|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| |
ABSTRACT |
|---|
|
|
|---|
Intrinsic positive end-expiratory pressure (PEEPi) is routinely determined under static conditions by occluding the airway at end-expiration (PEEPi,st). This procedure may be difficult in patients with chronic obstructive pulmonary disease (COPD) during spontaneous breathing, as both expiratory muscle activity and increased respiratory frequency often occur. To overcome these problems, we tested the hypothesis that the difference between maximum airway opening (MIP) and maximum esophageal (Ppl max) pressures, obtained with a Mueller maneuver from the end-expiratory lung volume (EELV), can accurately measure PEEPi,st. Using this method, we found that, in eight ventilator-dependent tracheostomized COPD patients (age 71 ± 7 yr), PEEPi,st averaged 13.0 ± 2.9 cm H2O. That measurement was validated by comparison with a reference static PEEPi (PEEPi,st-Ref ) taken at the same EELV adopted by patients during spontaneous breathing, and measured on the passive quasi-static pressure-volume (P/V) curve of the respiratory system, obtained during mechanical ventilation. PEEPi,st-Ref averaged 13.1 ± 3.0 cm H2O, i.e., a value essentially equal to PEEPi,st measured by means of our technique. We conclude that PEEPi,st can be accurately assessed in spontaneous breathing COPD patients by the difference between MIP and Ppl max during the Mueller maneuver.
| |
INTRODUCTION |
|---|
|
|
|---|
Dynamic pulmonary hyperinflation and intrinsic positive end-expiratory pressure (PEEPi) are well recognized events in patients with severe chronic obstructive pulmonary disease (COPD) during acute respiratory failure as well as in stable condition (1). PEEPi provides an inspiratory threshold load which may be substantial (1), and must be counterbalanced by the patients' inspiratory muscles before starting inspiration (4). Recent work has shown that in ventilator-dependent COPD patients, PEEPi accounts for about 40%, on average, of the total ventilatory workload (5, 6). In these patients accurate measurement of PEEPi is important for the correct setting of the ventilator, and particularly for the level of PEEP and continuous positive airway pressure (CPAP) to be applied (5, 7). In fact, low levels of PEEP/CPAP, compared with the actual PEEPi, will not provide sufficient assistance, whereas excessive levels of PEEP/CPAP can worsen pulmonary hyperinflation, thus reducing further the inspiratory muscle pressure generation capacity (8). Therefore, measurement of PEEPi is important in these patients. Unfortunately, the end-expiratory occlusion method commonly used to quantify PEEPi (4) seldom works in actively breathing patients during assisted ventilation. The esophageal balloon technique estimates the value of dynamic PEEPi (PEEPi,dyn) (5, 9). However, in patients with advanced COPD, PEEPi,dyn may be significantly lower than the mean end-expiratory static recoil pressure of the total respiratory system (PEEPi,st) measured with the end-respiratory occlusion technique (10). In this study we suggest a new method for measuring PEEPi,st in actively breathing patients. The rationale of this method is provided by the well-established relationship between the elastic recoil of the respiratory system and the pressure-generating capacity of the inspiratory muscles (13), such that the difference between the decrease in mouth pressure (Pao) and esophageal pressure (Ppl) during a sustained maximal or submaximal inspiratory effort against occluded airway at end-expiration should provide the value of PEEPi,st. The aim of our study was to assess the reliability of this method for measuring PEEPi,st during spontaneous breathing in ventilator-dependent COPD patients (5) in whom correct measurement of PEEPi is important.
| |
METHODS |
|---|
|
|
|---|
Patients
Eight ventilator-dependent COPD patients, admitted to the intensive care unit (ICU) of the Division of Pulmonary Disease, Medical Center for Rehabilitation, S. Maugeri Foundation, Veruno (NO), Italy, were recruited for this study. The diagnosis of COPD was confirmed by clinical history and pulmonary function tests (Table 1). Patients were all tracheostomized (cuffed tracheostomy cannula, 8 mm internal diameter and 11.5 cm in length; Shiley Inc., Irvine, CA) and had been mechanically ventilated (Evita ventilator, Dräger, Germany) for 45 ± 20 d before the study. All patients had undergone at least two weaning trials during the days preceding the study and were able to sustain a period of at least 20 min of unsupported breathing.
|
The study was conducted along the guidelines of the Declaration of Helsinki and the protocol was approved by the institution's ethics committee. Written informed consent was obtained from the patients or their next of kin.
