|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| |
ABSTRACT |
|---|
|
|
|---|
In 18 patients with chronic obstructive pulmonary disease intubated and mechanically ventilated, we
prospectively randomized 200 µg fenoterol-80 µg ipratropium bromide (four puffs) from a metered-dose inhaler (MDI) versus 1.25 mg fenoterol-500 µg ipratropium bromide in 5 ml saline from a nebulizer (NEB). Respiratory mechanics were assessed before and 30 min after the end of each delivery by
the rapid end-inspiratory airway occlusion technique. We did vary on single breaths the inflation flow
(
) from 0.2 to 1.2 L · s
1, at constant inflation volume. The total respiratory resistance of the respiratory system (Rrs) was partitioned into airway (Rint,rs) and tissue (
Rrs) resistances. We found that
Rrs was equivalently reduced, from 16.49 ± 1.37 to 14.85 ± 1.88 cm H2O · L
1 · s with MDI (p < 0.05)
and from 18.04 ± 1.85 to 15.15 ± 1.33 cm H2O · L
1 · s with NEB (p < 0.01). Whereas the prevailing
effect of MDI was to reduce Rint,rs, that of NEB was to decrease
Rrs. In addition, the
resistance of
the respiratory system over the whole range of
was significantly affected by NEB but not by MDI.
| |
INTRODUCTION |
|---|
|
|
|---|
The medical management of patients with chronic obstructive
pulmonary disease (COPD) who receive tracheal intubation
and mechanical ventilation for acute respiratory failure
(ARF) commonly includes the administration of bronchodilator aerosols in order to reduce the dynamic hyperinflation and
the markedly elevated airway resistance. Aerosols can be generated by two distinct devices, namely, metered-dose inhalers
(MDI) and small-volume nebulizers (NEB). The optimal deposition of aerosols to the lower respiratory tract in these patients depends on a lot of factors, including the device and its
location, the particle size, the drug dose given, factors related
to ventilator and circuit (tidal volume, duty cycle, humidification), presence and size of endotracheal tube, and disease severity (1). The choice between MDI and NEB is also based
upon the effects of the bronchodilator on the respiratory mechanics and cost-effectiveness considerations. In vitro investigations have been done to study the lung deposition of aerosols during mechanical ventilation with MDI or NEB by
varying some of the factors mentioned above (2). From
these in vitro studies, recommendations have been proposed
to optimally use MDI and NEB in intubated patients (1). Only
a few studies assessed the effectiveness of MDI or NEB in mechanically ventilated patients. Noncomparative studies have shown that MDI was effective in improving respiratory mechanics (8). Studies comparing MDI and NEB showed either no difference (12, 13) or a greater effect of NEB over
MDI (14). To our knowledge no study compared the cost-
effectiveness for MDI and NEB administration. Therefore,
some controversy remains regarding the choice of MDI versus
NEB administration of bronchodilator aerosols in intubated
and mechanically ventilated patients (15). Fenoterol-ipratropium bromide combines a selective
2-adrenergic agonist,
fenoterol, with a cholinergic antagonist, ipratropium bromide,
and has been found more effective than ipratropium alone (9).
No study in the literature compared the effects on respiratory mechanics of fenoterol-ipratropium bromide administered by
MDI or NEB to mechanically ventilated patients with COPD.
Therefore, in the present study in mechanically ventilated patients with COPD in ARF, we aimed at studying in detail the
respiratory mechanics, and specifically the flow-resistive properties of the respiratory system, after delivery of fenoterol-
ipratropium bromide by either MDI or NEB.
| |
METHODS |
|---|
|
|
|---|
Patients
We enrolled 18 patients (13 male) who received tracheal intubation and mechanical ventilation for ARF in our Medical Intensive Care Unit. Their mean ± SEM values of age, height, and weight were 67 ± 3 yr, 166 ± 3 cm, and 76 ± 3 kg, respectively. They all had COPD, which was diagnosed by clinical history, chest radiographs, and pulmonary function tests. The mean values of FEV1 and FEV1/VC before ARF were 0.77 ± 0.06 L (31 ± 3% of predicted) and 0.46 ± 0.04 (63 ± 6% of predicted), respectively (16). Acute respiratory failure had been triggered by an acute exacerbation in 10 patients and pneumonia in eight patients. They were investigated 1 to 10 d after the onset of tracheal intubation and mechanical ventilation (mean ± SEM 2.7 ± 0.7 d). The investigation was approved by the institutional ethics committee, and informed consent was obtained from the next of kin for each patient.
