An Experimental and Clinical Study |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| |
ABSTRACT |
|---|
|
|
|---|
Nebulized aerosols are commonly used to deliver drugs into the lungs of patients with cystic fibrosis (CF). The aim of this study was to assess the effectiveness of pressure-support (PS) ventilation in increasing aerosol deposition within the lungs of children with CF. An in vitro study demonstrated the feasibility of coupling a breath-actuated nebulizer to a PS device. An in vivo study was done with 18 children (ages 6 to 21 yr) with clinically stable CF, each of whom underwent both a standard and a PS-driven ventilation scan (control session and PS session, respectively). In addition, a perfusion scan was used to determine lung outlines and to construct a geometric model for quantifying aerosol deposition by radioactivity counting in MBq. Homogeneity of nebulization was evaluated from the four first-order moments of aerosol distribution in the peripheral and central lung regions. The time-activity nebulization curve was linear in all patients, with higher slopes during the PS than during the control session (0.43 ± 0.07 [mean ± SD] MBq/min and 0.32 ± 0.23 MBq/min, respectively; p < 0.018). Quantitatively, aerosol deposition was about 30% greater after the PS session (4.4 ± 2.7 MBq) than after the control session (3.4 ± 2.1 MBq; p < 0.05). Similarly, deposition efficacy (as a percentage of nebulizer output) was significantly better during the PS session than during the control session (15.3 ± 8.3% versus 11.5 ± 5.7%, p < 0.05). No differences in the regional deposition pattern or in homogeneity of uptake were observed. In conclusion, our data show that driving the delivery of a nebulized aerosol by noninvasive PS ventilation enhances total lung aerosol deposition without increasing particle impaction in the proximal airways.
| |
INTRODUCTION |
|---|
|
|
|---|
Nebulized aerosols are commonly used to deliver drugs to the lungs of patients with cystic fibrosis (CF). Any potential advantages of aerosol therapy would lie in an ability to deliver smaller doses of drugs directly to the site of interest (i.e., the lungs). However, evaluating both the amount of medication actually delivered to the lung and its exact site of deposition remain crucial problems. Few data are available on quantitative drug deposition in lungs affected by CF (1, 2) or on drug deposition in the lungs of children with CF (3). In addition to total lung deposition, the distribution of drug deposition across lung regions is important. In CF, the goal is to deposit a drug in the peripheral bronchi and bronchioles.
Bronchoconstriction or airway obstruction by mucus or inflammation diverts more airflow to nonobstructed airways. Heterogeneous drug deposition has been demonstrated in adults with CF (6). Deposition of an aerosol in the lung may vary widely according to many parameters including the type of nebulizer and type of compressor used to produce the aerosol, the nebulizer fill, injected flow, and breathing pattern (7, 8). In particular, the deposition fraction may change substantially when the breathing pattern changes. The optimal breathing pattern for aerosolization combines a large tidal volume (VT) and a slow inspiratory flow rate; higher flow rates seem to increase drug deposition in the oropharynx. Deep breathing has been shown to increase aerosol deposition in the lungs of both adults and children with CF (7, 9).
Inspiratory pressure-support (PS) ventilation is a method of ventilatory assistance that maintains a constant, preset, positive airway pressure during spontaneous inspiration, with the goal of decreasing the patient's inspiratory work of breathing. Allowing patients with severe chronic obstructive pulmonary disease (COPD) to control their own respiratory rate (RR), VT, and inspiratory time (TI) was associated with a significant decrease in the work performed by the inspiratory muscles, especially the diaphragm, and with significant improvements in CO2 elimination and oxygenation (10). One of the main advantages of PS ventilation is its good acceptability by patients, owing to the patient's control of the frequency and duration of inspiratory assistance, rather than control by the machine (10). Recently, we showed that PS ventilation given to CF patients via a nasal mask during chest physiotherapy was beneficial in preventing oxygen desaturation (11). In these patients, we observed a decrease in RR and an increase in both VT and minute ventilation (VE). These changes brought the breathing pattern closer to that associated with optimal aerosol deposition. When the present study was begun, the effects of PS ventilation on aerosol transport and particle deposition in the lungs were unknown. We designed the study to evaluate these effects.
The aim of our study was to conduct a precise evaluation of the efficacy of PS ventilation in optimizing aerosol deposition in vitro and in vivo. In the first part of the study, we tested the in vitro performances of two breath-actuated nebulizers of similar design, of which one was used alone (i.e., was triggered by the patient's inspiratory flow), whereas the other was coupled to the PS device (i.e., was triggered by the positive pressure generated by the PS device). These two nebulizer systems were compared on the basis of: (1) the total nebulizer output; and (2) the amount of solution captured on an absolute filter located at the end of a cranked tube mimicking the upper airways to the site of the carina. The optimal nebulizer and home PS ventilator settings for the clinical study were determined from this bench study.