Measurements
Flow, volume, and pleural and gastric pressures were measured as previously described (5, 9). Flow (
) was measured with a heated Lilly
pediatric-type pneumotachometer connected to a differential pressure transducer (Screenmate 701240-015009; Jaeger, Wurzburg, Germany), and inserted between the proximal tip of the tracheal cannula
and the Y connector of the ventilator (Figure 1). Volume (V) was calculated through numerical integration of the flow signal. Changes in
pleural (Ppl) and abdominal (Pab) pressures were estimated from
changes in esophageal and gastric pressures, respectively. Both Ppl
and Pab were measured using two balloon-tipped catheter systems
connected to two differential pressure transducers (143PC03D; Micro
Switch, Honeywell, Freeport, IL). The catheters were 80 cm in length
and 1.7 mm in internal diameter; the balloons were 10 cm in length
and 2.4 cm in circumference. The correct positioning of the esophageal balloon was assessed by means of the occlusion test (14). Another catheter, similar in length and internal diameter, and another
pressure transducer (143PC03D; Micro Switch, Honeywell) were used
to sample the pressure at the airway opening (Pao) via a side port inserted between the tracheostomy tube and the pneumotachometer.
Transpulmonary (PL) and transdiaphragmatic (Pdi) pressures were
obtained by subtracting Ppl from Pao and Pab, respectively. Minute
ventilation (
E), tidal volume (VT), inspiratory (TI) and expiratory
time (TE), total respiratory cycle duration (Ttot) breathing frequency (f), mean inspiratory flow (VT/TI), and the duty cycle (TI/Ttot were calculated as average values from one-minute continuous recordings of
flow and volume.
|
Dynamic pulmonary compliance (CdynL) and pulmonary resistance at midinspiratory volume (RL) were computed from PL,
, and V records as previously described (5). RL included the endotracheal tube resistance. PEEPi,dyn was measured as the decrease in Ppl preceding the start of inspiratory flow (5, 9) in patients who did not show
expiratory muscle activity (Figure 2). Five patients exhibited expiratory muscle activity, as indicated by a rise in Pab during expiration followed by a fall at the beginning of the inspiratory effort (Figure 3). In
these patients PEEPi,dyn measurement was corrected for the overestimation induced by expiratory muscle activity by subtracting the decrease in Pab from the drop in Ppl when both are present in the interval between the onset of inspiratory effort and the point of zero flow
(5, 9). Expiratory flow limitation during tidal breathing was qualitatively assessed by means of the technique of external positive end-
expiratory pressure (PEEP) withdrawal (15). Maximum inspiratory
pressure (MIP) and maximum pleural (Ppl max) pressure swings were
measured in each subject by means of the Mueller maneuver, i.e., a
sustained, maximum inspiratory effort against occluded airway starting at EELV (5, 9, 16).
|
|
Measurement of PEEPi,st. PEEPi,st during spontaneous breathing
was measured as the difference between MIP and Ppl max at the plateau. Patients enrolled for the present study were severe COPD patients. In this condition, it has been shown that an accurate evaluation
of inspiratory muscle strength (Pmus) is obtained only by adding the
total respiratory system recoil pressure (Prs) to maximum airway
opening pressure, as indicated by the equation: Pmus = Prs + MIP
(Equation 1) (13, 17). Because Pmus is equivalent to the change in
Ppl between the start and the end of an occluded inspiratory effort
(
Ppl) (18), and because the total respiratory system recoil pressure
is PEEPi,st, an alternative way to write Equation 1 is: Ppl max = PEEPi,st + MIP (Equation 2). A rearrangement of Equation 2 produces a third equation: PEEPi,st = MIP
Ppl max (Equation 3),
given that MIP and Ppl max are both negative values. In summary,
from a theoretical point of view, it is possible to assess PEEPi,st during spontaneous breathing by the difference between MIP and Ppl
max. In the five patients who had expiratory muscle activity, PEEPi,st
measured as MIP
Ppl max was corrected with the same method
used in the measurement of PEEPi,dyn (9), this correction amounting
to 2.0 ± 1.6 cm H2O on average.
