|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| |
ABSTRACT |
|---|
|
|
|---|
Constant continuous positive airway pressure (CPAP) is the treatment of choice for the obstructive sleep apnea syndrome (OSAS). To enable the pressure to be matched more accurately to actual requirements, and thus increase patient acceptance, an autoadjusting device based on the measurement of upper airway impedance was developed (APAPFOT). We investigated the efficacy and compliance in continuous use at home. Fifty-two patients were treated (randomized crossover) with CPAP and APAPFOT for 6 wk each. Respiratory disturbances, sleep profile, and arousals improved significantly with both modes (AHI: baseline, 35.1 ± 26/h; APAPFOT, 5.0 ± 5.2; CPAP, 4.3 ± 6.3; p < 0.001 baseline versus each mode). The mean pressure with APAPFOT was significantly reduced as compared with CPAP (CPAP, 7.8 ± 1.5 cm H2O; APAPFOT, 5.7 ± 1.8 cm H2O; p < 0.001). Under APAPFOT the pressure was lower than that under CPAP for 81.5 ± 21% of the time. Although overall use did not differ, 75% of the patients preferred APAPFOT for home treatment. We conclude that APAPFOT is as efficacious as constant CPAP in the treatment of OSAS. The treatment pressure can be reduced significantly, and sleep microstructure improved with APAPFOT. These might be the reasons for patient preference of automatic therapy.
| |
INTRODUCTION |
|---|
|
|
|---|
Since it was first described by Sullivan and colleagues (1), nasal continuous positive airway pressure (nCPAP) has developed into the standard form of treatment of obstructive sleep apnea syndrome (OSAS). CPAP treatment involves the application throughout the entire treatment period of a constant pressure titrated manually on the basis of the data provided by polysomnography (PSG). The aim of this titration is to determine the lowest pressure capable of reliably preventing all obstructive events during the course of the night. As further developments of this form of treatment, automatic devices (APAP) were designed that make it possible to continuously adjust the pressure to the level actually needed to keep the upper airways patent. With such devices, two objectives are possible: either to determine the pressure level for fixed CPAP home treatment (autotitrating systems), or to replace fixed CPAP by automatically varying pressure during the entire treatment period (autoadjusting systems). In the latter way, it is hoped to reduce pressure-associated side effects of CPAP, thus avoiding treatment failures because of noncompliance. Earlier automated nCPAP devices measured airway obstruction on the basis of a reduction in airflow, flattening of the inspiration flow contour, vibrations of the pharyngeal wall, or the flow regime provided by the CPAP compressor (2).
In 1956, DuBois and colleagues (12) described the forced oscillation technique for the measurement of complex resistances in the airways (impedance). In contrast to flow, flattening, and esophageal pressure, measurement of impedance is independent of patient activity and ventilatory effort, and directly reflects the resistance offered by the airways. Using this method, upper airway obstructions during sleep, that is, obstructive apneas and hypopneas, and the upper airway resistance syndrome could be detected noninvasively (13). In such situations, impedance correlated with the esophageal pressure. During polysomnography, it was possible to distinguish central apneas with open airways from obstructive events (13- 15). In addition, Navajas and colleagues (16) showed that manual nCPAP titration based on impedance measurement permitted an adequate determination of the pressure for constant-pressure treatment. The constant CPAP determined in this way did not differ significantly from that determined by conventional titration. The automatic device based on the forced oscillation technique (APAPFOT) enables the treatment pressure to be adjusted to the actual requirement (autoadjusting device) (19). In short-term observations in the sleep laboratory, we were able to show that with APAPFOT, effective treatment of OSAS was possible with significantly reduced mean treatment pressures as compared with constant CPAP (19).
