A Bioassay Using Methacholine |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| |
ABSTRACT |
|---|
|
|
|---|
Because chlorofluorocarbons (CFCs) contribute to depletion of stratospheric ozone, CFC-containing metered-dose inhalers (MDIs) such as Ventolin and Proventil are being phased out of production. In terms of delivery of albuterol to the lungs, we compared an alternative delivery system, the Spiros dry-powder inhaler (DPI), with Ventolin, using a methacholine challenge-based clinical bioassay. Twenty-four adults and adolescents with asthma completed this double-blind, four-period crossover study. Doses evaluated were one and three actuations each of Spiros and Ventolin (90- and 270-µg albuterol base). A methacholine challenge (Cockcroft method) was initiated 3 h before and 0.25 h after albuterol. Predose PC20FEV1 values were not significantly different between study days. Postdose PC20FEV1 results met standard bioassay study validity criteria: i.e., a significant dose-response relationship was present (p = 0.0002); tests for deviation from parallelism and overlap of dose-response curves were nonsignificant (p = 0.08, 0.69). By using Finney 2-by-2 bioassay analysis, we estimate that each Spiros actuation delivers 1.12 times as much albuterol to the airways as one Ventolin actuation (90% confidence interval, 0.68 to 1.94). There were no significant differences in markers of systemic effects (vital signs, potassium, and blood glucose concentrations). We conclude that Spiros and Ventolin inhalers deliver comparable quantities of albuterol to the airways.
| |
INTRODUCTION |
|---|
|
|
|---|
Inhaled
-adrenergic drugs play an extremely important role
in the management of asthma (1). In the United States as well as in much of the rest of the world, inhaled albuterol is the most commonly prescribed drug in this class (2). Metered-dose inhaler (MDI) formulations of albuterol marketed in the
United States all employ chlorofluorocarbon (CFC) compounds as propellants (Ventolin, Proventil, and at the time of
this writing, four approved generic equivalents to Ventolin or
Proventil sold under 17 brand names). With each actuation of
the inhaler, expansion of the CFC propellant provides the energy needed to aerosolize the drug (3).
Release of CFCs into the atmosphere has been shown to cause degradation of stratospheric ozone concentrations which, in turn, allows increased penetration of ultraviolet radiation to the surface of the earth. Adverse effects on the environment and human health are associated with this increase (4). Consequently, by international agreement (5), CFC production and use are being phased out. While use of CFCs in MDIs is currently permitted under a temporary "essential use exemption," it is clear that these CFC-containing products will eventually be phased out as well.
This has led to diligent efforts by the pharmaceutical industry to develop alternative aerosol delivery systems that do not use CFCs (3, 6, 7). Proventil HFA, developed by 3M Health Care (St. Paul, MN) and marketed in the United States by Schering-Plough Pharmaceuticals (Madison, NJ), is the first non-CFC metered-dose inhaler approved in the United States (8, 9). This system uses a more environmentally friendly propellant, hydrofluoroalkane (HFA 134a), to deliver albuterol. Another strategy has been to develop sophisticated, multidose, dry powder inhaler (DPI) systems that derive the energy necessary to aerosolize drugs for inhalation from sources other than expansion of propellant (typically from the patient's inspiratory effort) (10, 11).
With any non-CFC inhaled albuterol formulation intended
to serve as a replacement for CFC-containing inhalers, it is
critically important that clinicians know whether the new inhaler delivers essentially the same, more, or less drug to the
relevant
-adrenergic receptors in the airways as do marketed
CFC formulations. If a replacement inhaler delivers significantly less drug to the airways, patients may experience insufficient relief of symptoms when they use the inhaler. If
the new formulation delivers significantly more drug, adverse
events associated with excessive doses of inhaled
-agonist use
may be enhanced (12).