The patients were orotracheally intubated (Mallinckrodt cuffed-endotracheal tube 7.5, 8, or 8.5 mm ID; Mallinckrodt Laboratories, Athlone, Ireland) and mechanically ventilated (Cesar; Taema, Anthony, France) under volume-controlled mode with a squared inspiratory flow on zero end-expiratory pressure (ZEEP). During the study all patients were sedated with midazolam (0.2 mg/kg) and paralyzed with atracurium bresylate (0.3 to 0.6 mg/kg). The mean values of the
baseline ventilatory settings, which were kept constant throughout the
experiment, are listed in Table 1. The inspiratory to total cycle duration ratio (TI/Ttot) was 0.27 ± 0.01. Airflow (
) was measured with a
heated pneumotachograph (Fleisch No.2; Fleisch, Lausanne, Switzerland) inserted between the endotracheal tube and the Y-piece of the
ventilator. The pressure drop across the two ports of the pneumotachograph was measured with a differential piezoelectric pressure
transducer (163PC01D36, ± 12.7 cm H2O; Micro Switch, Freeport,
IL). The response of the pneumotachograph was linear over the experimental range of
. Pressure at the airway opening (Pao) was measured proximal to the endotracheal tube with a piezoelectric pressure
transducer (143PC03D, ± 176 cm H2O; Micro Switch). Tracheal pressure (Ptr) was measured via a polyethylene catheter (1.5 mm ID),
with multiple side holes and an occluded end hole, placed 2 cm past
the carinal end of the endotracheal tube and connected to a piezoelectric pressure transducer (143PC03D, ± 176 cm H2O; Micro Switch).
With the system used to measure Pao and Ptr, there was no appreciable shift or alteration in amplitude up to 20 Hz. The equipment dead
space (not including the endotracheal tube) was 150 ml. All variables were recorded on an IBM compatible computer by a 12-bit analog digital board (DT2801-A) interfaced with data acquisition software (Labdat; RHT-Infodat Inc., Montreal, Canada) at a sample frequency of 100 Hz. Subsequent data analysis was made with Anadat (RHT-Infodat). In this analysis, inflation volume (
V) was obtained by digital integration of the
signal. Special care was taken to avoid gas
leaks in the equipment and around the tracheal cuff. To reduce the effects of the compliance and resistance of the system connecting the
subjects to the ventilator on the mechanics measurements, we used a
single length of standard adult low compliance tubing supplied with
the machine (2 cm ID, 110 cm long) and omitted the humidifier during
the assessment of respiratory mechanics.
|
Arterial blood gases were measured with an ABL 500 blood gas analyzer (Radiometer, Copenhagen, Denmark). Before the experiment, PaO2 averaged 84 ± 6 mm Hg, PaCO2 averaged 53 ± 3 mm Hg, and pH was 7.41 ± 0.02 under the baseline ventilatory settings listed in Table 1. Heart rate, systemic blood pressure measured noninvasively (HPM 1008B; Hewlett-Packard, Waltham, MA), and transcutaneous O2 saturation (SpO2) (HPM 1020A; Hewlett-Packard) were continuously monitored.
Experimental Design
Basically, we underwent a prospective single-blind, randomized, crossover study. Patients received sequentially in a random order (random order table) fenoterol-ipratropium bromide by MDI and NEB administered by the same respiratory therapist. A period of at least 10 h was allowed between the administration of the bronchodilator with the two modalities. Fenoterol-ipratropium bromide delivered by MDI was given as four puffs (50 µg fenoterol and 20 µg ipratropium bromide per puff) by using a nonvalved MDI adapter (Medispacer; Baxter Laboratories, Maurepas, France). The MDI adapter was positioned in the inspiratory limb of the ventilator circuit 15 to 20 cm from the Y piece. After shaking the MDI canister, MDI was actuated just before the onset of the mechanical breath and a 4-s inflation hold was then applied. We waited 60 s between each actuation and repeated the same maneuver until the total dose was delivered. During this procedure, the ventilatory settings were not changed. Treatment with NEB was given at a dosage of 1.25 mg fenoterol and 500 µg ipratropium bromide in 5 ml of saline. The drug was placed in a nebulizer (Misty-Neb; Baxter Laboratories), which was inserted at the same place in the ventilator circuit as for the MDI inhalation. This nebulizer has a dead volume of 2 to 5 ml of gas and of 0.5 ml of saline. The gas flow from the ventilator was 5 L/min. The device was run until almost dry by visual inspection, which was obtained after 30 min on average. NEB was connected to a specific port at the ventilator allowing us to use the ventilator's flow during nebulization. A correction factor was used to keep the tidal volume delivered by the ventilator constant during nebulization. Fenoterol-ipratropium bromide specimens and MDI adapters and nebulizers were provided by Boehringer-Ingelheim Laboratories (Reims, France) and Baxter Laboratories, respectively.
Respiratory mechanics and vital signs (heart rate, SpO2, and systemic blood pressure) were assessed before and 30 min after the end of each modality of administration. Before inhalation of bronchodilator and before respiratory signal sampling, the trachea was gently suctioned to remove secretions. Six patients received inhaled bronchodilators other than fenoterol-ipratropium bromide before entry into the present study. In these patients, this treatment was withheld for at least 4 h before the onset of the investigation.