To date, few data are available on the continuous use of automated treatment systems (8, 22), in particular in comparison with the standard form of treatment (9, 10, 23, 24). As far as we are aware, no results have yet been published on the application of APAPFOT under these conditions. In a prospective, randomized study comparing APAPFOT with standard treatment with constant CPAP, therefore, we investigated the following points:
1. Effectiveness of APAPFOT treatment in terms of polysomnography data over 6 wk.
2. Profile of treatment pressures in continuous use.
3. Patient acceptance and side effects of APAPFOT.
| |
METHODS |
|---|
|
|
|---|
Patients
Fifty-two patients (45 men, 7 women, 54.7 ± 10.1 yr of age; BMI, 32.4 ± 5.8 kg/m2) (all data given as means ± SD) referred to a university sleep laboratory by pneumologists and general practioners (7/99 to 10/99) were recruited to the study. They underwent polysomnography for the
first time in order to investigate hypersomnia, and OSAS was diagnosed (AHI
10/h). All the patients underwent a basic examination of lung function (FEV1, 104.9 ± 26% predicted; Rtot, 0.25 ± 0.1 kPa/
L/s, Body Master Lab, Jaeger, Höchberg, Germany), and all gave their
written consent to participate. The study was approved by the Ethics
Commission of the University of Witten/Herdecke. Forty-seven patients completed the study. Of five patients who terminated their participation prematurely, one patient failed to turn up for either of the
follow-up examinations. Two patients receiving automated treatment
and two receiving constant-pressure CPAP left the study. They were
all recommended to continue treatment with standard CPAP. One patient was excluded after a bronchial carcinoma had been diagnosed.
Impedance and the APAPFOT Device
The forced oscillation technique (FOT) is a noninvasive method for determining complex resistances in the upper and lower airways (12). With the aid of a pump, an oscillating flow is applied to the airways. During the measurement with oscillating flows, not only "real" resistances but also "imaginary" resistances are involved (inertance: inertia of the air column moved, capacitance: extensibility of airways and chest). The term impedance thus includes these complex resistances and describes the mechanical properties of the upper airways, that is, the degree of obstruction in the sleep apnea syndrome (Figure 1).
|
As described elsewhere, (19) the system investigated here (Somnosmart; Weinmann, Hamburg, Germany) is an automatic CPAP device that regulates the treatment pressure in accordance with the impedance of the upper airways. The patient breathes through a tube to which an oscillating flow is applied. This oscillating flow is split between the airways and the device unit. A pressure sensor at the mask picks up a signal that comprises the current treatment pressure (CPAP) and the pressure generated by the oscillating measuring flow (oscillation pressure). Together the tube and device form an unchanging reference resistance, and the relative changes in oscillation pressure thus reflect the impedance of the upper airways, and serve to regulate the CPAP generator via a central processing unit. The frequency of the applied oscillations is 20 Hz. Measurement of flow via a pneumotachograph, and thus computation of absolute impedance, is not provided (19).
During the first 5 min of each application, the device determines a reference value during quiet breathing, which serves as baseline and forms the basis for pressure regulation. An increase in pressure is triggered when the reference value is exceeded by a factor of 1.6 (obstructive event) (19). The pressure increase increment is 0.2 cm H2O/s. At the end of an obstructive event, the treatment pressure is decreased again at a maximum rate of 0.1 cm H2O/s (19). Under APAPFOT the treatment pressure can be varied between 4 and 18 cm H2O. The algorithm defines a central respiratory disturbance when the oscillatory pressure signal is equal to, or not more than 10% lower than, the reference. During these events the pressure is kept constant as long as 2 min.
The patients breathed via a mass-produced standard nose mask (Bubble Mask; ResMed, North Ryde, Australia, SomnoMask; Weinmann, Hamburg, Germany). Humidifiers were not used because we intended to compare the amount of side effects with the different modes. However, the APAPFOT device functions properly with a humidifier in-line or other masks. Nasal prongs could not be used because the pressure sensor can only be applied in a mask. Constant CPAP treatment was performed without a pressure ramp.
Study Design
After establishing daytime sleepiness with the Epworth Sleepiness
Scale (ESS) (25), and performing diagnostic polysomnography, manual titration of the treatment pressure was carried out. During this
procedure, the treatment pressure was increased in incremental steps
of 1 cm H2O/h until respiratory disturbances were minimized and respiration-related arousals reduced to
5/h. The pressure thus titrated
(P man) was used as the treatment pressure under constant-pressure
CPAP. On the following two nights, the effect of treatment with constant CPAP (CPAP-first night) and APAPFOT (APAPFOT-first night)
was studied in the sleep laboratory with PSG. Thereafter, the patients
were treated in randomized order for 6 wk each with both constant
CPAP and APAPFOT. Twenty-four patients began the study with
APAPFOT, 28 with constant CPAP. At the end of each 6-wk treatment
period, PSG was carried out in the sleep laboratory (CPAP-6-wk and
APAPFOT-6 wk), and the patients were questioned about possible side
effects and their personal assessment of the treatment. At the end of
the study, the patients were asked which form of treatment they would
prefer for home use. The compliance data (days/time used) and the
pressure data were read out of the integrated processor. The detailed
testing in the laboratory was performed to confirm adequate treatment before starting the home therapy (CPAP-first night and APAPFOT-first night). After the 6-wk periods of regular use we wanted to establish the benefits based on both polysomnographic and subjective criteria (CPAP-6 wk and APAPFOT-6 wk).