The second part of the study evaluated the potential clinical utility of nebulization coupled to PS in children with CF by comparing the lung deposition of a radiolabeled aerosol produced by nebulization alone and by nebulization with PS. We quantified total and regional lung depositions of the aerosol and studied the distribution dynamics of the radiolabeled solution delivered by the two systems. Both our in vitro results and our in vivo findings in children with CF suggest that aerosol deposition at the sites of interest in CF can be improved by using a nebulizer coupled to a PS ventilator.
| |
METHODS |
|---|
|
|
|---|
Experimental In Vitro Study
Apparatus tested. The PS device used in the study was a home ventilator commonly used in our department (Onyx; Mallinckrodt, Les Ulis, France), whose performance is very similar to that of intensive-care PS devices (12). To minimize deposition of the aerosol in the ventilator circuit, PS device was connected by a T-tube to the mouthpiece of the nebulizer used in the study.
Two similar breath-actuated nebulizers were used. Both nebulizers deliver drug in a timed pulse during the inspiratory phase of respiration. The pulse can be timed from 200 ms to 900 ms according to
the patient's breathing pattern. A constant volume of solution is thus
available for inhalation during each spontaneous inspiration. The volume of nebulized solution increases with the inspiratory nebulization
time. With a 3.6-bar oxygen source, a low-resistance valve is opened
at the beginning of inspiration and a 12 L/min
peak flow through the
mouthpiece is rapidly achieved. The delivery time of the nebulizer
must always be set to less than the patient's inspiratory time to maximize driving of the pulse into the airways. Thus, the number and duration of pulses determine the available mass of drug.
The two nebulizers used in the study are triggered in different
ways. The nebulizer used alone in the control session (Optineb; Air
Liquide Santé, Paris, France) is actuated by the inspiratory flow of the
patient (control session), whereas the nebulizer used with PS in the PS
session (Optiplus; Air Liquide Santé) is actuated by the positive pressure generated by the PS device. Great care was taken to obtain similar performances from the two nebulizers to ensure that any differences between the two sessions would be ascribable only to the
presence or absence of PS. The mass median aerodynamic diameter
(MMAD) and geometric standard deviation (
g) of the generated
aerosol, measured with a laser device (MS 1000; Malvern Instruments
Ltd, Worcs, UK) at an inspiratory time of 400 ms, were 3.21 ± 0.13 (mean ± SD) µm for the nebulizer alone and 3.16 ± 0.02 µm for the
nebulizer coupled to the PS device.
Experimental setup. The output of the nebulizer was driven through a conventional mouthpiece and a glass cranked tube (internal diameter = 18 mm; length of the horizontal and vertical components = 13 cm and 19 cm, respectively) that mimicked the anatomy of the upper airways to the site of the carina. An absolute filter (Type A/E Glass; Pall Gelman Laboratories, Ann Arbor, MI) was placed at the end of the cranked tube.
The nebulizer solution used for the in vitro study was a 2.5% KCl solution. The MMAD of the KCl solution was similar to that of the radiolabeled solution used in the clinical study. Indeed, the MMAD of the KCl solution ranged from 3 µ to 5 µ (i.e., large enough to avoid Brownian oscillations and small enough to avoid impaction in the upper airways). The mass of KCl deposited on the absolute filter was measured using optical spectrometry.
To simulate repeated inspiratory efforts, the PS device was connected to a two-chamber test lung (dual adult TTL 1600 training test lung; Michigan Instruments Inc., Grand Rapids, MI) as previously described (12). One chamber of the test lung (the driving chamber) was connected to and powered by a ventilator (César; Taema, Antony, France), whereas the other chamber (the pressurized chamber) was connected to the PS ventilator being tested. The two chambers were connected to each other by a small metal component that allowed the driving chamber to lift the PS-pressurized chamber, mimicking the patient's contribution to inspiration. Thus, generation of positive pressure in the driving chamber by the ventilator lowered the pressure in the PS-pressurized chamber to atmospheric levels, simulating the production of negative alveolar pressures in vivo when the inspiratory muscles contract. This effect was detected by the triggering system of the PS ventilator being tested. Because the metal component was not secured to the PS-pressured chamber, this chamber, once effectively pressurized, could rise above the driving chamber. Compliance was set at 50 ml/cm H2O for both chambers. Positive end-expiratory pressure (PEEP) was applied to the driving chamber at a level that ensured synchronism of motion of the two chambers at onset of inspiration. A resistance of 17 cm H2O at 1 L/s was used to connect the PS device being tested to the lung model. The ventilator was set to provide a respiratory frequency of 20 cycles/min, a TI of 1 s, and a constant inspiratory flow rate of 10 L/min.
Various PS device settings were evaluated. Inspiratory pressure
ranged from 8 to 10 cm H2O. The inspiratory trigger was set at
0.7 cm
H2O, the lowest value not associated with self-triggering in pediatric
clinical practice. The rate of initial inspiratory flow, or rate of PS insufflation, was set at the highest possible value. Exhalation was totally
passive, and began when the inspiratory flow fell below a threshold
value of about 30% of the peak inspiratory flow. The PS device and
nebulizer settings optimal for the clinical study were determined during this experimental study.