It has been usual practice to study
2 agonist-induced bronchodilation to obtain this information. However, in the typical study of subjects with resting asthmatic airway obstruction, it is frequently not possible to distinguish between the effects of
different doses of the reference product (i.e., a significant dose-response relationship is not present). This usually appears to be because one actuation of the reference product is
near the top of the dose-response curve. Measurement of
drug-induced changes in nonspecific airway responsiveness
appears to have additional advantages over studies of bronchodilation (13). Measurement of airway responsiveness
with bronchial provocation with histamine or methacholine is
highly reproducible (16) and can reflect broad changes in inhaled
2-agonist activity in the lung under conditions in which
bronchodilation has "plateaued" near maximal achievable levels (13, 14). Information obtained from bronchoprovocation studies appears to be at least as clinically relevant as information obtained from studies of bronchodilation, because airway
responsiveness to histamine and methacholine itself correlates
with exercise-induced bronchospasm (17), cold air-induced
bronchospasm (18), overnight drop in peak flows (19), medication requirements (19, 20), and severity of asthmatic symptoms. Loss of
-agonist protective effects against methacholine-induced bronchospasm (21), which occurs with regular
salmeterol administration, is paralleled by loss of protection
against exercise- and allergen-induced bronchospasm (22, 23).
The Spiros inhaler is an effort-assisted, battery-powered,
DPI delivery system that is breath actuated to minimize the
need for patient coordination. It contains a removable, circular, multiple-dose cassette advanced by opening and closing
a clear plastic lid. A twinblade impeller generates an aerosol
cloud for delivery to the lungs with a slow, full breath. This
study used a prototype version of the Spiros inhaler that contained identical powder path and aerosol delivery specifications, but that used a short lever for cassette advancement. We
used a methacholine challenge-based clinical bioassay of pulmonary drug delivery to compare the relative quantity of albuterol delivered to the
-receptors by the Spiros DPI and
Ventolin MDI, the international reference product. It was our
hypothesis that the two systems would deliver comparable amounts of albuterol to the airways.
| |
METHODS |
|---|
|
|
|---|
Study Design
Subjects
12 yr of age with mild to moderate asthma were recruited
for this study (1). At an initial screening visit, FEV1 and PC20 (provocative concentration of methacholine causing a 20% fall in FEV1)
were required to be
65% of predicted and
4 mg/ml, respectively,
and PC20 was required to increase by at least eightfold in response to
administration of two actuations of Ventolin. At subsequent visits,
FEV1 had to be
65% of predicted, and PC20 had to be within twofold of the screening value. Subjects were excluded if they used more
than an average of one
-agonist inhaler per month (determined on
the basis of prescription refills), had had a respiratory tract infection
within 30 d, used oral corticosteroids within 3 mo of screening, had a
history of life-threatening asthma, or had significant illness in addition
to asthma. Stable doses of theophylline, inhaled corticosteroid, oral
or inhaled
-agonist, oral antihistamines (excluding astemizole), and
intranasal corticosteroids were permitted. None of the subjects were
taking ipratropium, cromolyn, or nedocromil before or during the study.
Thirty-one nonsmoking subjects were enrolled in this double-blind, randomized, crossover study. Twenty-four subjects completed all four treatment periods (15 males, 9 females; mean age, 26.2 yr; age range, 12 to 46 yr; Table 1). Two subjects were discontinued before study completion secondary to use of a drug prohibited in the protocol (prednisone) in response to an adverse event (asthma exacerbation). Four subjects were discontinued because of failure to meet protocol-specified criteria for between-visit reproducibility of baseline FEV1 or baseline PC20 and one subject was discontinued for failure to meet protocol-specified criteria for minimal responsiveness of PC20 to albuterol administration. Data from the 24 subjects completing the trial were used in the efficacy analyses. Data from all 31 subjects randomized into the study were used in safety assessments.
|
Four treatments were studied: one and three actuations of Ventolin MDI (90 µg of albuterol base per actuation) and one and three actuations of Albuterol Spiros DPI (108 µg of albuterol sulfate, equivalent to 90 µg of albuterol base per actuation). One of these treatments was administered on each of four separate study days. Blinding to both dose and device was accomplished by use of a double-dummy technique. The protocol received institutional review board approval, and written informed consent was obtained from subjects before participation.