Physiologic Measurements and Data Analysis
Patients were investigated in a semirecumbent position.
Respiratory Mechanics
Respiratory mechanics were assessed by the constant-
rapid airway
occlusion method previously described in detail (17). Whereas baseline
V was kept constant (Table 1),
was changed randomly from
0.2 to 1.2 L/s for single test breaths by regulating TI with the appropriate knob of the ventilator (iso-
V experiment). The end-inspiratory
occlusions, obtained by pressing the end-inspiratory hold knob on the
ventilator, lasted 5 s. Before each test breath, an end-expiratory occlusion was performed by pressing the end-expiratory hold button on the
ventilator. This allowed us to quantify PEEPi and to start the test
breath from a fixed static elastic equilibrium condition. It should be
noted that on ZEEP the ventilator generated a slightly positive end-expiratory pressure, averaging 1.13 ± 0.11 cm H2O. Therefore, the
pressure measured by the end-expiratory occlusion is the sum of
PEEPi and the positive pressure generated by the ventilator. We
termed this pressure PEEP total (PEEPt). Because PEEPi implies dynamic pulmonary hyperinflation (i.e., that the end-expiratory lung
volume during mechanical ventilation exceeds the relaxation volume
of the respiratory system [Vr]), we also measured the difference between end-expiratory lung volume and Vr (EELV) by reducing the
ventilator frequency to its lowest value during baseline expiration,
thus prolonging expiratory duration to allow the patient to exhale to
Vr. Vr was achieved when expiratory flow became nil and expiratory
occlusion resulted in no change in airway pressure (i.e., no PEEPi).
After each test breath, the baseline ventilation was resumed until
V,
, Pao, and Ptr returned to their baseline values (usually within a few
breaths). Each measurement was done twice.
After end-inspiratory airway occlusions, Ptr and Pao exhibited an
initial rapid drop (Pmax-P1) followed by a slow decay to an apparent
plateau pressure. During this period, the contribution of reduction in
pressure caused by volume loss by continuing gas exchange should be
negligible. Ptr measured at 5 s was taken to represent the static end-inspiratory elastic recoil pressure of the respiratory system (Pst,rs).
Dividing Pmax,tr-Pst,rs and Pmax,tr-P1,tr by the
immediately preceding the occlusion, total resistance (Rrs) and interrupter resistance
(Rint,rs) of the respiratory system were obtained. The additional tissue resistance (
Rrs) was calculated as the difference between Rrs
and Rint,rs. In computation of Rint,rs the errors caused by the closing
time of the ventilator valve were corrected as previously described
(18). The static elastance of the respiratory system (Est,rs) was computed by dividing the values of Pst,rs-PEEPt by baseline
V.
Model and Curve-fitting
Our data were analyzed in terms of the model of the respiratory system proposed by D'Angelo and colleagues (18) for humans (Figure 1). This model comprises two compartments in parallel. The first is a dashpot representing Rint,rs, which explains the initial fast pressure drop observed in Prs immediately after the end-inspiratory occlusion. Rint,rs is the sum of the lung and chest wall. Bates and colleagues (19) found that in dogs, in the absence of the chest wall, the initial pressure jump reflects only the airway resistance. Contrary to dogs (20), in normal anesthetized paralyzed humans (21) and in patients with COPD (17), the chest wall does not substantially contribute to Rint,rs. The second compartment of the model in Figure 1 is a Kelvin body, which consists of a standard static elastance (Est,rs) in parallel with a Maxwell body, i.e., a spring E2 and a dashpot R2 arranged serially. In normal anesthetized paralyzed subjects, E2 and R2 probably virtually entirely reflect the viscoelastic properties of the tissues of the lungs and chest wall (21).
|
During constant-
inflation, the model in Figure 1 predicts that
Rrs should increase with TI according to the following function (14):
|
(1) |
where
2 = R2/E2.
Because during constant-
inflation TI =
V/
, Equation 1 can be
rewritten (14)
|
(2) |
In the present study, we could fit Equation 1 to our experimental
data of
Rrs before and after the inhalation of fenoterol-ipratropium bromide by MDI and NEB in all of the patients, as shown in Figure 2
in a representative patient.