The physicians and technicians doing the manual CPAP titration were not involved in either the evaluation of the results of PSG or the questioning of the patients. Those who did the questioning and cared for the patients in the hospital and sleep laboratory and those who evaluated the results of PSG were blinded to the mode of treatment employed. Only the person who managed the randomization knew about the order of the treatment modes in the individual patients. This person adjusted the devices to the different treatment modes.
The patients were not informed about the sequence in which the two treatment modes were applied. In order to avoid product-specific differences such as design, they used only one device for both treatment periods, which permits ventilation in both modes. The patients were informed that they would be treated in two different modes, one with constant pressure, one with variable pressure, but they were not told which mode was the actual standard of treatment and which was the new mode. The patients were asked to contact one of us (W.R.) in case of problems with the treatment during the 6-wk periods. For this purpose a telephone emergency hotline (24-h) was at their disposal.
Polysomnography
The following parameters were recorded: electroencephalogram C4A1 or C3A2, submental or pretibial electromyogram, electrooculogram, effort (thoracic and abdominal induction plethysmography, Consumable Sensor Respiratory Band; Compumedics, Abbotsford, Australia), respiratory flow (thermo-elements, Thermistor Typ 5K3A1; Delta Regeltechnik, München, Germany), snoring signals (laryngeal microphone, ProTec, Woodinville, WA), oxygen saturation (finger pulse oximeter, Oxisensor II D25; Nellcor Puriton Bennett Inc., Pleasanton, CA). Data were recorded electronically (Compumedics). The polysomnography was scored manually. The analysis of the sleep stages was carried out in accordance with the guidelines of Rechtschaffen and Kales (26) and the ASDA criteria (27). Arousals were defined as respiration-induced when they occurred at the earliest with the onset, at the latest 2 s after the end, of an apnea or hypopnea. An apnea was defined as the cessation of respiratory flow for at least 10 s. A hypopnea was defined as a reduction in effort of 50% in comparison with baseline for at least 10 s together with a decrease in oxygen saturation of at least 4%. To quantify snoring, the number of epochs (30 s/page) with evidence of microphone signals for at least 2 s outside of movement artefacts were counted.
Statistics
For the computation of significant differences in dependent parameters (between baseline measurement and each treatment mode, and between the two modes), the ANOVA test with Bonferroni correction (p < 0.05) was used. The McNemar test was used to detect significant differences in device preference. To establish a relationship between treatment pressure and anthropometric and polysomnographic data, Pearson's correlation and simple regression were determined. To determine distribution as normal the Kolmogorov-Smirnov test was used (28).
| |
RESULTS |
|---|
|
|
|---|
Both constant CPAP and APAPFOT improved respiratory disturbances highly significantly at both measuring time points. AHI was reduced from 35.1 ± 26/h (baseline) to 5.3 ± 5.6 (APAPFOT-first night), 4.6 ± 4.8 (CPAP-first night), 5.0 ± 5.2 (APAPFOT-6 wk) and 4.3 ± 6.3 (CPAP-6 wk) (p < 0.001 between baseline and each treatment mode). Snoring was also decreased highly significantly in comparison with baseline: 49 ± 36/h (baseline), 10 ± 17 (APAPFOT-first night), 6 ± 11 (CPAP-first night), 13 ± 20 (APAPFOT-6 wk), and 6 ± 13 (CPAP-6 wk) (p < 0.001 between baseline and each treatment mode) (see Table E1 in online data supplement at www.atsjournals.org).