Methacholine challenges were initiated 3 h before (predose) and
15 min after (postdose) albuterol dosing. The protocol required that
short-acting
-agonists (e.g., albuterol) be withheld for 6 h, salmeterol
for 48 h, theophylline for 48 h, hydroxyzine for 96 h, astemizole for
3 mo, all other antihistamines for 48 h, and
blockers and nonsteroidal antiinflammatory drugs for 7 d before each methacholine challenge. Safety of each treatment was assessed by monitoring the nature
and frequency of adverse events; by measurement of pulse rate, blood
pressure, blood glucose, and serum potassium before and after each
treatment; and by performing a 12-lead electrocardiogram before and
after treatment.
Methacholine challenges were performed according to the method of Cockcroft and coworkers (19, 20). Spirometry was performed with a KOKO pneumotachograph and software (Pulmonary Data Systems Research, Louisville, CO). The software was specifically adapted to accommodate the tidal breathing (19) method of methacholine challenge. Results were expressed as provocative concentration of methacholine (mg/ml) required to produce a 20% decrease in FEV1 (PC20). For purposes of data presentation, results are also expressed as the activity ratio: Activity ratio = postdose PC20/geometric mean predose PC20 for that subject. This indicates the number of times by which the PC20 was increased by albuterol treatment. Results of statistical analysis of data expressed in these two ways (PC20 and activity ratio) are mathematically identical. Thus, only the results of statistical analysis for PC20 are presented.
Data Analysis
Results were analyzed using repeated-measures analysis of variance (ANOVA) with loge(PC20) as the dependent variable. Loge transformation of PC20 was necessary in order to meet the normality and equal variance assumption of ANOVA. Treatment, study day, and subject were factors in the ANOVA. The relative potency of the Spiros and Ventolin inhalers was calculated according to the 2-by-2 method of Finney (24). Study validity was evaluated by established criteria that test for a significant dose-response relationship (regression contrast), for the presence of a significant deviation from parallelism of the Spiros and Ventolin dose-response curves (parallelism contrast), and for the presence of a significant difference in mean responses to Spiros and Ventolin (preparations contrast). A significant preparations contrast would indicate that the Spiros and Ventolin dose- response curves did not overlap sufficiently for a valid potency estimate to be obtained. A significant parallelism contrast would indicate that curves for Spiros and Ventolin were not parallel. When this is true, relative potency is not constant, and changes as dose changes. Relative potency is expressed as the number of actuations of Ventolin required to produce an increase in PC20 equal to that produced by one actuation of Spiros DPI. For the purposes of this study, we prospectively specified that the two inhalers would be considered "comparable" if one actuation of the Spiros inhaler was at least half as potent and no more than twice as potent as one actuation of Ventolin. This requires that the 90% confidence interval for relative potency of the Spiros inhaler be contained within an interval of 0.5 to 2.0 (25). Conceptually, this is equivalent to demonstrating that the DPI is more than half as potent (p < 0.05) and less than twice as potent (p < 0.05) as Ventolin. This statistical approach, known as two, one-sided hypothesis testing, is accepted and widely used by the Food and Drug Administration (FDA) in tests of bioequivalence.
An interim analysis was carried out after the first 11 subjects had completed the protocol. The intent of the interim analysis was to permit early detection of a marked difference in potency of the Spiros and Ventolin inhalers. If the per-actuation potency of the Spiros inhaler had been markedly different from that of Ventolin, the study sponsor planned to stop this study and to modify the amount of albuterol contained in one actuation of the DPI before proceeding with further clinical evaluation. On the basis of the results of a preestablished decision algorithm, the conclusion from the interim analysis was that the study should be extended to completion (24 evaluable subjects). Final analyses were adjusted for performance of the interim analysis, using the method of Lan and Demets and the O'Brien-Fleming spending function (26, 27).