Clinical Study
Apparatus tested. The two nebulizer systems and the home PS ventilator used in the clinical study were the same as those used in the in vitro study.
Patients. Eighteen clinically stable CF patients (seven girls and 11 boys) aged 12 ± 4 (mean ± SD) yr (range: 6 to 21 yr) were studied. The subject's mean Shwachmann score was 74 ± 16 (range: 40 to 100) (13). Their VC and FEV1 were 77 ± 21% and 72 ± 26% predicted, respectively. Four patients were colonized with Pseudomonas aeruginosa and one with Burkholderia cepacia. Inhaled bronchodilators and corticosteroids were used on a long-term basis by eight and nine patients, respectively. Three patients were taking recombinant human deoxyribonuclease. None of the patients changed their usual treatment during the study period. All patients were receiving follow-up at our CF clinic, and were selected on the basis of disease stability, willingness to participate in the study, and availability for spending two afternoons in the hospital for the study.
The study protocol was approved by our institutional review board. Informed consent was given by all patients and their parents.
Imaging protocol. Each subject underwent one control session and one PS session. The two lung scans were performed in random order, during the afternoon, within the same week but at least 3 d apart. Before starting the protocol, the patient and parents were given information on the breath-actuated nebulizers and PS device, and the patient went through a practice session. A perfusion scan was performed after either the control or the PS session.
An anhydrous sodium solution containing 185 MBq of 99mTc-phytates (Phytacis; CIS Bio International, Gif-sur-Yvette, France), diluted in 4 ml of normal saline, was placed in a lead-shielded unit in the nebulizer (Sidestream; Medic-Aid, Sussex, UK). Before and after nebulization, radioactivity in the nebulizer was counted with a gamma camera (single-headed DS7; Sopha Medical Vision, Buc, France), with the same counting geometry for both counts. The exact amount of nebulized radioactivity (in MBq) was calculated as the count difference divided by the initial activity in the unit. Within the range of radioactivities used in this study, we found a linear relationship between actual radioactivity (x = 30 to 300 MBq with the same counting geometry) and activity measured by the camera (y = 124,415 x; R2 = 0.9994).
Patients inhaled the aerosol solution through a mouthpiece while in the sitting position and wearing a nose-clip. Dynamic aerosol deposition within the lungs was assessed on a posterior view during continuous breathing of VT for 10 min. Ten consecutive 60-s frames were thus acquired in 64 × 64-pixel matrices. Patient movements were prevented by having the patient sit in a fixed seat with the patient's back to the camera, and by maintaining the patient's head and the nebulizer in a sagittal median plane orthogonal to the collimator. Patients were asked to relax, and no attempt was made to control the breathing pattern. Immediately after each session, a gargle was provided for decontamination. After completion of the nebulization, additional, 300-s static images in the anterior and posterior projections were acquired in 256 × 256-pixel matrices.
For practical reasons (essentially decontamination purposes), we were unable to measure patient breathing patterns during the two nebulizer sessions.
An intravenous injection of 99mTc-macroaggregates in an amount providing 18.5 MBq of radioactivity was given via a small peripheral catheter, which was rinsed after the injection. The exact amount of injected radiotracer was determined by measuring syringe radioactivity before and after the injection. Perfusion image acquisition was begun immediately after the injection, with the patient seated, and with the same gamma-camera being used to obtain two 120-s static views (anterior and posterior) in 256 × 256-pixel matrices. Perfusion images were obtained for two reasons: (1) to allow an accurate determination of lung outlines; and (2) to determine an individual counting correction (ICC). This information was then used to calculate the absolute aerosol deposition within the lungs (14, 15).
Dynamic studies were analyzed by plotting the total count within each of the 10 frames against the corresponding acquisition time to produce a time-activity curve. Posterior views were analyzed.