|
In vitro Assessment of Inhaled Mass
Inhaled mass, defined as the quantity of drug actually reaching the end of the endotracheal tube (22), was determined in vitro under conditions that replicated the clinical study. The mean values of the ventilatory settings given in Table 1 were used. The drug was delivered with the same devices and at the same dosages. With NEB, the duration of nebulization was set to 30 min. With MDI, four puffs were given at 1-min intervals and then the mechanical ventilation was prolonged to a 30-min duration. A circular low-resistance absolute filter 70 mm in diameter (Gelman Sciences, Ann Arbor, MI) was interposed between the distal extremity of the endotracheal tube and a low-resistive bag (3). At the end of the 30-min period, the filter was removed and the bronchodilator molecules were extracted. Extraction was achieved by placing the filter in a glass bag beaker containing 10 ml 0.1N NaOH that was vortexed for 1 min. Then the solution was centrifuged between 0.500 and 0.600 × g for 10 min. Six experiments were conducted with NEB and 12 with MDI then averaged. Fenoterol was assayed as described by Diot and colleagues (5). Briefly, 2-ml samples of supernatant were read by spectrophotometer (Beckman DU-6; Beckman Instruments, Irvine, CA) at a wavelength of 240 nm, as determined by a previous spectrometer analysis of fenoterol in a 0.1 N NaOH solution. The blank was prepared by processing a circular filter 70 mm in diameter without fenoterol, as described previously. The standard curve was established by assaying known amounts of fenoterol between 50 and 300 µg pipetted onto the surface of a circular filter 70 mm in diameter. Its reproducibility was established on three different days. Assuming that the maximum mass of ipratropium bromide deposited on the filters would be 35% and 25% of the mass of drug delivered by the MDI or NEB, respectively (23), the maximum estimated concentrations of ipratropium bromide that could be obtained after washing the filter with 10 ml 0.1 N NaOH would be 2.8 and 12.5 µg/ml, respectively. To our knowledge, there is no method that would allow us to assay ipratropium bromide at such low concentrations, and we focused the analysis on fenoterol.
Particle Size Distribution
The particle size distribution of fenoterol-ipratropium bromide obtained with each aerosol generator was determined in vitro by using a laser velocimeter (APS 33 T. S. I., St. Paul, MN). The bronchodilator was delivered under the same conditions as in the clinical study (same dosages, same devices, same ventilator with respiratory parameters producing the same flow rate by using the mean values of the ventilatory settings, as given in Table 1).
Data Analysis
The investigators who performed the postsampling analysis of the respiratory signals were blinded to the treatment modality. The comparisons of the values of respiratory mechanics and vital signs before and
after inhalation were made within and between delivery modalities by
using Student's paired t tests. We applied Bonferroni's correction to
obtain the significance level that an individual p value must satisfy to
achieve significance at a p value < 0.05. Regression analysis was made
by the least-squares method. Comparisons of the values of respiratory
resistances obtained at different
during the iso
V experiments were
made by repeated measures analysis of variance (ANOVA). The values were expressed as their mean ± SEM. For our statistics, we used
SPSS software (SPSS for Microsoft Windows V 6.0.1; SPSS Inc., Chicago, IL).
| |
RESULTS |
|---|
|
|
|---|
All patients investigated in the present protocol exhibited
PEEPi that averaged 6.52 ± 0.89 cm H2O and was associated
with expiratory flow limitation. In all patients indeed, the passive expiratory
-
V curve exhibited a concavity toward the
volume axis, a finding that has been found associated with expiratory flow limitation (24). Consistent with PEEPi, all patients exhibited dynamic hyperinflation, as reflected by the
values of EELV (Table 2).
|
Respiratory Mechanics at Baseline Ventilatory Settings
As shown in Table 2, at the baseline ventilatory settings, the
mean values of inspiratory resistances, PEEPt, EELV, and
Est,rs before inhalation of fenoterol-ipratropium bromide
were not different between MDI and NEB. Both modalities of
administration reduced to the same extent the mean values
of Rrs, EELV, and PEEPt. However, the two components of
Rrs were not similarly affected. Indeed, with MDI, Rint,rs
decreased significantly, with no change in
Rrs. By contrast,
with NEB, an opposite finding was observed, namely, a significant reduction of
Rrs without any significant change in
Rint,rs. The values of Est,rs were not different before MDI and NEB and did not change significantly after inhalation of
fenoterol-ipratropium bromide with either mode of delivery
(Table 2).
Respiratory Resistances During iso-
V Experiment
In all patients, Rint,rs increased linearly with increasing
, at
constant
V (Figure 3) according to the following function:
|
(3) |
|
where K1 and K2 are Rohrer's constants whose mean ± SEM
values are given in Table 3. The mean values of K1 and K2
were not different before inhalation between MDI and NEB
and did not change significantly after inhalation of fenoterol-ipratropium bromide from MDI. The same was true for the
values of K1 after inhalation from NEB. By contrast, the values of K2 decreased significantly after administration by NEB
and therefore were significantly lower than the values of K2
obtained after inhalation by MDI (Table 3). However, by applying repeated-measures ANOVA to the data of Figure 3, we
found that Rint,rs was significantly influenced by
(p < 0.0001), but not by the modality of delivery.
