|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| |
ABSTRACT |
|---|
|
|
|---|
During dynamic hyperinflation, the ventilatory pump is facing increased demand because it must
overcome the intrinsic positive end-expiratory pressure (PEEPi) and decreased capacity since it must
operate at a dynamically increased end-expiratory lung volume (EELV). The aim of this study was to
evaluate the relative pressure contribution by the diaphragm and inspiratory rib cage muscles (RCMs) during dynamic hyperinflation. In six healthy subjects, dynamic hyperinflation was induced
by limiting expiratory flow. The global inspiratory muscle pressure (
Pmus,i) and transdiaphragmatic pressure (
Pdi) were partitioned into the portion used to overcome PEEPi and the portion used to inflate the respiratory system. The
Pdi/
Pmus,i ratio was used to estimate the pressure contribution of RCMs relative to that of the diaphragm. Our results suggest that (1) with increasing severity of dynamic hyperinflation, there is a significant increase in the inspiratory pressure contribution of RCMs
relative to that of the diaphragm for inflating the respiratory system; (2) during dynamic hyperinflation, especially at high EELV, the major pressure contribution of the diaphragm is to overcome the
PEEPi-imposed inspiratory threshold load, whereas the inspiratory pressure needed for the subsequent task of inflating the respiratory system is largely contributed by RCMs. This arrangement is
consistent with the change in mechanical advantages of RCMs and the diaphragm during the development of dynamic hyperinflation.
| |
INTRODUCTION |
|---|
|
|
|---|
Dynamic hyperinflation is common in patients with chronic obstructive pulmonary disease (COPD) during acute exacerbations requiring mechanical ventilation (1, 2), but it has also been reported in patients with stable COPD during resting breathing (3, 4). During dynamic hyperinflation, end-expiratory lung volume (EELV) is above the relaxation volume, resulting in a positive end-expiratory pressure (PEEPi) that has to be overcome by the inspiratory muscles before inspiratory flow can start (5). Dynamic hyperinflation thus decreases the capacity of and increases the demand to the inspiratory muscles.
Most of the previous studies of dynamic hyperinflation reported methods to measure PEEPi and/or discussed various factors likely affecting the reliability of these measurements (2, 6). How the different inspiratory muscles contribute to breathing during dynamic hyperinflation has not been investigated in detail. In particular, how the diaphragm, the major inspiratory muscle, contributes to overcoming the PEEPi-imposed inspiratory threshold load and subsequently to inflating the respiratory system has not been reported before. Because with increasing EELV, the diaphragm progressively shortens and its ability to generate pressure is gradually impaired (9, 10), we hypothesized that during dynamic hyperinflation, although the diaphragm may still play an important part in generating pressure to overcome the PEEPi-induced inspiratory threshold load, there is a need for a proportionately greater increase in pressure contribution of the inspiratory rib cage muscles (RCMs) over that of the diaphragm to overcome PEEPi and to inflate the respiratory system. Accordingly, the major purposes of the present study were to measure the pressure contribution of the diaphragm during dynamic hyperinflation and to estimate the pressure contribution of RCMs relative to that of the diaphragm to overcome the PEEPi-induced inspiratory threshold load and to inflate the respiratory system. For these purposes, we induced dynamic hyperinflation in healthy subjects by limiting expiratory flow by a Starling resistor (8, 11, 12) and measured inspiratory muscle pressure responses. We interpreted our experimental findings on the basis of different changes in mechanical advantages of the RCMs and diaphragm during the development of dynamic hyperinflation.
| |
METHODS |
|---|
|
|
|---|
Subjects
Six healthy subjects, all belonging to our laboratory staff and therefore familiar with respiratory maneuvers such as respiratory muscle relaxation, volunteered. Only one knew the precise purpose of the study. The study was approved by our institutional ethics committee, and informed consent was obtained from all the subjects.