| |
RESULTS |
|---|
|
|
|---|
Geometric mean predose PC20 values were not significantly different (0.65, 0.65, 0.59, and 0.64 mg/ml for one and three actuations of Spiros, and one and three actuations of Ventolin, respectively; p = 0.798). The geometric mean postdose PC20 values for the four treatments were 7.50, 10.15, 5.55, and 12.27 mg/ml, respectively. The corresponding geometric mean activity ratios for these four treatments (which indicate the change in PC20 induced by the albuterol inhaler treatment) were 11.47, 15.62, 9.48, and 19.18, respectively. A significant dose-response relationship was present (p = 0.0002) and the parallelism and preparations contrasts were not significant (p = 0.079, 0.686; Figure 1). The estimated relative potency based on these data indicated that one actuation of the Spiros inhaler is equivalent to 1.12 actuations of Ventolin (90% confidence interval 0.68 to 1.94). Adverse event profiles were similar for the two inhalers (Table 2) and no serious events were reported. Safety assessment (pulse rate, blood pressure, serum potassium, blood glucose, and electrocardiogram) showed no clinically meaningful differences between the Spiros and Ventolin inhalers.
|
|
| |
DISCUSSION |
|---|
|
|
|---|
Results presented here provide evidence that the dose of albuterol delivered to the site of relevant
-receptors in the lung by Spiros DPI is "comparable" to that delivered by Ventolin.
The relative potency estimate indicates that each actuation of
Spiros inhaler delivers approximately the same quantity of albuterol to the vicinity of these
-receptors, commonly referred to as the "biophase," as 1.12 actuations of Ventolin.
The fact that the 90% confidence interval for this estimate,
0.68 to 1.94, is completely contained within the 0.5-to-2.0 interval indicates that the Spiros inhaler meets the definition of
comparability we set out prospectively for the purposes of this
study. Markers of systemic effects (blood pressure, pulse rate,
serum potassium, blood glucose, and electrocardiogram) indicate no important difference in systemic absorption of albuterol for the two delivery systems.
Current FDA policies draw a clear distinction between inhaled albuterol preparations being developed as generic equivalents to Ventolin or Proventil and new inhaled albuterol
delivery systems being developed to replace these CFC-containing inhalers as they are phased out of use. Ventolin and
Proventil deliver bioequivalent quantities of albuterol to the
lung (28) and all generic brands currently marketed in the
United States have been rigorously tested for in vivo bioequivalence to one or both of these "innovator" albuterol
brands.* This requires demonstration (with 90% confidence) that the generic inhaler delivers between 0.67 and 1.5 times as much drug to the biophase as Ventolin or Proventil.
In contrast, current policies do not require that rigorous standards of
"bioequivalence" to Ventolin and/or Proventil be met by new inhalers, but that they be shown to be "comparable" to the
CFC formulations. Standards for demonstration of comparability do not appear to be clearly defined.
This regulatory distinction between generic formulations and new non-CFC formulations of currently marketed drugs is not clearly made in
other countries (29). Nor is it likely to be important to clinicians and patients who simply desire that "new" inhalers not
be less effective or more risky than those they currently use.
We believe that the prospective definition of comparability
used in this study is appropriate from a clinical perspective. We chose a lower relative potency limit of 0.5, which indicates that one actuation of Spiros (with 90% confidence) should be
more than half as potent as Ventolin. In many situations in
which patients use Ventolin or Proventil MDI, one or two inhalations are sufficient to provide relief of or protect from
asthma symptoms (e.g., exercise-induced bronchospasm). In
other circumstances (e.g., when the patient awakens with nocturnal asthma), as many as five actuations may be required to
normalize lung function (30). If a new albuterol preparation
were, in truth, half as potent as Ventolin, the number of actuations required to produce clinical benefit would double. This is
not likely to present a large practical problem for patients. In
fact, two other commonly used
-agonist MDIs (metaproterenol and terbutaline) are no more than half as potent per actuation as Ventolin (15, 31) (Figure 2). On the other hand, if the new preparation, in truth, delivered one-quarter or one-third as much drug to the biophase as Ventolin, then three to four
or more times as many actuations would be needed. This situation is likely to be impractical for many patients.