Also the total number of arousals and the number of respiration-related arousals were highly significantly reduced. Arousals decreased from 34.0 ± 21.7/h (baseline) to 10.3 ± 6.4 (APAPFOT-first night), 9.9 ± 5.5 (CPAP-first night), 10.9 ± 5.7 (APAPFOT-6 wk), and 12.6 ± 8.3 (CPAP-6 wk) (p < 0.001 between baseline and each treatment mode). With both modes, a significant improvement in REM sleep was observed. Both modes showed an improvement in daytime sleepiness, which reached the level of significance after 6 wk of treatment with APAPFOT (Table E1). No significant differences were found between the two treatment modes in terms of polysomnographic data.
Under APAPFOT, significantly lower mean pressures were seen at both time points (first night: 6.6 ± 2.4 cm H2O, p < 0.05; and 6 wk: 5.7 ± 1.8 cm H2O, p < 0.001) in comparison with manually titrated treatment pressure (7.8 ± 1.5) (see Table E2 in online data supplement). The 95th percentile of the treatment pressure under APAPFOT over 6 wk (9.0 ± 3.9) did not differ significantly from the constant CPAP. Although the device could apply treatment pressures between 4 and 18 cm H2O, the maximum pressure during the 6-wk period with APAPFOT was 12.1 ± 4.1 cm H2O. For 81.5 ± 21% of the 6-wk treatment period, the treatment pressure under APAPFOT was lower than the manually titrated pressure (Figure 2, Table E2). A significant correlation was found between BMI and Pman in CPAP, and Pmean in APAPFOT (Figure 3). In all subgroups of patients (grouped by BMI, age, sex, baseline AHI, ESS, and manually titrated treatment pressure), a reduction in Pmean vis-a-vis Pman was observed.
|
|
The patients used the devices 98.4% of the days, under both APAPFOT and constant CPAP. The duration of use was 315.4 ± 94.7 min/d (APAPFOT) and 315.4 ± 97.4 min/d (constant CPAP) (NS). Of the 52 patients initially recruited, 47 (90.4%) completed the study. Of these 47 patients, 35 (75%) gave preference to APAPFOT for long-term treatment at home, and 12 opted for constant CPAP (p < 0.01). Those showing a preference for APAPFOT had used this form of treatment for 334.9 ± 88.4 min/d, compared with 320.0 ± 108.8 min/d for CPAP (NS). Those preferring constant CPAP had used this mode for 323.5 ± 76.7 min/d and APAPFOT for 283.8 ± 92.5 min/d (p < 0.01). Side effects were mild under both modes, and no significant differences were observable. As expected, under APAPFOT, the patients were more aware of pressure variations. These variations in pressure under APAPFOT were also significantly more strongly felt to be unpleasant (scale 0 to 6: APAPFOT, 2.7 ± 2.3; CPAP, 1.3 ± 1.9; p < 0.01). This was particularly the case in those patients who on conclusion of the study gave preference to CPAP (APAPFOT, 4.4 ± 2.2; CPAP, 2.4 ± 2.2; p < 0.05).
| |
DISCUSSION |
|---|
|
|
|---|
As far as we are aware, our study presents for the first time data on patient acceptance and self-assessment of continuous home treatment using an automatic CPAP device in comparison with constant CPAP. Also for the first time it investigates the question wether the autoadjusting CPAP treatment on the basis of impedance measurement (APAPFOT) permits adequate treatment of OSAS under this condition. APAPFOT proved to be equally as effective as the standard treatment with constant CPAP in terms of the respiration parameters and sleep profile. In agreement with the polysomnographic findings, including the number of arousals, we were also able to demonstrate the effectiveness of the therapy on the basis of subjective patient criteria. Thus, daytime sleepiness as measured on the Epworth Sleepiness Scale significantly improved after 6 wk of treatment.
Normalization of respiratory disturbances and sleep architecture under APAPFOT was achieved with a significant reduction in mean treatment pressure measured over the whole observation period of more than 2 cm H2O, in comparison with constant CPAP (Figure 4). The pressure profile and the size of the pressure reduction under APAPFOT were investigated for possible dependence on anthropometric data and the data of baseline polysomnography. For each of the treatment modes, a significant correlation was found between body mass index (BMI) and treatment pressure (Figure 3). The extent of the reduction in treatment pressure under APAPFOT was equal in all subgroups of patients with differences in age, BMI, sex, AHI, sleep profile, or compliance. The reduction in pressure under APAPFOT confirmed our earlier results obtained in short-term treatment (19). The large majority of the patients investigated gave preference to autoadjusting CPAP treatment for long-term home use. It may be assumed that this highly significant difference (p < 0.01) in acceptance is based on the reduction in the treatment pressure under APAPFOT.