Experimental Procedure and Study Design
Patients were studied in a semirecumbent position throughout the
procedure. Following topical anesthesia (10% xylocaine spray), the
two balloon-tipped catheters were positioned through the nose into
the stomach and the esophagus as previously described (5, 9). The occlusion test was satisfactory in every instance, the average
Ppl/
Pao
being 1.015 ± 0.045 (range, 0.938 to 1.078).
Once the patient appeared to be accustomed to the experimental setting and measuring equipment, the ventilatory mode and setting prescribed by the caring physician was discontinued. The measuring equipment was composed of a connector for airway pressure sampling, the pneumotachometer, and a T-tube circuit with the same resistive characteristics as the ventilator tubings. The inspiratory and expiratory lines of the T-tube were separated by a Hans-Rudolph one-way valve to allow either inspiratory or expiratory line occlusions by means of two pneumatic shutters (Figure 1). Both shutters were used to perform the occlusion test. The assessment of maximum inspiratory pressures was performed by occluding only the inspiratory shutter.
After the spontaneous breathing pattern was recorded, the inspiratory limb of the circuit was occluded during a free expiration through the expiratory limb of the circuit. Therefore, the inspiratory occluded effort was performed at end-expiration. At this point, patients were encouraged to perform maximum inspiratory muscle contraction against the occluded airway (Mueller maneuver) (16). The occlusion was released after at least 2 s, and the maneuver was repeated after 30 s of spontaneous breathing until two values of MIP were within 5% of each other. A maximum of 5 maneuvers were sufficient in all patients.
Validation of PEEPi,st measurement. At the end of the above described procedure, the patients were sedated (propofol 2.5 mg/kg, intravenously) and ventilated in the control mode with the settings prescribed by the primary physician. On average, the settings of the ventilator were as follows: VT = 0.54 ± 0.09 L; f = 16.0 ± 3.7 breaths/ min; inspiratory flow = 0.83 ± 0.21 L/s; TI/Ttot = 0.23 ± 0.05. After a few minutes of mechanical ventilation to ensure for respiratory muscle relaxation, an end-inspiratory occlusion was performed using the appropriate "hold" button of the ventilator. After a plateau was reached in Pao, the occlusion was released, and a series of brief interruptions was performed throughout the following relaxed expiration by manual occlusion of the expiratory part of the ventilator, until expiratory flow became nil, according to the technique described in detail elsewhere (19, 20). After each airway occlusion, airflow fell to zero, lung volume did not change, and Pao became positive, exhibiting in all instances a plateau reflecting the elastic recoil pressure of the respiratory system at various lung volumes during lung deflation. We checked that the PL value at the plateau during the end-inspiratory occlusion was higher than that at the plateau recorded during the Mueller maneuver performed in the first part of the procedure. The quasi-static pressure-volume (P/V) relationship of the respiratory system, as well as of its lung and chest wall compartments during lung deflation were determined by plotting the plateau in postinterruption Pao, PL, and Ppl pressures, respectively, against the corresponding volume above relaxation volume (Vr, i.e., the end-expiratory position observed when expiratory flow and Pao became nil) (19, 21).
The PL/V and Pao/V curves were used to compute the EELV during spontaneous breathing and the reference PEEPi,st (PEEPi,st-Ref) at that volume, respectively. To this purpose, the absolute value of PL
at the plateau during the Mueller maneuver (Figures 2 and 3) were
considered to be the result of the static lung elastic recoil at the EELV
of the breath preceding the occluded inspiratory effort. The mirror
image of PL plateau (
PL plateau) was superimposed on the mirror
image of the passive PL/V curve (
PL/V) obtained during mechanical
ventilation (Figure 4, left). The intercept between
PL plateau and
PL/V curve identified the EELV during spontaneous breathing referenced to Vr. The intercept between EELV referenced to Vr and the
Pao/V curve gave the value of PEEPi,st-Ref for that particular lung
volume (22) (Figure 4, right).