|
The average values of the "viscoelastic" constants of the
respiratory system of our patients with COPD are shown in
Table 4. The values of R2,
2, and E2 were not different before
administration between the two modalities and did not change
significantly after inhalation. The average iso-
V relationships of
Rrs with
are illustrated in Figure 4. The values of
Rrs decreased progressively and significantly (p < 0.0001)
with increasing
and were significantly reduced with NEB as
compared with MDI over the whole range of
(p < 0.05 by
repeated-measures ANOVA).
|
|
The average iso-
V relationship of Rrs with
is displayed
in Figure 5. Before inhalation, the values of Rrs initially decreased until a minimum value was reached at a
of 0.6 L/s;
thereafter, Rrs tended to increase slightly with further increase in
. The values of Rrs were significantly reduced by
NEB administration (p < 0.01 by repeated-measures ANOVA).
|
Vital Signs
As shown in Table 5, heart rate, systolic and diastolic systemic blood pressure, and SpO2 did not change significantly with either modality of inhalation.
|
| |
DISCUSSION |
|---|
|
|
|---|
In this study of intubated and mechanically ventilated patients
with COPD in ARF, we have found that: (1) 30 min after inhalation of fenoterol-ipratropium bromide, Rrs, EELV, and
PEEPt significantly decreased; (2) at baseline
Rrs was reduced through a decrease in Rint,rs with MDI but through a
decrease in
Rrs with NEB; and (3) over a range of
the
Rohrer's constant K2 and the iso-
V changes of
Rrs were significantly influenced by NEB but not by MDI administration.
To our knowledge, only one study reported the effects of
inhaled fenoterol-ipratropium bromide in intubated and mechanically ventilated patients with COPD (9). In this study,
Fernandez and coworkers (9) administered by MDI two puffs
of fenoterol-ipratropium bromide (half the dose we presently
used) in 12 patients with COPD intubated for an acute exacerbation. They observed that 60 min later PEEPi and Rrs were
reduced on average by 24% and 16%, respectively. These
changes were greater than those obtained in the present study,
suggesting that the maximal effect of fenoterol-ipratropium bromide could be obtained later than the 30-min period we
have chosen; however, in nonintubated asthmatics, the maximal effects on FEV1 of nebulized fenoterol and of nebulized
ipratropium bromide were observed 30 min after drug administration (25). However, the study of Fernandez and coworkers (9) and our own are difficult to compare for the following
reasons. First of all, they administered the drug in a different
way than we did in that they used a catheter connecting the
MDI device to the endotracheal tube and delivered the bronchodilator through a manual inflation. Second, to our knowledge, the dose-response relationship of inhaled fenoterol-ipratropium bromide is unknown in intubated and mechanically
ventilated patients. Third, in their study, the measurements of
Rrs included the
resistance caused by the endotracheal
tube, resulting in an overestimation of the intrinsic total resistance of the respiratory system. Finally, the duration of the
end-inspiratory pause in their study was much shorter than what
we used, leading to an overestimation of the values of Pst,rs
and an underestimation of those pertaining to Rrs.
Surprisingly, in our study, we found that apparently the device used to deliver fenoterol-ipratropium bromide to the respiratory system could have influenced the components of the
inspiratory resistances differently. Intuitively, the main therapeutic effect of a bronchodilator we were expecting was a reduction in airway resistance. Only a few studies have partitioned Rrs into its airway and its tissue components. Dhand
and colleagues (10, 26), in mechanically ventilated patients
with COPD, found that the significant decrease of Rrs after
inhalation of albuterol was essentially due to a reduction of
Rint,rs. In our study, at baseline
applied during mechanical
ventilation, the values of Rint,rs were significantly reduced after MDI but not after NEB administration. The variability between patients might have blunted the significance of the difference in Rint,rs observed with NEB (Table 2). The baseline
values of Rint,rs in our study are lower than those previously reported in mechanically ventilated patients with COPD in
ARF (17, 27). This difference can be partly explained by
the relatively low
applied to our patients with COPD. We
tried to extend the results of Dhand and colleauges (10, 26)
obtained at a single
by varying
at a constant
V. The average value of the constant K2, which was thought to reflect
the additional resistance caused by the turbulent flow in the
central airways, was significantly reduced by NEB without any
significant effect of MDI.
Rrs was not significantly changed with albuterol administered by MDI in the patients studied by Dhand and colleagues
(10, 26). We have observed the same finding by using MDI to
deliver fenoterol-ipratropium bromide. In marked contrast, in
our study, the baseline values of
Rrs decreased significantly
after inhalation of fenoterol-ipratropium by using the NEB
device. In our patients,
Rrs fitted Equation 1 in the four experimental conditions. After inhalation of fenoterol-ipratropium bromide, with both devices, the average values of the
"viscoelastic constants" did not change significantly (Table 4).