Quantitative aerosol deposition was then assessed. Lung edges were determined automatically on the perfusion images by thresholding at 15% of the maximal count, which in earlier studies (6, 16) gave the best separation between pulmonary activity and background. The lung-edge outlines were then applied to the ventilation scans and were used to define further regions of interest (ROI) and for background subtraction. Each lung was manually included within an elastic grid composed of 40 equal-sized cells, in a five-column and eight-row matrix (Figure 1). This matrix allowed separation of the lung into central, middle, and peripheral regions, as previously described (16). Deposition in the central and peripheral regions was analyzed, the middle region being less informative in patients with obstructive lung disease. Homologous regions in the right and left lungs were used to evaluate the homogeneity of aerosol distribution by calculating the four first-order moments (i.e., mean count per pixel, SD, skew, and kurtosis [6, 17]). These parameters are indices that are corrected for the actual number of pixels, the mean pixel count, and the histogram SD (6, 17). Because these parameters are dimensionless, they allow the comparison of distribution histogram shapes regardless of lung size and total deposited radioactivity. Skew is a measure of asymmetry, and is low when distribution is uniform or symmetric. Kurtosis is considered a measure of the sharpness of distribution, and approaches zero when deposition is peripheral or diffuse. A penetration index was also calculated, as the ratio of peripheral to central counts (16). In addition, each lung was divided into three rectangular ROIs of equal size, consisting of the upper, middle, and lower thirds of the lung, to determine aerosol distribution from the base to the apex within each lung (expressed as the percentage of total lung deposition) (Figure 2) (17). Because of asymmetric pulmonary function caused by heart imprinting, the geometric means on the anterior and posterior views were calculated for both the mean count per pixel and for the upper-, middle- and lower-third percentages, and were then used to approximate these parameters in the median coronal plane. Absolute pulmonary deposition was assessed by multiplying these last parameters by the ICC, calculated as follows:
|
|
|
Statistical Analysis
Dynamic aerosol deposition was analyzed by subjecting the time- activity curves to linear regression. All continuous variables (including functional lung volumes, clinical scores, slopes of dynamic uptake, relative and quantitative count statistics, and uptake homogeneity) were compared for the control and PS sessions for each patient, using a paired Student's t test, and the null hypothesis was rejected if the value of p was less than 0.05. Additionally, patients were separated into two subgroups based on clinical or scintigraphic status, and differences between these subgroups were evaluated with an unpaired t test.
| |
RESULTS |
|---|
|
|
|---|
Experimental In Vitro Study
The experimental in vitro study analyzed interactions between the nebulizer and the PS ventilator to determine whether the nebulizer changed the functioning of the PS ventilator or vice versa. We found that when the nebulizer was coupled to the PS ventilator, the generated pressure remained unchanged. Setting the inspiratory time of the nebulizer above 400 ms resulted in premature discontinuation of PS due to the additional flow delivered by the nebulizer. We further found a linear relationship, with both nebulizer systems, between the mass of KCl deposited on the absolute filter and the number of inhalations (Figure 3). The mass of KCl on the absolute filter was larger with the PS system, but the difference did not reach statistical significance.
|
The yield of each system was calculated as the ratio of the amount of KCl that was nebulized for a fixed number of inhalations and the amount of KCl that was deposited on the absolute filter placed at the end of the cranked tube. This ratio remained stable with both systems from 50 to 400 inhalations, and was 70% and 75% for the control and PS session, respectively.
These data indicate that coupling of the nebulizer to the PS device in vitro did not impair the functioning of the nebulizer or PS device provided the nebulization time remained below 400 ms.
Clinical Study
All 18 patients completed the clinical study. The PS sessions were well tolerated, without side effects. Radioactivity in the nebulizer was similar before the two sessions (174 ± 8 [mean ± SD] MBq and 171 ± 10 MBq before the control and PS sessions, respectively, p = NS), whereas the amount of radioactivity that left the nebulizer during the nebulization time was slightly greater during the PS session (29 ± 11 MBq) than during the control session (25 ± 6 MBq; p = NS).
Dynamic lung deposition. Evaluation of the dynamic lung scans performed during aerosol inhalation showed a linear relationship between nebulization time and aerosol deposition in all patients (Figure 4). Dynamic aerosol deposition was significantly increased during the PS session as compared with the control session (average slopes: 0.43 ± 0.07 MBq/min and 0.32 ± 0.23 MBq/min for the PS and control sessions, respectively; p = 0.018).
|
Total aerosol deposition in the lungs. The absolute radioactivity deposited in the lungs after the nebulization session was about 30% greater after the PS session than after the control session (Table 1). Aerosol lung deposition was also significantly greater after the PS session when deposition was expressed as a percentage of the nebulizer output, which was the difference between the original radioactivity of the solution placed in the nebulizer unit and the residual radioactivity remaining after nebulization, or as the percentage of the total radioactivity present in the nebulizer unit.
|
Regional aerosol deposition. After both the control and PS sessions, approximately one-third of aerosol deposition in the lungs occurred in the central area (Table 1). The penetration index was 2.3 ± 0.5, and was remarkably similar for the two sessions. No difference in lung aerosol deposition between the two sessions was observed when deposition was analyzed from base to apex (Table 1).
Homogeneity of aerosol deposition. Skew and kurtosis were similar results after the control and PS sessions (Table 1).
Correlation of aerosol lung deposition with clinical characteristics of the patients. The group of 13 patients with larger amounts of deposited aerosol after the PS session ("responders") did not differ, in terms of clinical characteristics (age, height, Shwachmann score) or lung function parameters, from the group of five patients with unchanged amounts of deposited aerosol ("nonresponders"). However, the nonresponders to PS tended to have poorer lung function than the responders (FEV1 = 57 ± 33% and 77 ± 22% in the nonresponders and responders, respectively; p = 0.1).