|
For the present study, we chose a relative potency of 2.0 as
an upper acceptability limit, indicating that the new inhaler must deliver no more than twice as much albuterol to the airway biophase as Ventolin. This is based on concern about potential adverse effects, should too much albuterol be delivered. Concern over acute toxic effects is minimal because
inhaled albuterol is a drug with a wide margin of safety (except perhaps in sensitive patients such as the elderly). On the
other hand, there is long-standing concern over whether continued excessive use of albuterol (or other inhaled
agonists)
can degrade rather than enhance control of asthma symptoms,
or worse, increase the risk of life-threatening symptoms or
death from asthma. Existing data indicate that this risk is
greater for fenoterol than for albuterol (12). Because the marketed fenoterol preparation is approximately twice as potent as Ventolin, we reasoned that it would be undesirable for the potency of Spiros to be twice as great or more than that of Ventolin.
Implicit in the FDA criteria for demonstration of in vivo bioequivalence for generic albuterol preparations, and in the foregoing discussion of the definition of comparability for new albuterol preparations, is the presumption that comparisons between a new albuterol preparation and Ventolin should be made along the dose axis rather than the response axis of the dose-response curve. In other words, results should be expressed as a ratio of doses producing equal effects (synonymous with potency ratio or relative potency) rather than a simple comparison of the magnitude of responses observed at equal nominal doses of the two preparations. Comparisons of magnitude of response will yield different results depending on the nature and units of measurement of the specific response evaluated, characteristics of the asthmatic subjects studied, and the study design used. Consequently, albuterol preparations that appear comparable in one study may not be in another. It will be difficult or impossible to reconcile different answers concerning comparability that are obtained from different studies comparing response. In contrast, studies that estimate relative potency will all produce results that are expressed in the same units and that can be directly compared with one another even if study design and subject characteristics differ. Furthermore, statistical procedures for computing relative potency have well-established validity criteria and methods for determining confidence limits on potency ratio. These serve as an indicator of study quality (i.e., the ability to accurately distinguish differences in dose delivered). Studies comparing magnitude of response do not.
To provide relative potency information, studies must be specifically designed as clinical bioassays of the quantity of albuterol delivered, rather than as clinical trials of efficacy. This concept has been discussed in more detail elsewhere (33). In brief, each study must use a clinically relevant response measure, must give more than one dose of each of the two preparations being compared, must meet internal validity criteria (as laid out in METHODS), and finally, must have sufficient statistical power to provide clinically useful information.
Published studies provide instructive examples of both successful clinical bioassay experiments and problems that can arise when these principles are not followed. Dockhorn and coworkers (8), in a pivotal study assessing comparability of Proventil HFA to Ventolin, found that the bronchodilation responses to one and two actuations of each of these inhalers were both significantly different from responses to placebo but not from each other (hence "comparable"). Yet, in reality, this study lacks the power to clearly establish relative potencies of the two preparations. Statistical power of a bioassay experiment is a function of both the variability of responses measured (e.g., standard deviation or equivalent, commonly symbolized as "s") and the slope of the dose-response curve (commonly symbolized by statisticians as "b"). In fact, it is the ratio of these two factors, s/b, that is used in sample size computations for design of bioassay experiments (the lower the value of s/b, the fewer subjects required and the greater the power of the study) (24). For this reason, s/b is used on the x-axis for Figure 2A. When results of the study by Dockhorn and coworkers are reanalyzed as a bioassay experiment, a relatively large s/b value is obtained (Figure 2A) and the 90% confidence interval on the relative potency estimate is so wide that it is not clinically useful (Figure 2B). Results of this study violate one of the standard bioassay validity criteria in that a significant dose-response relationship is not present (regression contrast). Therefore, the potency estimate is of uncertain reliability. A study published by Kleerup and colleagues (9) also assessed comparability of Proventil HFA and Ventolin. This study has a low s/b value when reanalyzed using bioassay statistical procedures (Figure 2A). Consequently, there is a narrow confidence interval about the relative potency estimate (Figure 2B). Yet, this study violates two validity criteria: the dose-response curves for the two preparations differ significantly from being parallel (parallelism contrast), and the curves do not overlap sufficiently (preparations contrast). This makes the potency estimate from the study of uncertain reliability. Fortunately, another study of Proventil HFA, specifically designed as a bioassay experiment, has been done that meets the standard validity criteria. This study uses methacholine bronchoprovocation to demonstrate that the potency of Proventil HFA relative to Ventolin is 1.05. This estimate has a 90% confidence interval (0.78 to 1.42) that even falls within the FDA 0.67-1.50 bioequivalence range set for generic inhaled albuterol MDIs (not shown in Figure 2) (34). Similarly, the current study, and a similar evaluation of CFC-containing Proventil, have also successfully used a clinical bioassay approach to establish bioequivalence to Ventolin (28).