However, the iso
V relationships of
Rrs to
were not similar regarding the modality of inhalation (Figure 5). At any
given
, the mean values of
Rrs were significantly different
before and after NEB inhalation, which was not the case with
MDI administration. Some variability in our results may have blunted the significance of these changes. There is no previous data, to our knowledge, upon the effects of bronchodilator administration on the "viscoelastic constants" of the respiratory
system to compare with. The structural meaning of the viscoelastic constants to date remains unknown. Assuming that
the effects of NEB administration on
Rrs are true, how could
a bronchodilator affect tissue resistance and why such a difference between the two modalities of administration? In animals, the contribution of tissue resistance to the increase in
lung resistance observed after induced bronchoconstriction has been shown to be dramatic. Indeed, in rats challenged
with nebulized metacholine, Nagase and colleagues (30) found
that tissue resistance increased by 9-fold and represented
81.8% of total lung resistance after metacholine inhalation.
Because no bronchodilator was further administered we cannot argue from this study that bronchodilator agent may reverse bronchoconstriction by acting predominantly at the tissue level. In addition, in our patients, the low level of
applied during baseline ventilation may have made of
Rrs
the predominant component of Rrs relative to Rint,rs. Why
did the two modalities of administration behave differently in
this respect? This might be due to differences in drug deposition into the respiratory tract. In vitro, the inhaled mass of
fenoterol averaged 225 ± 11.6 µg with NEB and 60.3 ± 4.2 µg
with MDI (p < 0.001). These values corresponded to 18.1% of the mass of fenoterol placed in the nebulizer and to 30.1% of that delivered by the MDI device. As already mentioned, for
technical reasons, we did not measure the inhaled mass of ipratropium bromide. The results of the particle size distribution
of fenoterol-ipratropium bromide obtained with both aerosol
devices in vitro are shown in Table 6. The values of geometric
standard deviation (GSD) were those of polydisperse aerosols
with each generator. However, the mean value of mass median aerodynamic diameter (MMAD) obtained with MDI was
significantly higher than that obtained with NEB, whereas the
percentage of particles with MMAD < 5 µm was significantly lower with MDI than with NEB. The probability of deposition
in thoracic airways can be estimated from the model (31). Accordingly, an overall deposition of 34.7% and of 28.5% could
be computed for MDI and NEB, respectively. With MDI and
NEB, the probability of deposition can be estimated to 24%
and 21%, and to 10.7% and 7.5% at the alveolar and airway
levels, respectively. Hence, by merging the results of these two
in vitro experiments together, we could try to estimate, at least
for fenoterol, the intrathoracic quantitative mass deposition.
The total mass of fenoterol deposited could amount to 60 µg
with MDI and to 225 µg with NEB. The mass of fenoterol deposited in the airways could be around 6 µg with MDI and 17 µg
with NEB, whereas the corresponding values for the alveolar
deposition could be 14 and 47 µg, respectively. These in vitro
studies showed that the deposition of fenoterol could have
been different with the two modalities of inhalation, the NEB
modality delivering to the lungs four times more fenoterol than the MDI modality. The quantity of drug deposited at specific levels of the respiratory tract could have also been different. However, we did not obtain estimates of lung deposition
for ipratropium bromide. So, it is speculative to infer from
these results the difference of the values of the components of
Rrs we observed in the clinical study. In addition, the agreement between in vitro and in vivo studies is inconstant. For example, in the study of Fuller and colleagues (12) in mechanically ventilated patients, the lung deposition of radiolabeled
fenoterol was higher when the drug was administered by MDI
than by NEB, whereas the changes in peak inspiratory pressure were not different between the two modalities. It must be
stressed, however, that recent studies, where the inhaled mass
was accurately assessed (4, 5, 32), in fact showed that large
amounts of drug can be delivered to the patients with nebulizers. Hence, our present data are consistent with these studies,
whereas the initial study of Fuller and colleagues (12) is not.
|
In conclusion, our study has provided a systematic description of the effects on respiratory resistances of the administration of inhaled fenoterol-ipratropium bromide given by MDI and NEB in patients with COPD intubated and mechanically ventilated for acute respiratory failure. We have found that both devices equivalently reduced Rrs without side effects, but they apparently affected differently the components of Rrs. Finally, the in vitro assessment of the inhaled mass of fenoterol showed that the nebulizer delivered a larger amount of drug than the metered-dose inhaler we tested.
| |
Footnotes |
|---|
Correspondence and requests for reprints should be addressed to Dr. Claude Guerin, Service de Réanimation Médicale et d'Assistance Respiratoire, 93 Grande Rue de la Croix-Rousse, 69004 Lyon, France.
(Received in original form October 24, 1997 and in revised form November 10, 1998).
Acknowledgments: The writers wish to thank Dr. Patrice Diot, CHU Bretonneau, Tours, France, for his cooperation in performing the inhaled mass determinations.