When the patients were separated into two groups based on disease severity (mild, FEV1 > 60% predicted; or moderately severe: FEV1 < 60% predicted), the only parameter that differed significantly in the two groups was skew, with the difference being apparent after both the control and the PS session. Skew was 1.3 ± 0.4 and 0.9 ± 0.3 in the mild and moderately severe groups, respectively, after the control session (p = 0.02), and was 1.3 ± 0.4 and 0.9 ± 0.5, respectively, after the PS session (p = 0.04). Dynamic lung deposition, total and regional aerosol deposition, kurtosis, and penetration index showed no significant differences between the two groups.
In the overall study group (n = 18), total aerosol deposition was not correlated with age, height, or the percent predicted VC or FEV1. Neither was the time-course of aerosol deposition within the lungs correlated with the percent of predicted VC or FEV1. Skew, which reflects asymmetry and heterogeneity of aerosol deposition, was not correlated with age, but was inversely correlated with the percent predicted FEV1 after both the control and PS sessions (p = 0.0025 and p < 0.0001, respectively), indicating that poorer lung function, as assessed from FEV1, was associated with more asymmetric and more heterogeneous aerosol distribution. Skew was inversely correlated with the percent predicted VC after the control session (p = 0.003) but not after the PS session. There was also an inverse relationship between skew and the Shwachmann score, which was significant after the control session (p = 0.008) and of borderline significance after the PS session (p = 0.06). Skew was inversely correlated with the penetration index after both sessions (p = 0.05 and p = 0.02 after the control and PS sessions, respectively). Neither kurtosis nor the penetration index were correlated with any of the clinical or functional parameters studied.
| |
DISCUSSION |
|---|
|
|
|---|
The present study was the first in which both a bench test and a clinical study were used to evaluate the effect of PS ventilation on aerosol deposition in children with CF. Our results show that total lung aerosol deposition can be significantly enhanced when aerosol generation and transport through the airways are driven by noninvasive PS ventilation.
The major finding in our clinical study was that the total radioactivity count deposited in the lungs after nebulization, expressed as a percentage of the total radioactivity count initially present in the nebulizer unit, increased by about 30% after the PS session as compared with the control session.
The first issue is whether this increased aerosol deposition was related to a difference between the nebulizers used in the control and PS sessions. Two characteristics of aerosols have been shown to affect lung deposition of aerosol particles. One is particle size distribution with smaller particles more likely to deposit within the lungs than larger particles (4, 7). Particle size distributions in the aerosols generated by the two nebulizers used in our study were closely similar and were near the optimal values for aerosol deposition in children with CF (4). The other factor affecting aerosol particle deposition in the lungs is the intrinsic performance of the nebulizer. Better intrinsic performance of the PS-coupled nebulizer in our study would translate into larger values for both the volume of solution and the amount of radioactivity that leave the nebulizer unit during the nebulization time, as compared with the control nebulizer. Although we did find that the amount of radioactivity that left the nebulizer was slightly larger during the PS session than during the control session, this difference did not reach statistical significance, and was too small to explain the 37% increase in aerosol lung deposition during the PS session. These data indicate that the two nebulizers had similar performances and produced similar aerosols. Consequently, the increase in aerosol deposition in the lungs after the PS session can be ascribed to the use of PS.
The most likely explanation for the increased aerosol deposition after the PS session is the change in breathing pattern
change induced by PS ventilation. In a previous study, we
showed that children with CF had an increase in their VT
(from 0.4 L to 1.0 L) and decrease in their RR (from 18 breaths/min to 11 breaths/min) when treated with noninvasive
PS ventilation set at an inspiratory pressure of 12 cm H2O
(11). The magnitude of these changes in breathing pattern varied with the level of PS: higher levels of PS ventilation resulted
in a higher VT and a lower RR. In the present study, we chose
a moderate level of inspiratory pressure (10 cm H2O) to avoid
an excessive increase in inspiratory flow, which would have increased the risk of aerosol particle impaction in the upper and
central airways. The velocity of inspired gas is known to be an
important determinant of particle transport and deposition.
Higher inspiratory flow rates increase turbulent flow and produce stronger inertial forces, leading to impaction of particles
in more proximal airways (6). For practical reasons (essentially decontamination), we were unable to study inspiratory flow and VT during the two nebulization sessions. In our previous study, however, the increase in VT (
150%) exceeded the
increase in inspiratory flow (
50%) during PS ventilation set
at 12 cm H2O (11). This change in breathing pattern during PS
can be expected to minimize particle impaction in the proximal airways. The optimal breathing pattern for drug delivery
combines a low inspiratory flow rate and large tidal breaths.