In summary, we used a methacholine bioassay to demonstrate that Albuterol Spiros, a new DPI, delivers an estimated
1.12 times as much drug to the lung as Ventolin (90% confidence interval, 0.68 to 1.94). We propose that for this and
other new albuterol inhaler systems to be considered "comparable" to Ventolin, they should demonstrate, in a clinical bioassay, the ability to deliver between 0.5 and 2.0 times as much
albuterol to the
-receptor biophase in the lung.
| |
Footnotes |
|---|
Correspondence and requests for reprints should be addressed to Richard C. Ahrens, M.D., Department of Pediatrics, University of Iowa, 200 Hawkins Drive, Iowa City, IA 52242-1083.
(Received in original form June 17, 1998 and in revised form March 15, 1999).
* Division of Bioequivalence. 1994. Interim Guidance for Documentation of In Vivo Bioequivalence of Albuterol Inhalation Aerosols (Metered-Dose Inhalers). Office of Generic Drugs, Center for Drug Evaluation and Research, Food and Drug Administration, Rockville, MD. (http://www.fda.gov/cder/guidance/index.htm)
Summary basis of approval for albuterol inhalation aerosol manufactured by
Norton Limited. December 28, 1995. Obtained under the United States Freedom of Information Act.
Division of Oncology and Pulmonary Drug Products. 1994. Points to Consider:
Clinical Development Programs for MDI and DPI Drug Products. Center for Drug
Evaluation and Research, Food and Drug Administration, Rockville, MD.
Acknowledgments: The authors thank Donna Reihman for helping to prepare this manuscript.
Supported by Dura Pharmaceuticals, Inc. (San Diego, CA).
| |
References |
|---|
|
|
|---|
1. National Asthma Education and Prevention Program. 1997. Expert Panel Report 2: Guidelines for the Diagnosis and Management of Asthma. NIH, Bethesda, MD. NIH Publication No. 97-4051. 31-70.
2. Top 200 Rx drugs. 1997. In Drug Topics Red Book. Medical Economics Company, Montvale, NJ. 124.
3. Atkins, P. J., N. P. Barker, and D. Mathisen. 1992. The design and development of inhalation drug delivery systems. In A. J. Hickey, ed. Pharmaceutical Inhalation Aerosol Technology. Marcel Dekker, New York. 155-185.
4. Last, J. M.. 1993. Global change: ozone depletion, greenhouse warming, and public health. Annu. Rev. Public Health 14: 115-136 [Medline].
5. Montreal protocol on substances that deplete the ozone layer; September 1987. As amended in London, June 1990; and at Copenhagen, November 1992.
6. Alexander, D. J., and S. E. Libretto. 1995. An overview of the toxicology of HFA-134a (1,1,1,2-tetrafluoroethane). Hum. Exp. Toxicol. 14: 715-720 [Medline].
7. June, D. S., R. K. Schultz, and N. C. Miller. 1994. A conceptual model for the development of pressurized metered-dose hydrofluoroalkane-based inhalation aerosols. Pharm. Technol. October:40-52.
8. Dockhorn, R., J. A. Vanden, Burgt, B. P. Ekholm, D. Donnell, and M. T. Cullen. 1995. Clinical equivalence of a novel non-chlorofluorocarbon-containing salbutamol sulfate metered-dose inhaler and a conventional chlorofluorocarbon inhaler in patients with asthma. J. Allergy Clin. Immunol. 96: 50-56 [Medline].
9.
Kleerup, E. C.,
D. P. Tashkin,
A. C. Cline, and
B. P. Ekholm.
1996.