Experimental Setup
The experimental setup was similar to that in a previous study (8). The experiment was performed with the subject seated in a high-back armchair, breathing through a mouthpiece, and wearing a noseclip. The mouthpiece was connected via a pneumotachograph to a two-way nonrebreathing valve (Type 2600; Hans-Rudolph, St. Louis, MO). The latter was connected to a bag-in-box plethysmograph in a closed circuit. Respiratory flow was measured by the pneumotachograph (Fleisch No. 2; Fleisch, Lausanne, Switzerland) attached to a differential pressure transducer (Validyne Corp., Northridge, CA). Changes in EELV were measured by the bag-in-box system. Respiratory timing signals, including breathing frequency (f), inspiratory and expiratory time (TI and TE), were calculated from the flow signal. Esophageal and gastric pressures (Pes and Pga) were measured by balloon-tipped catheters positioned in the lower esophagus and stomach and connected to differential pressure transducers (Validyne). Transdiaphragmatic pressure (Pdi) was obtained by subtraction of Pes from Pga. End-expiratory carbon dioxide partial pressure (PETCO2) was measured at the mouthpiece by a CO2 analyzer (Ametek CD-3A; Sunnyvale, CA).
In order to limit expiratory flow and induce dynamic hyperinflation, a Starling resistor (8, 11, 12) was connected to the expiratory line of the breathing circuit by a three-way stopcock type valve (Hans Rudolph 2120C). The Starling resistor consisted of a compressible rubber tube contained in a Plexiglas® chamber. The chamber was pressurized by connecting it to the expiratory side of the valve system near the mouthpiece by a side tube. Expiratory mouth pressure was thus relayed to the chamber, which compressed the rubber tube, rendering flow independent of expiratory efforts at ~ 0.6 L/s.
Protocol
An individual static chest wall pressure-volume relationship was obtained by repeated respiratory muscle relaxation maneuvers during passive deflation from TLC to FRC against a high nonlinear resistance. Three successful relaxations were performed by each subject. A relaxation was judged successful by the reproducibility of the chest wall relaxation pressure-volume curves, a smooth decline of Pes and Pga from TLC to FRC, and a zero Pdi (13). Static pressure-volume relationships of the chest wall were constructed by plotting lung volume (VL) against Pes with the VL at the equilibrium volume (FRC) scaled to zero.
After a brief period of adjusting to breathing on the non-flow-limited breathing circuit, quiet breathing data were obtained for 5 min. Then, with the subjects remaining on the mouthpiece, the expiratory route was opened to the Starling resistor. To induce dynamic hyperinflation, the subjects were subsequently instructed to increase breathing frequency from their natural frequency to 20, 25, 30, 35, and 40 breaths/min using a metronome. No other instructions were given. Specifically, the subjects were left free to choose their tidal volume (VT), respiratory timing, and the preference of their respiratory muscle use. At each frequency, 6 to 10 breaths were recorded for data analysis.
Data Recording and Analysis
Respiratory flow, VL, Pes, Pga, and PETCO2 were preamplified by an eight-channel strip-chart recorder (HP 7758B; Hewlett-Packard, Waltham, MA) passed through a 12-bit analog-to-digital converter and recorded by a computer. Data analysis was performed on averaged breaths (five to eight breaths).
With dynamic hyperinflation, inspiratory muscle effort must start
prior to the start of inspiratory flow to overcome PEEPi. A sudden
decrease in Pes just before inspiratory flow has been used to measure
dynamic PEEPi (2). However, since end-expiratory Pes baseline is
subject to influences from activities of various respiratory muscles (2,
6), we used the individual chest wall pressure-volume curve as the
reference for measuring global inspiratory muscle pressure contribution (
Pmus,i) for a given lung volume. The schema for Pmus,i measurement is shown in Figure 1. A clockwise dynamic VL-Pes loop (thin
line) during breathing is superimposed on the static pressure-volume
curve of the chest wall (thick line). Points A and B represent the beginning and end of inspiratory flow, respectively. EELV is increased
because of dynamic hyperinflation as indicated by the level of AC.