|
Studies by Condos and colleagues (29) and Monserrat and coworkers (30) showed that a reduction in constant CPAP pressure of 2 mbar was associated with partial collapse of the upper airways (29, 30). Our present data are not incompatible with this. Although a highly significant reduction in the mean pressure over a period of 6 wk was to be seen under APAPFOT, the 95th percentile of the pressure range applied did not differ significantly from manually titrated constant CPAP. This implies that, in phases of high upper airways resistance, the pressure applied was identical to, or higher than, that titrated manually. This means that there was no constant reduction in treatment pressure; rather, the obvious reduction in mean treatment pressure was due to pressure-lowering during phases of low pressure requirement. In our view, the data on sleep microstructure and macrostructure show that APAPFOT reduced Pmean while still ensuring appropriate treatment of OSAS. The method of manual titration included oxygen desaturation in contrast to the impedance-controlled treatment. However, we do not believe that one of the algorithms is more rigorous in titration than the other because the amounts of oxygen desaturation and respiratory disturbances did not differ significantly between the two treatment modes.
To date, only few data are available on continuous treatment with autoadjusting CPAP systems, in particular in direct comparison with constant CPAP. In an open study involving patients pretreated with constant CPAP, Boudewyns and colleagues (22) discovered a significant improvement in the polysomnographic data with an autoadjusting CPAP system, as had been seen earlier with constant CPAP. Although that study did not collect data on compliance, four of the 15 patients investigated complained of an increase in excessive daytime sleepiness under autoadjusting treatment. Teschler and colleagues (23) reported a reduction in median pressure over a period of 8 wk under the automatic device based on flattening, but no significant differences in compliance data. Meurice and colleagues (10) found a significantly higher daily use over a 3-wk period as compared with constant CPAP. In a cohort study lasting 3 to 6 mo, Konermann and colleagues (24) demonstrated a significant improvement in the numbers of days of use of the automatic device with identical duration of application/d. In our randomized crossover study, no significant differences were to be found in the mean compliance data of the overall group. We believe that this is due to the fact that in our study, both systems were utilized on more than 98% of the days of the observation period, and for more than 5.3 h per day. We observed a greater compliance (significant in constant CPAP) on the part of patients vis-a-vis the respective system to which they finally gave preference. Therefore, if patients have the opportunity to test both modes their preference may be an indicator of the compliance. Subjective data on sleep quality, self-assessment, and patients' preference were not reported by Teschler (23) or Konermann (24). However, as we did not directly compare APAPFOT with other autoadjusting devices our data do not prove the superiority of one method.
According to Badia and colleagues (31) accurate measurement of impedance and detection of leakage mandate the simultaneous determination of airflow and pressure variations. Furthermore, in the event of leakages at the mouth, the elevation of the soft palate might close the posterior nasal airway. This closure results in an increase in impedance, which would prompt a pressure increase of the automatic device. The APAPFOT device does not prevent such an occurrence since it has no integrated facility for flow measurement. Although it is conceivable that these disturbances might trigger a pressure increase, a significant decrease in treatment pressure while providing effective treatment of OSAS has nevertheless been shown for APAPFOT over 6 wk of therapy too.
On the basis of the results of this study, therefore, we conclude that (1) autoadjusting treatment based on measurement of impedance enables effective treatment of OSAS over 6 wk at home. As is the case with conventional treatment with constant CPAP, a correlation between treatment pressure and BMI is found for APAPFOT too. (2) Under APAPFOT, a significant reduction in treatment pressure is achieved. (3) The majority of patients give preference to the automated system for long-term home treatment.
| |
Footnotes |
|---|
Correspondence and requests for reprints should be addressed to Winfried J. Randerath M.D., Klinik Ambrock, Department of Pneumology, Allergology and Sleep Medicine, Ambrocker Weg 60, 58091 Hagen, Germany. E-mail: winfried. randerath{at}dland.de
(Received in original form June 30, 2000 and in revised form November 9, 2000).