|
Data Analysis
Using a personal computer (Compaq 386 equipped with 80387 math coprocessor; Compaq, Houston, TX, and with DT2801/A A/D board; Data Translation, Marlboro, MA) all signals were analog-to-digital converted, displayed on-line through the procedure, and stored on 3.5-inch floppy diskettes at a sampling rate of 100 Hz. The data were collected and the tracings analyzed using appropriate software (Labdat and Anadat; RHT-InfoDat Inc., Montreal, PQ, Canada). Data are reported as mean ± SD unless otherwise specified. Simple linear regression techniques (23) were employed to calculate the correlation between PEEPi,st and PEEPi,st-Ref, and between PEEPi,dyn and PEEPi,st. Comparisons between MIP and Ppl max and between PEEPi,dyn and PEEPi,st were performed by means of the Wilcoxon signed rank test (24). A value of p < 0.05 was taken as the criterion for accepting statistically significant difference. PEEPi,st and PEEPi,st-Ref were also compared by means of the Bland and Altman analysis (25).
| |
RESULTS |
|---|
|
|
|---|
Table 2 shows mean values of breathing pattern and lung mechanics during spontaneous breathing. In general, VT was small and the frequency of breathing high. Lung mechanics was abnormal because of low dynamic compliance, high pulmonary resistance, and PEEPi (Table 2).
|
Mean values of Ppl max and MIP are also shown in Table
2. On average Ppl max was significantly greater than MIP (p < 0.05). The difference between Ppl max and MIP, i.e., the
change in PL occurring between the start of Ppl decrease and
the PL plateau value reflected PEEPi,st (Table 2). The
PL occurring during the Mueller maneuver is composed of two parts
(Figure 2): the first part is the change in PL between the start
of Ppl decrease and the point of zero flow corresponding to
PEEPi,dyn; the second is the further increase in PL occurring
during the occluded inspiratory effort after the point of zero
flow to reach the plateau. The PEEPi,dyn/PEEPi,st ratio averaged 0.55 ± 0.15. Individual values are shown in Figure 5.
|
Validation of PEEPi,st. The lung volume at which the end-inspiratory occlusion was performed during mechanical ventilation averaged 1.045 ± 0.342 L above Vr. The EELV of the
breath preceding the Mueller maneuver averaged 0.772 ± 0.289 L above Vr. On average, PEEPi,st-Ref computed on the
passive Pao/V curve for that lung volume amounted to 13.1 ± 3.0 cm H2O. The relationship between PEEPi,st measured as
MIP
Ppl max and Peepi,st-Ref computed from the quasi-static expiratory Pao/V curve is shown in Figure 6. All the individual data essentially lie on the identity line. The strong
agreement between individual values of PEEPi,st and PEEPi,st-Ref is confirmed by the Bland and Altman analysis
shown in Figure 7.