Supported by a grant from Baxter France Laboratories.
| |
References |
|---|
|
|
|---|
1.
Dhand, R., and
M. J. Tobin.
1997.
Inhaled bronchodilator therapy in mechanically ventilated patients.
Am. J. Respir. Crit. Care Med.
156:
3-10
2. O'Doherty, M. J., S. H. L. Thomas, C. J. Page, D. F. Treacher, and T. O. Nunan. 1992. Delivery of a nebulized aerosol to a lung model during mechanical ventilation. Am. Rev. Respir. Dis. 146: 383-388 [Medline].
3. O'Riordan, T. G., M. J. Greco, R. J. Perry, and G. C. Smaldone. 1992. Nebulizer function during mechanical ventilation. Am. Rev. Respir. Dis. 145: 1117-1122 [Medline].
4. O'Riordan, T. G., L. B. Palmer, and G. C. Smaldone. 1994. Aerosol deposition in mechanically ventilated patients. Am. Rev. Respir. Dis. 149: 214-249 .
5. Diot, P., L. Morra, and G. C. Smaldone. 1995. Albuterol delivery in a model of mechanical ventilation. Am. J. Respir. Crit. Care Med. 152: 1391-1394 [Abstract].
6. Thomas, S. H. L., M. J. O'Doherty, C. J. Page, D. F. Treacher, and T. O. Nunan. 1993. Delivery of ultrasonic nebulized aerosols to a lung model during mechanical ventilation. Am. J. Respir. Crit. Care Med. 148: 872-877 .
7. Fink, J. B., R. Dhand, A. G. Duarte, J. W. Jenne, and M. J. Tobin. 1996. Aerosol delivery from a metered-dose inhaler during mechanical ventilation. Am. J. Respir. Crit. Care Med. 154: 382-387 [Abstract].
8. Fernandez, A., A. Lazaro, A. Garcia, C. Aragon, and E. Cerda. 1990. Bronchodilators in patients with chronic obstructive pulmonary disease on mechanical ventilation. Am. Rev. Respir. Dis. 141: 164-168 [Medline].
9. Fernandez, A., J. Munoz, B. de la Calle, I. Alia, A. Ezpeleta, M.A. de la Cal, and A. Reyes. 1994. Comparison of one versus two bronchodilators in ventilated COPD patients. Intensive Care Med. 20: 199-202 [Medline].
10. Dhand, R., A. Jubran, and M. J. Tobin. 1995. Bronchodilator delivery by metered-dose inhaler in ventilator-supported patients. Am. J. Respir. Crit. Care Med. 151: 1827-1833 [Abstract].
11.
Manthous, C. A.,
W. Chatila,
G. A. Schmidt, and
J. B. Hall.
1995.
Treatment of bronchospasm by metered-dose inhaler albuterol in mechanically ventilated patients.
Chest
107:
210-213
12. Fuller, H. D., M. B. Dolovich, G. Posmituck, W. Wong, Pack, and M. T. Newhouse. 1990. Pressurized aerosol versus jet aerosol delivery to mechanically ventilated patients. Am. Rev. Respir. Dis. 141: 440-444 [Medline].
13.
Gay, P. C.,
H. G. Patel,
S. B. Nelson,
B. Gilles, and
R. D. Hubmayr.
1991.
Metered-dose inhaler for bronchodilator delivery in intubated,
mechanically ventilated patients.
Chest
99:
66-71
14. Manthous, C. A., J. B. Hall, G. A. Schmidt, and L. D. H. Wood. 1993. Metered-dose inhaler versus nebulized albuterol in mechanically ventilated patients. Am. Rev. Respir. Dis. 148: 1567-1570 [Medline].
15. Newhouse, M. T., and H. D. Fuller. 1993. Rose is a rose is a rose? Aerosol therapy in ventilated patients: nebulizers versus metered-dose inhalers: a continuing controversy. Am. Rev. Respir. Dis. 148: 1444-1446 [Medline].
16. Morris, J. F., A. Koski, and L. C. Johnson. 1971. Spirometric standards for healthy nonsmoking adults. Am. Rev. Respir. Dis. 103: 57-67 [Medline].
17.
Guérin, C.,
M. L. Coussa,
N. T. Eissa,
C. Corbeil,
M. Chassé,
J. Braidy,
N. Matar, and
J. Milic-Emili.
1993.
Lung and chest wall mechanics in mechanically ventilated COPD patients.
J. Appl. Physiol.
74:
1570-1580
18.
D'Angelo, E.,
E. Calderini,
G. Torri,
F. M. Robatto,
D. Bono, and
J. Milic-Emili.
1989.
Respiratory mechanics in anesthetized paralyzed
humans: effects of flow, volume and time.
J. Appl. Physiol.
67:
2556-2564
19.