During slow, relaxed breathing, particle sedimentation and
diffusion are the primary deposition processes, and particle
deposition increases in the peripheral airways as a result of the
proximity of lung surfaces in these airways. It has been shown
that optimal particle deposition in patients with airway obstruction is achieved with flow rates of no more than 0.2 L/s
(1). The flow rate providing optimal particle deposition in
children with CF is unknown, but we found that "targeting"
breath at a predetermined low flow rate by instructing the patient to inhale in such way that the flow rate was kept between
a pair of "tramtracks" on a screen enhanced amikacin deposition in the lungs (9). The most often reported advantage of PS
ventilation is control of the frequency and duration of inspiratory assistance by the patients themselves. However, because
expiration is unassisted, the decrease in respiratory rate induced by PS ventilation results in an increase in expiratory
time. In patients with airway obstruction, increasing the expiratory time may result in less alveolar gas retention and consequently in a decrease in the overall work of breathing. A
longer expiratory time may also have an important effect on
aerosol droplet deposition, facilitating sedimentation of particles that have entered the lung during inspiration (18).
An important message from this study is that coupling aerosol delivery with PS ventilation can enhance aerosol deposition in the lungs without increasing particle impaction in the proximal airways. One of our concerns was that an excessively high PS ventilation level might generate high axial velocities and higher levels of flow turbulence, thus increasing proximal particle impaction, with radial diffusion from turbulence minimizing the axial dispersion of aerosol. It is interesting to note that the moderate levels of inspiratory pressure (10 cm H2O) used in the present study provided a meaningful increase in aerosol deposition without altering the aerosol distribution, at least in our group of CF patients. Analysis of local lung deposition of aerosol is important in a disease such as CF (6). We therefore calculated the skew and the kurtosis of aerosol distribution. Skew and kurtosis in adults with CF have been shown to be significantly increased, indicating nonuniform distribution (6). In CF, bronchoconstriction or airway obstruction caused by mucus or inflammation diverts more of the airflow toward the nonobstructed airways. This increases particle deposition in healthy regions as a result of both higher flow rates, which facilitate particle impaction in airways, and longer residence times in the lung periphery, which favor sedimentation and diffusion. Furthermore, deposition immediately proximal to obstructed sites is also increased despite the reduced airflow through these regions, whereas deposition downstream from obstructed sites is severely reduced. At sites of mild to severe obstruction, high axial velocities and higher levels of flow turbulence seem to enhance particle deposition by impaction and by turbulent radial diffusion, resulting in concentration of aerosol at "hot spots" immediately downstream from the obstructed sites.
Skew and kurtosis have been found effective in separating normal subjects from CF patients on the basis of aerosol deposition (6). However, these indices of homogeneity of aerosol deposition provide no information about which anatomic sites receive less aerosol. We found an inverse correlation between skew and severity of lung disease as assessed by the Shwachmann score and by the percent of predicted FEV1. We also found a significant inverse correlation between skew and the penetration index, indicating that as the lung disease in CF progresses, aerosol deposition becomes both more central and less homogeneous. In contrast to others, we found no correlation between the penetration index and indices of bronchial obstruction such as FEV1 (1). This may be because both of the breath-actuated nebulizers used in our study are designed to reduce asynchronism between aerosol delivery and inspiration by the patient.
It would have been of interest to identify a clinical or functional parameter capable of predicting the effect of PS on aerosol deposition in individual patients. Unfortunately, no correlation was found between enhanced aerosol deposition and age, height, the Shwachmann score, VC, or FEV1. Furthermore, the addition of PS did not increase the amount of aerosol that was deposited in the lungs in five of the 18 children in our study. This troubling result can be explained by several factors. First, our population was small and heterogeneous. Some children had very severe lung disease whereas others had only moderate lung impairment. Second, not all the children had routine nebulization therapy at home. Despite the training session, some children were more familiar with the PS technique than were others. Third, the variability and unpredictability of aerosol deposition in healthy humans has been pointed out (19). This is likely to be even more true in CF, a disease in which the degree of bronchial obstruction is highly variable from one day to another (Figure 5). Fourth, we required a washout period of at least 3 d between the control and PS sessions because of the decay of the radioactivity in the radiolabeled aerosol. Despite careful exclusion of those children who were not in a stable condition, it is possible that the clinical condition of some patients was slightly different on the two sessions. Moreover, accurate analysis of the distribution of small amounts of aerosol in diseased lungs is even more difficult in children than in adults. Despite these factors, we observed a general tendency toward enhanced aerosol deposition with PS ventilation in patients with better lung function, but this may merely reflect the limited efficacy of aerosol transport in severely diseased lungs even with optimal nebulization techniques. Others also found no correlation between total lung aerosol deposition and age or height (3); however, in one study, aerosol concentration within the lungs (defined as the percentage of nebulizer output deposited normalized for the predicted TLC) was higher in children and teenagers than in adults, and in adults was negatively correlated with height and weight (20).
|
Our study provides convincing evidence that an in vitro model incorporating particle diameter, normal breathing frequency, and VT is less accurate than in vivo measurement of aerosol deposition, which takes into account all the factors that determine aerosol deposition, including the disease being treated. However, an important finding in our experimental study was that coupling the nebulizer to the PS ventilator did not alter the performances of the nebulizer or PS ventilator.