The
Pmus,i required to initiate the inspiratory flow (thus to overcome the PEEPi-imposed inspiratory threshold load) and to accomplish the whole inspiration was measured as the Pes value at the beginning and the end of inspiratory flow, respectively, relative to the Pes value on the chest wall pressure-volume curve for a given end-
expiratory and end-inspiratory VL (the pressure differences given by
AC and BD). The sign of
Pmus,i measured this way is always negative. The
Pmus,i to overcome the respiratory system elastance during inspiratory flow was calculated as BD-AC. The pressure contribution of the diaphragm to initiate inspiratory flow and to accomplish
the whole inspiration was measured as the Pdi value at the beginning
and end of inspiratory flow, respectively, relative to the Pdi baseline
during resting breathing, which was assumed to be zero when the subject sits relaxed in an upright position (13). The ratio
Pdi/
Pmus,i
was used to evaluate the pressure contribution of RCMs relative to
that of the diaphragm for a given condition.
|
The data are presented as individual values or mean ± SD. Paired t tests and least-squares linear regressions were used for statistical analysis when appropriate. A p value less than 0.05 was considered to be statistically significant.
| |
RESULTS |
|---|
|
|
|---|
All the subjects developed dynamic hyperinflation during expiratory Starling resistor breathing. The individual maximal
increases in EELV ranged from 0.87 to 2.32 L. The breathing
pattern parameters are given in Table 1 for quiet unloaded
breathing and for Starling resistor breathing at the highest
EELV. All subjects responded similarly to the Starling resistor breathing: VT remained relatively constant; TI and TE were
significantly shortened; minute ventilation (
E) was significantly increased (because of the imposed increase in f). PETCO2
significantly decreased (because of increased
E), from 38 ± 5 mm Hg during quiet unloaded breathing to 28 ± 4 mm Hg
during Starling resistor breathing at the highest EELV.
|
Pmus,i and
Pdi to overcome the PEEPi-imposed inspiratory threshold load and to initiate the inspiratory flow during
dynamic hyperinflation are shown in Figure 2. The negativity
of
Pmus,i linearly increased with increasing EELV in each
subject;
Pdi also increased, but with considerable variability
(see Subjects 3, 4, and 5).
Pmus,i and
Pdi to inflate the respiratory system (between the beginning and end of inspiratory flow) during dynamic hyperinflation are shown in Figure
3. The magnitude of
Pmus,i increased with increasing EELV
in four subjects (Subjects 2, 3, 5, and 6) and was relatively constant in the remaining two subjects. The
Pdi showed a tendency to decrease with increasing EELV in all but one subject
(Subject 5).
|
|
The
Pdi/
Pmus,i ratio is shown in Figure 4 before the
start of inspiratory flow (upper panels) and between the beginning and end of inspiratory flow (lower panels) during dynamic hyperinflation. With increasing EELV, the negativity of
Pdi/
Pmus,i before the start of inspiratory flow significantly
decreased in four subjects (Subjects 1 to 4) (p < 0.05), remained unchanged in one subject (Subject 5), and significantly
increased in one subject (Subject 6) (p < 0.05); the negativity
of the same ratio between the beginning and end of inspiratory flow significantly decreased in all the subjects (p < 0.05).
The individual slopes of
Pdi/
Pmus,i ratio against
EELV
relationships before inspiratory flow as well as between the
beginning and end of inspiratory flow are shown in Table 2.
The average slope before inspiratory flow was 0.23 ± 0.60, which was not significantly different from zero, whereas the
slope during inspiratory flow (1.04 ± 0.28) was not only significantly greater than zero but also significantly greater than the
former (p < 0.05).
|
|
In Figure 5, averaged
Pmus,i and
Pdi before the start of
inspiratory flow are expressed as a percentage of
Pmus,i and
Pdi as developed for the whole inspiration (from the beginning of inspiratory effort to the end of inspiratory flow). With
increasing EELV, about half of the total
Pmus,i was developed before inspiratory flow to overcome the PEEPi-imposed
inspiratory threshold load, whereas 75% of the total
Pdi was
developed for the same purpose (p < 0.01).