The devices were supplied by the Weinmann Company, Hamburg, Germany.| |
References |
|---|
|
|
|---|
1. Sullivan CE, Issa FG, Berthon-Jones M, Eves L. Reversal of obstructive sleep apnea by continuous positive airway pressure applied through the nares. Lancet 1981; 8225: 862-865 .
2. Behbehani K, Yen FC, Burk JR, Lucas EA, Axe JR. Automatic control of airway pressure for treatment of obstructive sleep apnea. IEEE Trans Biomed Eng 1995; 42: 1007-1016 [Medline].
3. Berthon-Jones M, Lawrence S, Sullivan CE, Grunstein R. Nasal continuous positive pressure treatment: current realities and future. Sleep 1996; 19: S131-S135 [Medline].
4. Fleury B, Rakotonanahary D, Hausser-Hauw C, Lebeau B, Guilleminault C. A laboratory validation study of the diagnostic mode of the AutoSet system for sleep-related respiratory disorders. Sleep 1996; 19: 502-505 [Medline].
5. Gugger M. Comparison of ResMed AutoSet (version 3.03) with polysomnography in the diagnosis of the sleep apnea/hypopnea syndrome. Eur Respir J 1997; 10: 587-591 [Abstract].
6. Sériès F. Auto CPAP in the treatment of sleep apnea hypopnea syndrome. Sleep 1996; 19: S281-283 [Medline].
7. Teschler H, Berthon-Jones M, Thompson AB, Henkel A, Henry J, Konietzko N. Automated continuous positive airway pressure titration for obstructive sleep apnea syndrome. Am J Respir Crit Care Med 1996; 154: 734-740 [Abstract].
8. Teschler H, Farhat AA, Exner V, Konietzko N, Berthon-Jones M. AutoSet nasal CPAP titration: constancy of pressure, compliance and effectiveness at 8 months follow-up. Eur Respir J 1997; 10: 2073-2078 [Abstract].
9. Sharma S, Wali S, Pouliot Z, Peters M, Neufeld H, Kryger M. Treatment of obstructive sleep apnea with a self-titrating continuous positive airway pressure (CPAP) system. Sleep 1996; 19: 497-501 [Medline].
10. Meurice J-C, Marc I, Sériès F. Efficacy of auto-CPAP in the treatment of obstructive sleep apnea/hypopnea syndrome. Am J Respir Crit Care Med 1996; 153: 794-798 [Abstract].
11. Meurice J-C, Paquereau J, Denjean A, Patte F, Sériès F. Influence of correction of flow limitation on continuous positive airway pressure efficiency in sleep apnoea/hypopnoea syndrome. Eur Respir J 1998; 11: 1121-1127 [Abstract].
12.
DuBois AB,
Brody AW,
Lewis DH,
Burgess BF.
Oscillation mechanics
of lungs and chest in man.
J Appl Physiol
1956;
8:
587-594
13. Hoster M, Schlenker E, Rühle KH. Influence of automated self-setting CPAP systems on sleep and breathing in sleep apnea syndrome. Wien Med Wochenschr 1996; 146: 385-387 [Medline].
14. Rühle KH. Oszillatorische Impedanz - Grundlagen und klinische Anwendungsmöglichkeiten. In: Rühle KH, editor. Oszillatorische impedanz bei schlafbezogenen atemregulationsstörungen, 1st ed. Stuttgart, New York: Thieme; 1996. p. 1-5.
15. Rühle KH, Schlenker E, Randerath W. Upper airway resistance syndrome. Respiration 1994; 64: 29-34 .
16. Farré R, Rotger M, Montserrat JM, Navajas D. A system to generate simultaneous forced oscillation and continuous positive airway pressure. Eur Respir J 1997; 10: 1349-1353 [Abstract].
17. Farré R, Peslin R, Rotger M, Navajas D. Inspiratory dynamic obstruction detected by forced oscillation during CPAP: a model study. Am J Respir Crit Care Med 1997; 15: 952-956 .
18.
Navajas D,
Farré R,
Rotger M,
Montserrat JM.
Assessment of airway
obstruction during CPAP by means of the forced oscillation technique
in patients with SAS.
Am J Respir Crit Care Med
1998;
157:
1526-1530
19.
Randerath W,
Parys K,
Feldmeyer F,
Rühle KH.
Self-adjusting nCPAP-therapy based on measurement of impedance: a comparison of two
different maximum pressure levels.