Bates, J. H. T.,
M. S. Ludwig,
P. D. Sly,
K. Brown,
J. G. Martin, and
J. J. Fredberg.
1998.
Interrupter resistance elucidated by alveolar pressure
measurement in open-chest normal dogs.
J. Appl. Physiol.
65:
408-414
20.
Bates, J. H. T.,
K. A. Brown, and
T. Kochi.
1989.
Respiratory mechanics
in the normal dog determined by expiratory flow limitation.
J. Appl.
Physiol.
67:
2276-2285
21.
D'Angelo, E.,
F. M. Robatto,
E. Calderini,
M. Tavola,
D. Bono,
G. Torri, and
J. Milic-Emili.
1991.
Pulmonary and chest wall mechanics in
anesthetized paralyzed humans.
J. Appl. Physiol.
70:
2602-2610
22. Smaldone, G. C.. 1991. Drug delivery via aerosol systems: concept of "aerosol inhaled". J. Aerosol Med. 4: 229-236 [Medline].
23. Coleman, D. M., H. W. Kelly, and B. C. McWilliams. 1996. Therapeutic aerosol delivery during mechanical ventilation. Ann. Pharmacother. 30: 644-655 [Abstract].
24. Gottfried, S. B., A. Rossi, B. D. Higgs, P. M. A. Calverley, L. Zocchi, C. Bozic, and J. Milic-Emili. 1985. Noninvasive determination of respiratory system mechanics during mechanical ventilation for acute respiratory failure. Am. Rev. Respir. Dis. 131: 414-420 [Medline].
25.
Bryant, D. H..
1985.
Nebulized ipratropium bromide in the treatment of
acute asthma.
Chest
88:
24-29
26. Dhand, R., A. G. Duarte, A. Jubran, J. W. Jenne, J. B. Fink, P. J. Fahey, and M. J. Tobin. 1996. Dose-response to bronchodilator delivered by metered-dose inhaler in ventilator-supported patients. Am. J. Respir. Crit. Care Med. 154: 388-393 [Abstract].
27. Broseghini, C., R. Brandolese, R. Poggi, G. Polese, E. Manzin, J. Milic-Emili, and A. Rossi. 1988. Respiratory mechanics during the first day of mechanical ventilation in patients with pulmonary edema and chronic airway obstruction. Am. Rev. Respir. Dis. 138: 355-361 [Medline].
28. Bernasconi, M., Y. Ploysongsang, S. B. Gottfried, J. Milic-Emili, and A. Rossi. 1988. Respiratory compliance and resistance in mechanically ventilated patients with acute respiratory failure. Intensive Care Med. 14: 547-553 [Medline].
29. Tantucci, C., C. Corbeil, M. Chassé, J. Braidy, N. Matar, and J. Milic-Emili. 1991. Flow resistance in patients with chronic obstructive pulmonary disease in acute respiratory failure. Am. Rev. Respir. Dis. 144: 384-389 [Medline].
30.
Nagase, T.,
A. Moretto, and
M. S. Ludwig.
1994.
Airway and tissue behavior during induced constriction in rats: intravenous vs. aerosol administration.
J. Appl. Physiol.
76:
830-838
31. James, A. C., W. J. Stahlhofen, G. Rudolf, R. Köbrich, J. K. Briant, M. J. Egan, W. Dixon, and A. Birchall. 1994. Deposition of inhaled particles. In H. Smith, editor. Human Respiratory Tract Model for Radiological Protection. Pergamon Press, Oxfordshire. 231-299.
32. Palmer, L. B., G. C. Smaldone, S. R. Simon, T. G. O'Riordan, and A. Cuccia. 1998. Aerosolized antibiotics in mechanically ventilated patients: delivery and response. Crit. Care Med. 26: 31-39 [Medline].
This article has been cited by other articles:
![]() |
M. Tzoufi, S. D. Mentzelopoulos, C. Roussos, and A. Armaganidis The Effects of Nebulized Salbutamol, External Positive End-Expiratory Pressure, and Their Combination on Respiratory Mechanics, Hemodynamics, and Gas Exchange in Mechanically Ventilated Chronic Obstructive Pulmonary Disease Patients Anesth. Analg., September 1, 2005; 101(3): 843 - 850. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. A. Volta, V. Alvisi, S. Petrini, S. Zardi, E. Marangoni, R. Ragazzi, M. Capuzzo, and R. Alvisi The Effect of Volatile Anesthetics on Respiratory System Resistance in Patients with Chronic Obstructive Pulmonary Disease Anesth. Analg., February 1, 2005; 100(2): 348 - 353. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Schonhofer and S. Sortor-Leger Equipment needs for noninvasive mechanical ventilation Eur. Respir. J., October 1, 2002; 20(4): 1029 - 1036. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Proc. Am. Thorac. Soc. | Am. J. Respir. Cell Mol. Biol. |