|
| |
DISCUSSION |
|---|
|
|
|---|
In the present study, we induced dynamic hyperinflation in
healthy subjects by expiration through a Starling resistor and measured
Pdi and
Pmus,i to assess the pressure contributions of the diaphragm and the overall inspiratory muscles during
breathing. We partitioned
Pdi and
Pmus,i into the contributions to overcome the PEEPi-imposed inspiratory threshold
load and to subsequently inflate the respiratory system. We further used the
Pdi/
Pmus,i ratio to evaluate the pressure contribution of RCMs relative to that of the diaphragm. We found
that before the start of inspiratory flow, the magnitude of both
Pmus,i and
Pdi increased with increasing EELV with no
consistent change in the
Pdi/
Pmus,i ratio as a function of
EELV, whereas between the beginning and the end of the inspiratory flow, there was a significant decrease in the negativity of the
Pdi/
Pmus,i ratio with increasing EELV. We interpret these results as follows: (1) with increasing magnitude of
dynamic hyperinflation, there is a greater increase in the inspiratory pressure contribution of RCMs relative to that of the
diaphragm for inflating the respiratory system than for overcoming PEEPi; (2) during dynamic hyperinflation, especially at
high EELV, the major pressure contribution of the diaphragm
is to overcome the PEEPi-imposed inspiratory threshold load,
whereas the inspiratory pressure needed for the subsequent
task of inflating the respiratory system is largely contributed
by RCMs.
Technical Considerations
In this study, we used
Pmus,i to assess the global inspiratory
muscle pressure contribution before the start of and between the beginning and end of inspiratory flow. The former is similar to, but more accurate than, dynamic PEEPi. Dynamic
PEEPi is determined as an abrupt decrease in Pes relative to
its immediate preceding baseline before inspiratory flow starts
(3, 4). However, expiratory muscle (2, 6) or tonic inspiratory muscle activities (8) during dynamic hyperinflation elevate or lower the expiratory Pes baseline, respectively, and
therefore dynamic PEEPi may overestimate or underestimate
the true inspiratory muscle pressure contributions. Our recent
work (8) using the Campbell diagram to evaluate PEEPi demonstrated that it is difficult to reliably estimate the inspiratory
muscle effort to overcome PEEPi without information of the
relaxation pressure-volume relationship of the chest wall. This
is because the static pressure-volume curve of the chest wall
describes the relationship between lung volume and pleural pressure when all respiratory muscles are relaxed and any departure from this curve requires respiratory muscle activity
(13). In the present study, we found that in all subjects, the
end-expiratory Pes fell to the right of the static pressure-volume relationship of the chest wall under most or all of the occasions during Starling resistor breathing, indicating expiratory muscle recruitment. The average maximal deviation was
10.3 ± 3.1 cm H2O. Hence, using the individual static pressure-volume curve of the chest wall as reference, our measurement of
Pmus,i for a given end-expiratory or end-inspiratory
lung volume presumably gives a reliable measurement of the
global inspiratory muscle pressure contribution under each condition.
As a modification of the original concept of Macklem and
coworkers (14), more recent investigators have employed the
Pga/
Pes ratio to partition the relative inspiratory muscle
pressure contributions (15). We did not use this method in
the present study. Instead, we chose to use the
Pdi/
Pmus,i
ratio for the same purpose because a change in
Pdi/
Pmus,i
can be interpreted similarly to that in
Pga/
Pes, but
Pmus,i
can be directly linked to PEEPi. On the basis of our definition,
inspiratory
Pmus,i, reflecting global inspiratory muscle pressure contribution, is negative, leading to a negative
Pdi/
Pmus,i ratio when
Pdi > 0. Decreasing the negativity of
Pdi/
Pmus,i ratio requires a decreasing
Pdi for a given
Pmus,i
or an increasing negativity of
Pmus,i for a given
Pdi, and
therefore indicates an increase in pressure contribution from
RCMs relative to that of the diaphragm. When
Pdi is zero,
Pdi/
Pmus,i ratio becomes maximal (zero). Under such a
condition, all the
Pmus,i must be generated by RCMs. In
fact, if the diaphragm is the only muscle contracting such as
during phrenic nerve stimulation (RCMs are relaxed), the calculated
Pdi/
Pmus,i ratio based on previous data (18, 19)
was around
2.5 at FRC, significantly more negative than the
same ratio shown in Figure 4 where both RCMs and diaphragm presumably contributed their pressures to a variable
extent.