|
|
| |
DISCUSSION |
|---|
|
|
|---|
The data of this study show that PEEPi,st can be measured in
actively breathing patients from the difference between MIP
and Ppl max during the maximum occluded inspiratory effort,
i.e., the Mueller maneuver commonly used to assess the inspiratory muscle strength. Clearly, an esophageal balloon is
needed to measure changes in Ppl and the occlusion test must
be satisfactory, which is the case when an equal decrease occurs in Pao and Ppl during respiratory efforts against occluded
airway (14) because the lung volume does not change, and
hence, PL (i.e., Pao
Ppl) remains constant throughout the
occlusion (14). Similarly, PL remains constant during the
Mueller maneuver (Figures 2 and 3) (14). In normal subjects
in whom the EELV is coincident with Vr, a simultaneous decrease in Pao and Ppl indicates the onset of the inspiratory effort against occluded airway. By contrast, in patients with severe COPD and dynamic pulmonary hyperinflation, the decrease in Pao at the beginning of the occluded effort is preceded by the decrease in Ppl (Figures 2 and 3). The value of
Ppl in that interval has been interpreted as the amount of
pressure needed to counterbalance PEEPi,dyn (4). Because
Pao does not change in that interval, the change in PL also
provides measurement of PEEPi,dyn. This interpretation is
valid provided that expiratory muscles are relaxed and airway
opening is not occluded at end-expiration. This was the case in
three of our eight patients in the present study (Figure 2),
whereas in the other five patients expiratory muscles were active during expiration such that correction for expiratory muscle relaxation at the beginning of the inspiratory effort was
needed (5, 9, 26). In all eight patients, after the initial change
due to PEEPi,dyn, PL exhibited an additional modification
before reaching the plateau which lasted throughout the occlusion. The fact that this additional change was present in
patients regardless of expiratory muscle activity strongly supports that it was not the result of abdominal muscle relaxation. We excluded an artifact due to bad positioning of the esophageal balloon not only because the occlusion test was satisfactory in all the patients, but also because, after the slight initial
change following the offset of PEEPi,dyn, PL exhibited a plateau throughout the occluded inspiratory effort in all patients
though both Pao and Ppl continued to decrease to the point of
minimum value, namely MIP and Ppl max, respectively. According to Equation 3, the difference between MIP and Ppl
max, and hence the
PL from the start of the inspiratory effort
to PL plateau represented PEEPi,st. This was confirmed by
the identity of PEEPi,st and the reference PEEPi,st measured
on the quasi-static P/V curve. Therefore we interpreted the
difference in PL between the end of PEEPi,dyn period and the plateau as due to stress adaptation phenomena of the lung tissues plus equilibration among lung units with different time
constants (
) during the occlusion. Our interpretation is in
agreement with the results of Petrof and colleagues (10) and
Maltais and colleagues (11) who have shown that in COPD
patients PEEPi,dyn is smaller than the true end-expiratory
elastic recoil, i.e., PEEPi,st.
The measurement of PEEPi,st as MIP
Ppl max could be
affected by substantial errors if correspondence between
Ppl
and
Pao, quantified by the
Ppl/
Pao ratio obtained during
the occlusion test (14), is lacking. Of course, the error would
increase with increasing Ppl max. It follows that Ppl max
should be corrected for
Ppl/
Pao ratio before subtracting
Ppl max from MIP to obtain PEEPi,st. The correction can be
accomplished by dividing the measured Ppl max by
Ppl/
Pao
found during the occlusion test.
A recent study by Yan and colleagues in normal subjects in whom dynamic hyperinflation has been induced by breathing through a Starling resistor (27) suggested that PEEPi,st could be underestimated because of persistent postinspiratory muscle activity. This could be the case in asthmatics (28, 29), but it is unlikely to be the case in patients with COPD in whom several studies have shown that the postinspiratory activity decays early in expiration (30, 31). In our study, the close agreement found between PEEPi,st measured during the Mueller maneuver and PEEPi,st-Ref measured from the quasi-static P/V curve (Figures 6 and 7) supports our conclusion. The Mueller maneuver to measure PEEPi does still require the invasive procedure of the esophageal and gastric catheters, but it helps to overcome the methodological problems of the end-expiratory occlusion maneuver in actively breathing patients. In fact, active patients react to the end-expiratory airway occlusion such that the continuous inspiratory and expiratory muscle activity may prevent the plateau in the pressure at the airway opening, thus negating noninvasive measurement of PEEPi,st without sedation. Moreover, severe COPD patients often exhibit an increased respiratory frequency (16) and breath-by-breath variability of the respiratory cycle duration that can make it impossible for the operator to select the correct time for airway occlusion. In contrast, during the Mueller maneuver with the technique used in this study, only the inspiratory side of the circuit is occluded, such that the patient can perform a complete expiration before the occluded inspiratory effort.