Cumulative dose-response study of non-CFC propellant HFA 134a salbutamol sulfate metered-dose inhaler in patients with asthma.
Chest
109:
702-707
10. Williams, J., and K. A. Richards. 1997. Ease of handling and clinical efficacy of fluticasone propionate Accuhaler/Diskus inhaler compared with the Turbohaler inhaler in paediatric patients: UK Study Group. Br. J. Clin. Pract. 51: 147-153 [Medline].
11. Richards, R., and M. Saunders. 1993. Need for a comparative performance standard for dry powder inhalers. Thorax 48: 1186-1187 [Abstract].
12. Sears, M. R., and D. R. Taylor. 1994. The beta2-agonist controversy: observations, explanations and relationship to asthma epidemiology. Drug Safety 11: 258-283 .
13. Ahrens, R. C., A. C. Bonham, G. A. Maxwell, and M. M. Weinberger. 1984. A method for comparing peak intensity and duration of action of aerosolized bronchodilators using bronchoprovocation with methacholine. Am. Rev. Respir. Dis. 129: 903-906 [Medline].
14. Salome, C. M., R. E. Schoeffel, K. Yan, and A. J. Woolcock. 1983. Effect of aerosol fenoterol on the severity of bronchial hyperreactivity in patients with asthma. Thorax 38: 854-858 [Abstract].
15. Ahrens, R. C., J. B. Harris, G. Milavetz, L. Annis, and R. A. Ries. 1987. Use of bronchial provocation with histamine to compare the pharmacodynamics of inhaled albuterol and metaproterenol in asthmatic patients. J. Allergy Clin. Immunol. 79: 876-882 [Medline].
16. Juniper, E. F., P. A. Frith, C. Dunnett, D. W. Cockcroft, and F. E. Hargreave. 1978. Reproducibility and comparison of responses to inhaled histamine and methacholine. Thorax 33: 705-710 [Abstract].
17. Anderton, R. C., M. T. Cuff, P. A. Frith, D. W. Cockcroft, J. L. C. Morse, N. L. Jones, and F. E. Hargreave. 1979. Bronchial responsiveness to inhaled histamine and exercise. J. Allergy Clin. Immunol. 63: 315-320 [Medline].
18. O'Bryne, P. M., G. Ryan, M. Morris, D. McCormick, N. L. Jones, J. L. C. Morris, and F. E. Hargreave. 1982. Asthma induced by cold air and its relationship to nonspecific bronchial responsiveness to methacholine. Am. Rev. Respir. Dis. 125: 281-289 [Medline].
19. Cockcroft, D. W., D. N. Killian, J. J. A. Mellon, and F. E. Hargreave. 1977. Bronchial reactivity to inhaled histamine: a method and clinical survey. Clin. Allergy 7: 235-243 [Medline].
20. Juniper, E. F., P. A. Frith, and F. E. Hargreave. 1981. Airway responsiveness to histamine and methacholine: relationship of minimum treatment to control symptoms of asthma. Thorax 36: 575-579 [Abstract].
21.
Cheung, D.,
M. D. Timmers,
A. H. Zwinderman,
E. H. Bel,
J. H. Dijkman, and
P. J. Sterk.
1992.
Long-term effects of a long-acting,
2-adrenoceptor agonist, salmeterol, on airway hyperresponsiveness in
patients with mild asthma.
N. Engl. J. Med.
327:
1198-1203
[Abstract].
22.
Giannini, D.,
A. Carletti,
F. L. Dente,
E. Bacce,
A. DiFranco,
B. Vagaggini, and
P. L. Paggiaro.
1996.
Tolerance to the protective effect of salmeterol on allergen challenge.
Chest
110:
1452-1457
23.
Simons, F. E.,
T. V. Gerstener, and
M. S. Cheang.
1997.
Tolerance to the
bronchoprotective effect of salmeterol in adolescents with exercise-
induced asthma using concurrent inhaled glucocorticoid treatment.