Chest
1999;
116:
991-999
20. Randerath W, Parys K, Feldmeyer F, Rühle KH. Selfadjusting nCPAP-therapy based on forced oscillation technique in obstructive sleep apnea syndrome. Respiration 2000; 67: 272-279 [Medline].
21. Randerath W, Galetke W, David M, Siebrecht H, Sanner B, Rühle KH. Prospective randomized comparison of impedance-controlled auto-CPAP (APAPFOT) with constant CPAP. Sleep Med (In press)
22.
Boudewyns A,
Grillier-Lanoir V,
Willemen MJ,
De Cock WA,
van de
Heyning PH,
de Backer WA.
Two months follow up of auto-CPAP
treatment in patients with obstructive sleep apnea.
Thorax
1999;
54:
147-149
23. Teschler H, Wessendorf TE, Farhat AA, Konietzko N, Berthon-Jones M. Two months auto-adjusting versus conventional nCPAP for obstructive sleep apnoea syndrome. Eur Respir J 2000; 15: 990-995 [Abstract].
24.
Konermann M,
Sanner BM,
Vyleta M,
Laschewski F,
Groetz J,
Sturm A,
Zidek W.
Use of conventional and self-adjusting nasal continuous
positive airway pressure for treatment of severe obstructive sleep apnea syndrome.
Chest
1998;
113:
714-718
25. Johns MW. A new method for measuring daytime sleepiness: the Epworth sleepiness scale. Sleep 1991; 14: 540-545 [Medline].
26. Rechtschaffen A, Kales A. A manual of standardized terminology techniques and scoring system for sleep stages of human subjects. University of California, Los Angeles: Brain Information Service; 1968.
27. ASDA Task Force. EEG arousals: scoring rules and examples. Sleep 1992;15:173-184.
28. Sachs L. Angewandte Statistik. Statistische Methoden und ihre Anwendungen, 5th ed. Heidelberg: Springer; 1978;230-246.
29. Condos R, Norman RG, Krishnasamy I, Peduzzi N, Goldring RM, Rapoport DM. Flow limitation as a noninvasive assessment of residual upper-airway pressure therapy of obstructive sleep apnea. Am J Respir Crit Care Med 1994; 150: 475-480 [Abstract].
30. Montserrat JM, Ballester E, Olivi H, Reolid A, Llobers P, Morello A, Rodriguez-Roisin R. Time-course of stepwise CPAP titration. Am J Respir Crit Care Med 1995; 152: 1854-1859 [Abstract].
31.
Badia JR,
Farré R,
Kimoff RJ,
Ballester E,
Hernández L,
Rotger M,
Navajas D,
Montserrat JM.
Clinical application of the forced oscillation technique for CPAP titration in the sleep apnea/hypopnea syndrome.
Am J Resp Crit Care Med
1999;
160:
1550-1554
This article has been cited by other articles:
![]() |
D. Damjanovic, A. Fluck, H. Bremer, J. Muller-Quernheim, M. Idzko, and S. Sorichter Compliance in sleep apnoea therapy: influence of home care support and pressure mode Eur. Respir. J., April 1, 2009; 33(4): 804 - 811. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Pevernagie, J. F. Masa, J. C. Meurice, R. Farre, O. Marrone, and D. Rodenstein Treatment of obstructive sleep-disordered breathing with positive airway pressure systems Eur. Respir. Rev., December 1, 2007; 16(106): 125 - 131. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Nilius, A. Happel, U. Domanski, and K.-H. Ruhle Pressure-Relief Continuous Positive Airway Pressure vs Constant Continuous Positive Airway Pressure: A Comparison of Efficacy and Compliance. Chest, October 1, 2006; 130(4): 1018 - 1024. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Rigau, J. M. Montserrat, H. Wohrle, D. Plattner, M. Schwaibold, D. Navajas, and R. Farre Bench model to simulate upper airway obstruction for analyzing automatic continuous positive airway pressure devices. Chest, August 1, 2006; 130(2): 350 - 361. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. M. Nolan, S. Ryan, T. M. O'Connor, and W. T. McNicholas Comparison of three auto-adjusting positive pressure devices in patients with sleep apnoea Eur. Respir. J., July 1, 2006; 28(1): 159 - 164. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. L. Dellaca, M. Rotger, A. Aliverti, D. Navajas, A. Pedotti, and R. Farre Noninvasive detection of expiratory flow limitation in COPD patients during nasal CPAP Eur. Respir. J., May 1, 2006; 27(5): 983 - 991. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Nussbaumer, K. E. Bloch, T. Genser, and R. Thurnheer Equivalence of Autoadjusted and Constant Continuous Positive Airway Pressure in Home Treatment of Sleep Apnea Chest, March 1, 2006; 129(3): 638 - 643. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. R. Westbrook, D. J. Levendowski, M. Cvetinovic, T. Zavora, V. Velimirovic, D. Henninger, and D. Nicholson Description and Validation of the Apnea Risk Evaluation System: A Novel Method To Diagnose Sleep Apnea-Hypopnea in the Home Chest, October 1, 2005; 128(4): 2166 - 2175. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Abdenbi, B. Chambille, and P. Escourrou Bench testing of auto-adjusting positive airway pressure devices Eur. Respir. J., October 1, 2004; 24(4): 649 - 658. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. A. Pevernagie, P. M. Proot, K. B. Hertegonne, M. C. Neyens, K. P. Hoornaert, and R. A. Pauwels Efficacy of Flow- vs Impedance-Guided Autoadjustable Continuous Positive Airway Pressure: A Randomized Cross-over Trial Chest, July 1, 2004; 126(1): 25 - 30. [Abstract] [Full Text] [PDF] |
||||
![]() |
O. Senn, T. Brack, F. Matthews, E. W. Russi, and K. E. Bloch Randomized Short-term Trial of Two AutoCPAP Devices versus Fixed Continuous Positive Airway Pressure for the Treatment of Sleep Apnea Am. J. Respir. Crit. Care Med., December 15, 2003; 168(12): 1506 - 1511. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. F. Fitzpatrick, C. E. D. Alloway, T. M. Wakeford, A. W. MacLean, P. W. Munt, and A. G. Day Can Patients with Obstructive Sleep Apnea Titrate Their Own Continuous Positive Airway Pressure? Am. J. Respir. Crit. Care Med., March 1, 2003; 167(5): 716 - 722. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. A. Massie, N. McArdle, R. W. Hart, W. W. Schmidt-Nowara, A. Lankford, D. W. Hudgel, N. Gordon, and N. J. Douglas Comparison between Automatic and Fixed Positive Airway Pressure Therapy in the Home Am. J. Respir. Crit. Care Med., January 1, 2003; 167(1): 20 - 23. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Farre, J. M. Montserrat, J. Rigau, X. Trepat, P. Pinto, and D. Navajas Response of Automatic Continuous Positive Airway Pressure Devices to Different Sleep Breathing Patterns: A Bench Study Am. J. Respir. Crit. Care Med., August 15, 2002; 166(4): 469 - 473. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. J. Randerath, M. Heise, R. Hinz, and K.-H. Ruehle An Individually Adjustable Oral Appliance vs Continuous Positive Airway Pressure in Mild-to-Moderate Obstructive Sleep Apnea Syndrome* Chest, August 1, 2002; 122(2): 569 - 575. [Abstract] [Full Text] [PDF] |
||||
![]() |
W.J. Randerath, J. Meier, H. Genger, U. Domanski, and K-H. Ruhle Efficiency of cold passover and heated humidification under continuous positive airway pressure Eur. Respir. J., July 1, 2002; 20(1): 183 - 186. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J. TOBIN Sleep-Disordered Breathing, Control of Breathing, Respiratory Muscles, and Pulmonary Function Testing in AJRCCM 2001 Am. J. Respir. Crit. Care Med., March 1, 2002; 165(5): 584 - 597. [Full Text] [PDF] |
||||
![]() |
F. Lofaso, K. Leroux, M.A. Quera-Salva, G. Mroue, M.P. D'Ortho, D. Isabey, and B. Louis Snoring detection during auto-nasal continuous positive airway pressure Eur. Respir. J., January 1, 2002; 19(1): 108 - 112. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Lévy and J.-L. Pépin Autoadjusting Continuous Positive Airway Pressure . What Can We Expect? Am. J. Respir. Crit. Care Med., May 1, 2001; 163(6): 1295 - 1296. [Full Text] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Proc. Am. Thorac. Soc. | Am. J. Respir. Cell Mol. Biol. |