This study was a feasibility study that assessed the possibility of combining nebulization and noninvasive PS ventilation in children with CF. It was not an efficacy study evaluating the therapeutic potential of a drug (a bronchodilator or an antibiotic), but rather a deposition study. It seems important to us to know, before the routine use of a nebulized drug, the site of its deposition (especially in diseased lung such as in CF), and the exact amount of drug that is deposited in the lungs. These essential points are most often lacking in studies of the effect of a nebulized drug in children.
Future treatments for CF patients will require targeted delivery to specific cell types or specific receptor populations. It
is therefore critical to identify and to learn how to control the
sites of aerosol deposition. For example, similar total deposition values can mask wide differences in the regional distribution of an aerosol (7). Our study can be viewed as a first attempt to manage drug distribution with assisted ventilation.
Methods of assisted ventilation other than PS need to be investigated. Several of these methods have been used in an attempt to enhance aerosol deposition. The effects of continuous positive airway pressure (CPAP) at 10 cm H2O on aerosol
kinetics and bronchodilator efficacy have been evaluated in
nine stable asthmatic adults (21). Despite a significant reduction in total aerosol delivery (from 6.85 ± 1.52% to 1.32 ± 0.37%) when a facemask was used, the bronchodilator response was not affected, probably because of an intrinsic effect of CPAP itself (21). Another study with adults attending
an emergency room for acute asthma exacerbations showed
that the delivery of a
-agonist by noninvasive inspiratory PS
ventilation with PEEP resulted in significantly greater improvement in the peak expiratory flow rate than did delivery of the
drug by nebulization alone (22). Two other studies found that
intermittent positive-pressure breathing and high-frequency
oscillation did not increase drug delivery to the lungs (23, 24).
In conclusion, this study is the first to show that coupling PS ventilation to a breath-actuated nebulizer enhances aerosol deposition in the lungs of children with CF. This finding may have three clinical applications. First, in patients with CF in whom noninvasive mechanical ventilation is indicated, simultaneous administration of aerosol therapy and PS ventilation can enhance aerosol deposition in the lung, thereby helping to unload respiratory muscles, improve oxygenation, and reduce the time spent in treatment each day, possibly resulting in better compliance with treatment. Second, coupling PS to nebulization may be effective in other lung diseases known to respond to nebulized treatments: in status asthmaticus, for instance, coupling bronchodilator nebulization to noninvasive PS may increase treatment efficacy. Clearly this may require a functional study, which could be the logical continuum of a deposition study as that described here. Third, we suggest that nebulization of a radiolabeled aerosol may be an appropriate means of tailoring the amount of drug deposited in the lungs to the needs of each individual CF patient.
| |
Footnotes |
|---|
Correspondence and requests for reprints should be addressed to Brigitte Fauroux, M.D., Hôpital d'Enfants Armand Trousseau, 28 Avenue du Docteur Arnold Netter, 75012 Paris, France. E-mail: bfauroux{at}aol.com
(Received in original form March 13, 2000 and in revised form July 14, 2000).
Acknowledgments: The authors are deeply grateful to the CF patients and their parents who participated in this study with unfailing enthusiasm.
Supported in part by Air Liquide Santé International and Mallinckrodt Chemical Co.
| |
References |
|---|
|
|
|---|
1. Ilowite JS, Gorvoy JD, Smaldone GC. Quantitative deposition of aerolized gentamycin in cystic fibrosis. Am Rev Respir Dis 1987; 136: 1445-1449 [Medline].
2.
Thomas SHL,
O'Doherty MJ,
Graham A,
Page CJ,
Blower P,
Geddes DM,
Nunan TO.
Pulmonary deposition of nebulised amiloride in cystic fibosis: comparison of two nebulisers.
Thorax
1991;
46:
717-721
3. Chua HL, Collis GG, Newbury AM, Chan K, Bower GD, Sly PD, Le Souef PN. The influence of age on aerosol deposition in children with cystic fibrosis. Eur Respir J 1994; 7: 2185-2191 [Abstract].
4. Mallol J, Rattray S, Walker G, Cook D, Robertson CF. Aerosol deposition in infants with cystic fibrosis. Pediatr Pulmonol 1996; 21: 276-281 [Medline].
5. Alderson PO, Secker-Walker RH, Strominger DB, Markham J, Hill RL. Pulmonary deposition of aerosols in children with cystic fibrosis. J Pediatr 1974; 84: 479-484 [Medline].
6.
Laube BL,
Links JM,
LaFrance ND,
Wagner HN,
Rosenstein BJ.
Homogeneity of bronchopulmonary distribution of 99mTc aerosol in normal subjects and in cystic fibrosis patients.
Chest
1989;
95:
822-830
7.
Newman SP,
Woodman G,
Clarke SW.