As previously mentioned, a possible limitation of the method
we propose in the present study is the need to insert both gastric and esophageal balloons, a procedure that can be uncomfortable for the patient (32). However, this limitation is common to the method of end-expiratory occlusion to exclude or
correct for expiratory muscle activity. On the other hand, during the Mueller maneuver, all the following measurements can
be obtained: an estimation of the pressure-generating capacity
of the inspiratory muscles; the evaluation of PEEPi,st; both
PEEPi,dyn and PEEPi,st measurements within the same breath,
i.e., at the same end-expiratory volume, provided that appropriate corrections are done for effects of
Ppl/
Pao ratio and
expiratory muscle activity on the measurement.
The possibility of assessing PEEPi,st during spontaneous breathing provides some clinical advantages. First, PEEPi,st represents a more accurate estimate of the true elastic lung recoil than PEEPi,dyn, and some definition of lung inhomogeneities through the PEEPi,dyn/PEEPi,st ratio (Figure 5) (11). Second, and more important, PEEPi,st could provide a better guideline to set the optimal level of PEEP and CPAP in ventilator-dependent patients with flow limitation (11). Hence, our technique could prove very useful in clinical practice.
In summary, our study has shown that the occluded inspiratory effort maneuver (Mueller maneuver) is a reliable procedure for measuring PEEPi,st in patients with severe COPD during spontaneous breathing. This method provides advantages with respect to the traditional end-expiratory occlusion maneuver, and may prove suitable for both diagnostic and therapeutic purposes.
| |
Footnotes |
|---|
Correspondence and requests for reprints should be addressed to Andrea Purro, M.D., Fondazione Salvatore Maugeri, Clinica del Lavoro e della Riabilitazione, IRCCS, Instituto di Riabilitazione di Veruno, Via per Revislate n.13, 28010 Veruno (NO), Italy.
(Received in original form February 4, 1997 and in revised form November 5, 1997).
Acknowledgments: The authors thank the medical and nursing staff of the ICU for their skill and kind cooperation. They would also like to thank Ms. R. Allpress for her help in the preparation of this manuscript.
Supported by the S. Maugeri Foundation, Pavia, Italy.
| |
References |
|---|
|
|
|---|
1. Similowski, T., J. Milic-Emili, and J.-P. Derenne. 1996. Respiratory mechanics during acute respiratory failure of chronic obstructive pulmonary disease. In J.-P. Derenne, W. A. Whitelaw, and T. Similowski, editors. Acute Respiratory Failure in Chronic Obstructive Pulmonary Disease, 1st ed. Marcel Dekker, New York. 23-46.
2. Rossi, A., G. Polese, and J. Milic-Emili. 1996. Mechanical ventilation in the passive patient: theory and clinical investigation. In J.-P. Derenne, W. A. Whitelaw, and T. Similowski, editors. Acute Respiratory Failure in Chronic Obstructive Pulmonary Disease, 1st ed. Marcel Dekker, New York. 709-746.
3. Pride, N. B., and J. Milic-Emili. 1995. Lung mechanics. In P. Calverley and N. Pride, editors. Chronic Obstructive Pulmonary Disease, 1st ed. Chapman & Hall Medical, London. 161-184.
4. Rossi, A., G. Polese, G. Brandi, and G. Conti. 1995. Intrinsic positive end-expiratory pressure (PEEPi). Intensive Care Med. 21: 522-536 [Medline].
5. Appendini, L., A. Purro, A. Patessio, S. Zanaboni, M. Carone, E. Spada, C. F. Donner, and A. Rossi. 1996. Partitioning of inspiratory muscle workload and pressure assistance in ventilator-dependent COPD patients. Am. J. Respir. Crit. Care Med. 154: 1301-1309 [Abstract].
6. Ranieri, V. M., S. Grasso, L. Mascia, S. Martino, T. Fiore, A. Brienza, and R. Giuliani. 1997. Effects of proportional assist ventilation on inspiratory muscle effort in patients with chronic obstructive pulmonary disease and acute respiratory failure. Anesthesiology 86: 79-91 [Medline].
7. Gottfried, S. B. 1995. The role of PEEP or CPAP in the mechanically ventilated COPD patient. In C. Roussos, editor. The Thorax, Part C, 2nd ed. Marcel Dekker, New York. 2471-2500.