Effect of Dynamic Hyperinflation on Inspiratory Muscle Use
Earlier investigators (20) have studied the effect of continuous positive airway pressure (CPAP) on inspiratory muscle responses. Although CPAP induces acute hyperinflation, it
moves the inspiratory volume-pressure relationship to the right
along the pressure axis, so that inspiration is assisted and
PEEPi is not present (23). During dynamic hyperinflation,
generation of inspiratory pressure is required not only for inflating the respiratory system but also for counteracting the inspiratory threshold pressure brought about by PEEPi. This
task has to be accomplished when inspiratory muscles are in a
mechanical disadvantage secondary to the shortening of the
inspiratory muscles with an operating lung volume above the
relaxation volume (24). Hence, dynamic hyperinflation loads the inspiratory pump by reducing the capacity of, as well as by increasing the demand to, the inspiratory muscles. This feature is unique and probably differs from all other types of inspiratory loads. As shown previously, even though both muscle
groups shorten with increasing lung volume, major differences
exist between RCMs and the diaphragm. First, at FRC, the
resting length of the diaphragm is close to its optimal length,
whereas that of the parasternal intercostals is 15% longer than
its optimal length (25). Thus, increasing EELV from FRC
would initially move the parasternal intercostals towards their
optimal length. Second, from FRC to TLC, the diaphragm
shortens by 30 to 40% (10), whereas the scalenes and parasternal intercostals shorten by only 6 and 10%, respectively (26,
27). Third, with increasing lung volume, the two muscular parts of the diaphragm (crural and costal) change from a parallel to an in series configuration, which impairs the function of
the diaphragm as a pressure generator (28). Consequently, at
TLC, the ability of the diaphragm to generate pressure is dramatically impaired (18, 19), but that of the parasternal intercostals is well preserved (29). On this basis, one can expect
that during dynamic hyperinflation, an increased pressure
contribution from RCMs will take over the task of generating
pressure to maintain breathing. This reasoning is consistent with
our observations for the inspiratory pressure contributions to
inflate the respiratory system (between the beginning and end
of inspiratory flow) during dynamic hyperinflation. Indeed,
the amplitude of
Pmus,i increased with increasing EELV in
most subjects, presumably reflecting a greater respiratory system elastance at high volumes, whereas
Pdi remained relatively constant or decreased (Figure 3), leading to a significant
decrease in the negativity of
Pdi/
Pmus,i ratio with increasing EELV in all subjects (Figure 4, lower panels, and Table 2).
By contrast, we did not find a consistent change in
Pdi/
Pmus,i ratio as a function of EELV before the beginning of
inspiratory flow (Figure 4, upper panels, and Table 2). To
overcome the PEEPi-imposed inspiratory threshold load,
Pdi
increased with increasing
Pmus,i nearly proportionately despite a progressive increase in EELV (Figure 2). This suggests
that relative pressure contributions of RCMs and diaphragm
to overcome the PEEPi-imposed inspiratory threshold load
remain relatively constant over different levels of dynamic hyperinflation. Further, the diaphragm spent as much as 75% of
the total
Pdi generated within one breathing cycle to overcome the PEEPi-imposed inspiratory threshold load. This was
significantly more than that from
Pmus,i, which amounted to
50% (Figure 5). We therefore conclude that during dynamic
hyperinflation, the major pressure contribution of the diaphragm is to overcome PEEPi, whereas the inspiratory pressure required to inflate the respiratory system is largely generated by RCMs.