Deposition of carbenicillin aerosols in cystic fibrosis: effects of nebuliser system and breathing pattern.
Thorax
1988;
43:
318-322
8.
Hess D,
Fisher D,
Willaims P,
Pooler S,
Kacmarek RM.
Medication nebulizer performance. Effects of diluent volume, nebulizer flow, and
nebulizer brand.
Chest
1996;
110:
498-505
9. Faurisson F, Prévot B, Just J, Fauroux B, Chevret S, Grimfeld A. Respiratory self-monitoring of inhalation increases the deep deposition of aerolized amikacin in cystic fibrosis patients [abstract]. Am Rev Respir Dis 1994; 149: A666 .
10. Brochard L, Harf A, Lorino H, Lemaire F. Inspiratory pressure support prevents diaphragmatic fatigue during weaning from mechanical ventilation. Am Rev Respir Dis 1989; 139: 513-521 [Medline].
11. Fauroux B, Boulé M, Lofaso F, Zérah F, Clément A, Harf A, Isabey D. Chest physiotherapy in cystic fibrosis: improved tolerance with nasal pressure support ventilation. Pediatrics 1999; 103: e93 .
12. Lofaso F, Brochard L, Hang T, Lorino H, Harf A, Isabey D. Home versus intensive care pressure support devices: experimental and clinical comparison. Am J Respir Crit Care Med 1996; 153: 1591-1599 [Abstract].
13. Shwachmann H, Kulczucki LL. Long-term study of one hundred and five patients with cystic fibrosis. Am J Dis Child 1968; 96: 6-15 .
14. Iacono AT, Smaldone GC, Keenan RJ, Diot P, Dauber JH, Zeevi A, Burckart GJ, Griffith BP. Dose-related reversal of acute rejection by aerosolized cyclosporine. Am J Respir Crit Care Med 1997; 155: 1690-1698 [Abstract].
15. Smaldone GC, Fuhrer J, Steigbigel RT, McPeck M. Factors determining pulmonary deposition of aerosolized pentamidine in patients with human immunodeficiency virus infection. Am Rev Respir Dis 1991; 143: 727-737 [Medline].
16. Agnew JF, Francis RA, Pavis D, Clarke SW. Quantitative comparison of 99Tcm-aerosol and 81Krm ventilation images. Clin Phys Physiol Meas 1982; 3: 21-30 [Medline].
17. Garrard CS, Gerrity TR, Schreiner JF, Yeates DB. Analysis of aerosol deposition in the healthy human lung. Arch Environ Health 1981; 39: 184-193 .
18. Newman SP, Pavia D, Garland N, Clarke SW. Effects of various inhalation modes on the deposition of radioactive pressurized aerosols. Eur J Respir Dis 1982;63(Suppl)119:57-65.
19. Thomas SHL, Langford JA, George RJD, Geddes DM. Variability in the measurement of nebulized aerosol deposition in man. Clin Sci 1988; 75: 535-542 [Medline].
20.
O'Doherty MJ,
Thomas SHL,
Gibb D,
Page CJ,
Harrington C,
Duggan C,
Nunan TO,
Bateman NT.
Lung deposition of nebulised pentamidine in children.
Thorax
1993;
48:
220-226
21. Parkes SN, Bersten AD. Aerosol kinetics and bronchodilator efficacy during continuous positive airway pressure delivered by face mask. Thorax 1997; 52: 171-175 [Abstract].
22.
Pollack CV,
Fleish KB,
Dowsey K.
Treatment of acute bronchospasm
with
-adrenergic agonist aerosols delivered by a nasal bilevel positive
airway pressure circuit.
Ann Emerg Med
1995;
26:
552-557
[Medline].
23. Dolovich MB, Killian D, Wolff RK, Obminski G, Newhouse MT. Pulmonary aerosol deposition in chronic bronchitis: intermittent positive pressure breathing versus quiet breathing. Am Rev Respir Dis 1997; 115: 397-402 .
24. Thomas SH, Langford JA, George RJ, Geddes DM. Aerosol deposition in the human lung: effect of high-frequency oscillation on the deposition characteristics of an inhaled nebulized aerosol. Clin Sci 1988; 75: 535-542 .
This article has been cited by other articles:
![]() |
M. Griese, J. Ramakers, A. Krasselt, V. Starosta, S. van Koningsbruggen, R. Fischer, F. Ratjen, B. Mullinger, R. M. Huber, K. Maier, et al. Improvement of Alveolar Glutathione and Lung Function but Not Oxidative State in Cystic Fibrosis Am. J. Respir. Crit. Care Med., April 1, 2004; 169(7): 822 - 828. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J. TOBIN Pediatrics, Surfactant, and Cystic Fibrosis in AJRCCM 2000 Am. J. Respir. Crit. Care Med., November 1, 2001; 164(9): 1581 - 1594. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Proc. Am. Thorac. Soc. | Am. J. Respir. Cell Mol. Biol. |