8. Decramer, M.. 1997. Hyperinflation and respiratory muscle interaction. Eur. Respir. J. 10: 934-941 [Abstract].
9. Appendini, L., A. Patessio, S. Zanaboni, M. Carone, B. Gukov, C. F. Donner, and A. Rossi. 1994. Physiologic effects of positive end-expiratory pressure and mask pressure support during exacerbation of chronic obstructive pulmonary disease. Am. J. Respir. Crit. Care Med. 149: 1069-1076 [Abstract].
10. Petrof, B. J., M. Legaré, P. Goldberg, J. Milic-Emili, and S. B. Gottfried. 1990. Continuous positive airway pressure reduced work of breathing and dyspnea during weaning from mechanical ventilation in severe chronic obstructive pulmonary disease. Am. Rev. Respir. Dis. 141: 281-289 [Medline].
11. Maltais, F., H. Reissman, P. Navalesi, P. Hernandez, A. Gursahaney, V. M. Ranieri, M. Sovilj, and S. B. Gottfried. 1994. Comparison of static and dynamic measurements of intrinsic PEEP in mechanically ventilated patients. Am. J. Respir. Crit. Care Med. 150: 1318-1324 [Abstract].
12. Zakynthinos, S. G., T. Vassilakopoulos, E. Zakynthinos, and C. Roussos. 1997. Accurate measurement of intrinsic positive end-expiratory pressure: how to detect and correct for expiratory muscle activity. Eur. Respir. J. 10: 522-529 [Abstract].
13. Agostoni, E., and R. E. Hyatt. 1986. Static behavior of the respiratory system. In P. Fishman, editor. Handbook of Physiology: The Respiratory System, Vol. 3. American Physiologic Society, Bethesda, MD. 113-144.
14. Baydur, A., P. K. Behrakis, W. A. Zin, M. J. Jaeger, and J. Milic-Emili. 1982. A simple method for assessing the validity of the esophageal balloon technique. Am. Rev. Respir. Dis. 126: 788-791 [Medline].
15. Rossi, A., R. Brandolese, J. Milic-Emili, and S. B. Gottfried. 1990. The role of PEEP in patients with chronic obstructive pulmonary disease during assisted ventilation. Eur. Respir. J. 3: 818-822 [Abstract].
16. Tobin, M. J., F. Laghi, and J. M. Walsh. 1994. Monitoring of respiratory neuromuscular function. In M. J. Tobin, editor. Principles and Practice of Mechanical Ventilation, 1st ed. McGraw-Hill, New York. 945- 966.
17. Rochester, D. F., N. S. Arona, N. M. T. Braun, and S. K. Goldberg. 1979. The respiratory muscle in chronic obstructive pulmonary disease (COPD). Bull. Eur. Physiopathol. Respir. 15: 951-975 [Medline].
18. Derenne, J., Ph, P. T. Maklem, and C. Roussos. 1978. The respiratory muscles: mechanics, control, and pathophysiology. Am. Rev. Respir. Dis. 118: 119-133 [Medline].
19. Gottfried, S. B., A. Rossi, B. D. Higgs, P. M. A. Calverley, L. Zocchi, C. Bozic, and J. Milic-Emili. 1985. Noninvasive determination of respiratory system mechanics during mechanical ventilation for acute respiratory failure. Am. Rev. Respir. Dis. 131: 414-420 [Medline].
20. Broseghini, C., R. Brandolese, R. Poggi, G. Polese, E. Manzin, J. Milic-Emili, and A. Rossi. 1988. Respiratory mechanics during the first day of mechanical ventilation in patients with pulmonary edema and chronic airway obstruction. Am. Rev. Respir. Dis. 138: 355-361 [Medline].
21. Ranieri, V. M., R. Giuliani, L. Mascia, S. Grasso, V. Petruzzelli, F. Bruno, T. Fiore, and A. Brienza. 1996. Chest wall and lung contribution to the elastic properties of the respiratory system in patients with chronic obstructive pulmonary disease. Eur. Respir. J. 9: 1232-1239 [Abstract].
22.
Polese, G.,
A. Rossi,
L. Appendini,
G. Brandi,
J. H. T. Bates, and
R. Brandolese.
1991.