Pediatrics
99:
655-659
24. Finney, D. J. 1978. Statistical Methods in Biological Assay, 3rd ed. Charles Griffin & Co., London. 105-147.
25. Schuirmann, D. J.. 1987. Pharmacometrics. A comparison of the two one-sided tests procedure and the power approach for assessing the equivalence of average bioavailability. J. Pharmacokinet. Biopharm. 15: 657-680 [Medline].
26. Lan, K. K. G., and D. L. Demets. 1983. Design and analysis of group sequential tests based on the type I error spending function. Biometrika 74: 149-154 .
27. O'Brien, P. C., and T. R. Fleming. 1979. A multiple testing procedure for clinical trials. Biometrics 35: 549-556 [Medline].
28. Ahrens, R. C., J. S. Nelson, C. Lux, P. Ecklund, M. Teresi, and G. Milavetz. 1992. A method for determining bioequivalence of beta agonist metered-dose inhalers. J. Allergy Clin. Immunol. 89(Part 2):340. [Abstract 783]
29. Rogers, D. F., and J. N. Ganderte. 1995. Determining equivalence of inhaled medications. Consensus statement from a workshop of the British Association for Lung Research, held at Royal Brompton National Heart & Lung Institute, London, June 24, 1994. Respir. Med. 89: 253-261 [Medline].
30. Beaty, R., E. Harman, L. Molino, and L. Hendeles. 1992. The dose- response of albuterol MDI during acute attacks of nocturnal asthma (abstract). Am. J. Respir. Crit. Care Med. 145: A66 .
31. Harris, J. B., R. C. Ahrens, G. Milavetz, R. A. Ries, L. Annis, and M. E. Eddy. 1986. Relative potencies and rates of decline in effect of inhaled albuterol and terbutaline. J. Allergy Clin. Immunol. 77: 147 .
32. Petitti, D. B. 1994. Meta-Analysis, Decision Analysis, and Cost-Effectiveness Analysis. Oxford University Press, New York. 119-121.
33. Ahrens, R. C.. 1991. On comparing beta adrenergic agonists. Ann. Allergy 67: 296-298 [Medline].
34. Parameswaran, K., M. Inman, M. Morris, E. Summers, D. Kamada, J. Otis, P. O'Byrne, and F. E. Hargreave. 1998. Bioequivalence of CFC salbutamol pressurized MDI and HFA salbutamol pressured MDI on bronchoconstriction to methacholine (abstract). Eur. Respir. J. 12: 65s .
This article has been cited by other articles:
![]() |
L. Hendeles, G. L. Colice, and R. J. Meyer Withdrawal of Albuterol Inhalers Containing Chlorofluorocarbon Propellants N. Engl. J. Med., March 29, 2007; 356(13): 1344 - 1351. [Full Text] [PDF] |
||||
![]() |
M. B. Dolovich, R. C. Ahrens, D. R. Hess, P. Anderson, R. Dhand, J. L. Rau, G. C. Smaldone, and G. Guyatt Device Selection and Outcomes of Aerosol Therapy: Evidence-Based Guidelines: American College of Chest Physicians/American College of Asthma, Allergy, and Immunology Chest, January 1, 2005; 127(1): 335 - 371. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J. Coffey, B. Wilfond, and L. F. Ross Ethical Assessment of Clinical Asthma Trials Including Children Subjects Pediatrics, January 1, 2004; 113(1): 87 - 94. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. C. AHRENS, M. E. TERESI, S.-H. HAN, D. DONNELL, J. A. VANDEN BURGT, and C. R. LUX Asthma Stability after Oral Prednisone . A Clinical Model for Comparing Inhaled Steroid Potency Am. J. Respir. Crit. Care Med., October 1, 2001; 164(7): 1138 - 1145. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. A. Stewart, R. C. Ahrens, S. Carrier, M. Frosolono, C. Lux, S.-H. Han, and G. Milavetz Demonstration of In Vivo Bioequivalence of a Generic Albuterol Metered-Dose Inhaler to Ventolin Chest, March 1, 2000; 117(3): 714 - 721. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Proc. Am. Thorac. Soc. | Am. J. Respir. Cell Mol. Biol. |