodic lung inflations: effect of state and
CO2.
Am J Respir Crit Care Med
1999;
160:
950-960
This article has been cited by other articles:
![]() |
J. D. Pierce, C. Goodyear-Bruch, S. Hall, and R. L. Clancy Effect of dopamine on rat diaphragm apoptosis and muscle performance Exp Physiol, July 1, 2006; 91(4): 731 - 740. [Abstract] [Full Text] [PDF] |
||||
![]() |
J.-L. Vincent, M. P. Fink, J. J. Marini, M. R. Pinsky, W. J. Sibbald, M. Singer, P. M. Suter, D. Cook, P. E. Pepe, and T. Evans Intensive care and emergency medicine: progress over the past 25 years. Chest, April 1, 2006; 129(4): 1061 - 1067. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Battisti, D. Tassaux, J.-P. Janssens, J.-B. Michotte, S. Jaber, and P. Jolliet Performance Characteristics of 10 Home Mechanical Ventilators in Pressure-Support Mode: A Comparative Bench Study Chest, May 1, 2005; 127(5): 1784 - 1792. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Fauroux, F. Nicot, S. Essouri, N. Hart, A. Clement, M.I. Polkey, and F. Lofaso Setting of noninvasive pressure support in young patients with cystic fibrosis Eur. Respir. J., October 1, 2004; 24(4): 624 - 630. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J. Tobin Writing a Review Article for AJRCCM Am. J. Respir. Crit. Care Med., October 1, 2003; 168(7): 732 - 734. [Full Text] [PDF] |
||||
![]() |
T. Sharshar, G. Desmarais, B. Louis, G. Macadou, R. Porcher, A. Harf, J.-C. Raphael, D. Isabey, and F. Lofaso Transdiaphragmatic Pressure Control of Airway Pressure Support in Healthy Subjects Am. J. Respir. Crit. Care Med., October 1, 2003; 168(7): 760 - 769. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. de Chazal and R. D. Hubmayr Novel aspects of pulmonary mechanics in intensive care Br. J. Anaesth., July 1, 2003; 91(1): 81 - 91. [Full Text] [PDF] |
||||
![]() |
E. Kondili, G. Prinianakis, and D. Georgopoulos Patient-ventilator interaction Br. J. Anaesth., July 1, 2003; 91(1): 106 - 119. [Full Text] [PDF] |
||||
![]() |
M. Vitacca, L. Barbano, S. D'Anna, R. Porta, L. Bianchi, and N. Ambrosino Comparison of Five Bilevel Pressure Ventilators in Patients With Chronic Ventilatory Failure: A Physiologic Study Chest, December 1, 2002; 122(6): 2105 - 2114. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Parthasarathy and M. J. Tobin Effect of Ventilator Mode on Sleep Quality in Critically Ill Patients Am. J. Respir. Crit. Care Med., December 1, 2002; 166(11): 1423 - 1429. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Yang, J. Luo, J. Bourdon, M.-C. Lin, S. B. Gottfried, and B. J. Petrof Controlled Mechanical Ventilation Leads to Remodeling of the Rat Diaphragm Am. J. Respir. Crit. Care Med., October 15, 2002; 166(8): 1135 - 1140. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Chiumello, P. Pelosi, E. Calvi, L.M. Bigatello, and L. Gattinoni Different modes of assisted ventilation in patients with acute respiratory failure Eur. Respir. J., October 1, 2002; 20(4): 925 - 933. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Younes, J. Kun, K. Webster, and D. Roberts Response of Ventilator-Dependent Patients to Delayed Opening of Exhalation Valve Am. J. Respir. Crit. Care Med., July 1, 2002; 166(1): 21 - 30. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J. TOBIN Critical Care Medicine in AJRCCM 2001 Am. J. Respir. Crit. Care Med., March 1, 2002; 165(5): 565 - 583. [Full Text] [PDF] |
||||
![]() |
N Ambrosino and A Rossi Proportional assist ventilation (PAV): a significant advance or a futile struggle between logic and practice? Thorax, March 1, 2002; 57(3): 272 - 276. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Proc. Am. Thorac. Soc. | Am. J. Respir. Cell Mol. Biol. |