The maintenance of a relatively constant
Pdi/
Pmus,i ratio before the beginning of inspiratory flow may at first appear
surprising because the diaphragm, compared with RCMs, is
more significantly deprived of its mechanical advantage by contracting at high lung volumes, as stated above. Our results would
suggest that to overcome the PEEPi-imposed inspiratory threshold load, the activation of the diaphragm must be progressively increased beyond that of RCMs in order to keep the
Pdi/
Pmus,i ratio relatively constant or to be less affected by
the increasing degree of dynamic hyperinflation. If this were
the case, the question arises why the subjects had a greater decrease in the negativity of this ratio during inspiratory flow as
EELV increased. The answer may be found in the characteristic features of a PEEPi-induced inspiratory threshold load.
During inspiratory resistive and elastic loading, inspiratory flow
starts instantaneously with inspiratory muscle contraction so
that the processes to overcome the load and to inflate the respiratory system occur simultaneously. During inspiratory
threshold loading, the situation is different. The total inspiratory effort can be divided clearly into two phases by the beginning of inspiratory flow, with the inspiratory muscle velocity
of shortening much slower in the first phase before flow starts
(determined by chest wall distortion and gas compression)
than in the second phase during flow. According to the force-velocity relationship, both RCMs and the diaphragm should
be better at generating inspiratory pressure in the first phase
than in the second phase. However, if the amount of shortening of the diaphragm is greater than that of RCMs between
FRC and TLC (10, 26, 27), the velocity of shortening of the diaphragm must be greater than that of RCMs during flow. In
other words, the capacity of the diaphragm as an inspiratory
pressure generator will be more affected during flow. This then
results in most of the pressure contribution of the diaphragm
being developed to overcome the PEEPi-imposed inspiratory threshold load, leaving RCMs as the major inspiratory pressure generator to inflate the respiratory system. Alternatively,
these results may be explained by failure to maintain the increased diaphragmatic drive throughout the whole inspiration
(30) as dynamic hyperinflation develops.
Finally, it needs to be pointed out that the current results were obtained from healthy subjects. Their inspiratory muscles presumably had sufficient reserve for extra demands. In patients with COPD, the diaphragm is weak because of chronic hyperinflation, and the capacity of the diaphragm to increase its pressure output is therefore diminished. With dynamic hyperinflation, the tasks of overcoming the PEEPi-imposed inspiratory threshold load and inflating the respiratory system would consequently both depend largely on RCMs. However, in these patients, the inspiratory pressure contribution from RCMs is already increased during resting breathing to compensate for diaphragmatic weakness (16), which may consequently lead to reduced RCMs reserve. As dynamic hyperinflation requires further recruitment of RCMs, a poor reserve of inspiratory muscle function may lead to reduced ventilation and therefore cause ventilatory failure (31).
| |
Footnotes |
|---|
Correspondence and requests for reprints should be addressed to Dr. S. Yan, Montréal Chest Institute Research Center, 3650 St. Urbain Street, Room 526, Montréal, PQ, H2X 2P4 Canada. E-mail: sheng{at}meakins.lan.mcgill.ca
(Received in original form November 19, 1996 and in revised form April 3, 1997).
Dr. Kayser is the recipient of post-doctoral fellowship from Merck-Frosst.Acknowledgments: Supported by the Montréal Chest Institute, the Association Pulmonaire du Québec, and the T. J. Costello Memorial Research Fund.
| |
References |
|---|
|
|
|---|
1. Rossi, A., S. B. Gottfried, L. Zocchi, B. D. Higgs, S. Lennox, P. M. A. Calverley, P. Begin, A. Grassino, and J. Milic-Emili. 1985. Measurement of static compliance of the total respiratory system in patients with acute respiratory failure during mechanical ventilation: the effect of intrinsic positive end-expiratory pressure. Am. Rev. Respir. Dis 131: 672-677 [Medline].
2. Appendini, L., A. Patessio, S. Zanaboni, M. Carone, B. Gukov, C. F. Donner, and A. Rossi. 1994. Physiological effects of positive end-expiratory pressure and mask pressure support during exacerbations of chronic obstructive pulmonary disease. Am. J. Respir. Crit. Care Med 149: 1069-1076 [Abstract].