Partitioning of respiratory mechanics in mechanically ventilated patients.
J. Appl. Physiol.
71:
2425-2433
23. Snedecor, G. W., and W. G. Cochran. 1980. Statistical Methods, 7th ed. Iowa State University Press, Ames. 334-364.
24. Colton, T. 1974. Statistics in Medicine, 1st ed. Little, Brown, Boston. 219-227.
25. Bland, J. M., and D. G. Altman. 1986. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1: 307-310 [Medline].
26. Ninane, V.. 1997. "Intrinsic" PEEP (PEEPi): role of expiratory muscles. Eur. Respir. J. 10: 516-518 [Medline].
27. Yan, S., B. Kayser, M. Tobiasz, and P. Sliwinski. 1996. Comparison of static and dynamic intrinsic positive end-expiratory pressure using the Campbell diagram. Am. J. Respir. Crit. Care Med. 154: 938-944 [Abstract].
28. Martin, J. G., E. Powell, S. Shore, J. Emrich, and L. A. Engel. 1980. The role of respiratory muscles in the hyperinflation of bronchial asthma. Am. Rev. Respir. Dis. 121: 441-447 [Medline].
29.
Muller, N.,
A. C. Byran, and
N. Zamel.
1980.
Tonic inspiratory muscle
activity as a cause of hyperinflation in histamine-induced asthma.
J.
Appl. Physiol.
49:
869-874
30. Fleury, B., C. Murciano, C. Talamo, M. Aubier, R. Pariente, and J. Milic-Emili. 1985. Work of breathing in patients with chronic obstructive pulmonary disease in acute respiratory failure. Am. Rev. Respir. Dis. 131: 822-827 [Medline].
31. Haluszka, J., D. A. Chartrand, A. E. Grassino, and J. Milic-Emili. 1990. Intrinsic PEEP and arterial PCO2 in stable patients with chronic obstructive pulmonary disease. Am. Rev. Respir. Dis. 141: 1194-1197 [Medline].
32. Similowski, T., and J. P. H. Derenne. 1994. Inspiratory muscle testing in stable COPD patients. Eur. Respir. J. 7: 1871-1876 [Abstract].
This article has been cited by other articles:
![]() |
M. I. Polkey Surgical procedures in emphysema: any impact on dynamic hyperinflation? Eur. Respir. Rev., December 1, 2006; 15(100): 96 - 99. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. HAYOT, M. RAMONATXO, S. MATECKI, J. MILIC-EMILI, and C. PREFAUT Noninvasive Assessment of Inspiratory Muscle Function during Exercise Am. J. Respir. Crit. Care Med., December 1, 2000; 162(6): 2201 - 2207. [Abstract] [Full Text] |
||||
![]() |
A. PURRO, L. APPENDINI, A. DE GAETANO, M. GUDJONSDOTTIR, C. F. DONNER, and A. ROSSI Physiologic Determinants of Ventilator Dependence in Long-term Mechanically Ventilated Patients Am. J. Respir. Crit. Care Med., April 1, 2000; 161(4): 1115 - 1123. [Abstract] [Full Text] |
||||
![]() |
S. G. ZAKYNTHINOS, T. VASSILAKOPOULOS, E. ZAKYNTHINOS, C. ROUSSOS, and G. E. TZELEPIS Correcting Static Intrinsic Positive End-expiratory Pressure for Expiratory Muscle Contraction . Validation of a New Method Am. J. Respir. Crit. Care Med., September 1, 1999; 160(3): 785 - 790. [Abstract] [Full Text] |
||||
![]() |
L. APPENDINI, A. PURRO, M. GUDJONSDOTTIR, P. BADERNA, A. PATESSIO, S. ZANABONI, C. F. DONNER, and A. ROSSI Physiologic Response of Ventilator-dependent Patients with Chronic Obstructive Pulmonary Disease to Proportional Assist Ventilation and Continuous Positive Airway Pressure Am. J. Respir. Crit. Care Med., May 1, 1999; 159(5): 1510 - 1517. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Proc. Am. Thorac. Soc. | Am. J. Respir. Cell Mol. Biol. |