3. Haluszka, J., D. A. Chartrand, A. E. Grassino, and J. Milic-Emili. 1990. Intrinsic PEEP and arterial PCO2 in stable patients with chronic obstructive pulmonary disease. Am. Rev. Respir. Dis 141: 1194-1197 [Medline].
4. Vecchio, L. D., G. Polese, R. Possi, and A. Rossi. 1990. "Intrinsic" positive end-expiratory pressure in stable patients with chronic obstructive pulmonary disease. Eur. Respir. J 3: 74-80 [Abstract].
5. Pepe, P. E., and J. J. Marini. 1982. Occult positive end-expiratory pressure in mechanically ventilated patients with airflow obstruction: the auto-PEEP effect. Am. Rev. Respir. Dis 126: 166-170 [Medline].
6. Ninane, V., J.-C. Yernault, and A. De Troyer. 1993. Intrinsic PEEP in patients with chronic obstructive pulmonary disease: role of expiratory muscles. Am. Rev. Respir. Dis 148: 1037-1042 [Medline].
7. Lessard, M. R., F. Lofaso, and L. Brochard. 1995. Expiratory muscle activity increases intrinsic positive end-expiratory pressure independently of dynamic hyperinflation in mechanically ventilated patients. Am. J. Respir. Crit. Care Med 151: 562-569 [Abstract].
8. Yan, S., B. Kayser, M. Tobiasz, and P. Sliwinski. 1996. Comparison of static and dynamic intrinsic positive end-expiratory pressure using the Campbell diagram. Am. J. Respir. Crit. Care Med 154: 938-944 [Abstract].
9.
Braun, N. M. T.,
A. S. Narinder, and
D. F. Rochester.
1982.
Force-length
relationship of the normal human diaphragm.
J. Appl. Physiol
53:
405-412
10.
Gauthier, A. P.,
S. Verbanck,
M. Estenne,
C. Segebarth,
P. T. Macklem, and
M. Paiva.
1994.
Three-dimensional reconstruction of the in vivo
human diaphragm shape at different lung volumes.
J. Appl. Physiol
76:
495-506
11. Sliwinski, P., S. Cala, and P. T. Macklem. 1994. Effect of dynamic hyperinflation on ventilation during CO2 rebreathing (abstract). Eur. Respir. J 7: 286S .
12. Kayser, B., P. Sliwinski, S. Yan, M. Tobiasz, and P. T. Macklem. 1996. Effects of moderate expiratory airflow limitation during exercise on the perception of breathlessness in healthy humans (abstract). Am. J. Respir. Crit. Care Med 153: A116 .
13. Agostoni, E., and J. Mead. 1964. Statics of the respiratory system. In W. O. Fenn and H. Rahn, editors. Handbook of Physiology, Section 3: Respiration, Vol. 1, 1st ed. American Physiological Society, Washington, DC. 387-410.
14.
Macklem, P. T.,
D. Gross,
A. Grassino, and
C. Roussos.
1978.
Partitioning of inspiratory pressure swings between diaphragm and intercostal/ accessory muscles.
J. Appl. Physiol
44:
200-208
15.
Criner, G. J., and
B. R. Celli.
1987.
Ventilatory muscle recruitment in exercise with O2 in obstructed patients with mild hypoxemia.
J. Appl. Physiol
63:
195-200
16. Martinez, F. J., J. I. Couser, and B. R. Celli. 1990. Factors influencing ventilatory muscle recruitment in patients with chronic airflow obstruction. Am. Rev. Respir. Dis 142: 276-282 [Medline].
17. Montes de Oca, M., J. Rassulo, and B. R. Celli. 1996. Respiratory muscle and cardiopulmonary function during exercise in very severe COPD. Am. J. Respir. Crit. Care Med 154: 1284-1289 [Abstract].
18.
Smith, J., and
F. Bellemare.
1987.
Effect of lung volume on in vivo contraction characteristics of human diaphragm.
J. Appl. Physiol
62:
1893-1900
19.
Yan, S.,
T. Similowski,
A. P. Gauthier,
P. T. Macklem, and
F. Bellemare.
1992.
Effect of fatigue on diaphragmatic function at different lung volumes.
J. Appl. Physiol
72:
1064-1067
20.
Agostoni, E..
1962.
Diaphragm activity and thoracoabdominal mechanics
during positive pressure breathing.
J. Appl. Physiol
17:
215-220
21. Green, M., J. Mead, and T. A. Sears. 1978. Muscle activity during chest wall restriction and positive pressure breathing in man. Respir. Physiol 35: 283-300 [Medline].
22.
Flenley, D. C.,
L. D. Pengelly, and
J. Milic-Emili.
1971.
Immediate effects of positive pressure breathing on the ventilatory response to CO2.
J. Appl. Physiol
30:
7-11
23. Daubenspeck, J. A. 1995. Mechanical aspects of loaded breathing. In C. Roussos, editor. The Thorax, Part A: Physiology. Lung Biology in Health and Disease, Vol. 85, 2nd ed. Marcel Dekker, New York. 953-985.
24.
Kim, M. J.,
W. S. Drus,
J. Danon,
W. Machnach, and
J. T. Sharp.
1976.
Mechanics of the canine diaphragm.
J. Appl. Physiol
41:
369-382
25.
Farkas, G. A.,
M. Decramer,
D. F. Rochester, and
A. De Troyer.
1985.
Contractile properties of intercostal muscles and their functional significance.
J. Appl. Physiol
59:
528-535
26.
Decramer, M., and
A. De Troyer.
1984.
Respiratory changes in parasternal intercostal length.
J. Appl. Physiol
57:
1254-1260
27.
Farkas, G. A., and
D. F. Rochester.
1986.
Contractile characteristics and
operating lengths of canine neck inspiratory muscles.
J. Appl. Physiol
61:
220-226
28.
Zocchi, L.,
N. Garzaniti,
S. Newman, and
P. T. Macklem.
1987.
Effect of
hyperinflation and equilibration of abdominal pressure on diaphragmatic action.
J. Appl. Physiol
62:
1655-1664
29. Jiang, T. X., K. Deschepper, M. Demedts, and M. Decramer. 1989. Effect of acute hyperinflation on the mechanical effectiveness of the parasternal intercostals. Am. Rev. Respir. Dis 139: 522-528 [Medline].
30.
Levine, S.,
M. Gillen,
P. Weiser,
G. Feiss,
M. Goldman, and
D. Henson.
1988.
Inspiratory pressure generation: comparison of subjects with
COPD and age-matched normals.
J. Appl. Physiol
65:
888-899
31. Rossi, A., G. Polese, and G. Brandi. 1991. Dynamic hyperinflation. In J. J. Marini and C. Roussos, editors. Ventilatory Failure, Vol. 15. Springer-Verlag, Berlin. 199-218.
This article has been cited by other articles:
![]() |
J. A. Simpson, K. R. Brunt, C. P. Collier, and S. Iscoe Hyperinflation-induced cardiorespiratory failure in rats J Appl Physiol, July 1, 2009; 107(1): 275 - 282. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. D. Eves, S. R. Petersen, M. J. Haykowsky, E. Y. Wong, and R. L. Jones Helium-Hyperoxia, Exercise, and Respiratory Mechanics in Chronic Obstructive Pulmonary Disease Am. J. Respir. Crit. Care Med., October 1, 2006; 174(7): 763 - 771. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. YAN Sensation of Inspiratory Difficulty during Inspiratory Threshold and Hyperinflationary Loadings . Effect of Inspiratory Muscle Strength Am. J. Respir. Crit. Care Med., November 1, 1999; 160(5): 1544 - 1549. [Abstract] [Full Text] |
||||
![]() |
M. I. Polkey, C.-H. Hamnegard, P. D. Hughes, G. F. Rafferty, M. Green, and J. Moxham Influence of acute lung volume change on contractile properties of human diaphragm J Appl Physiol, October 1, 1998; 85(4): 1322 - 1328. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Similowski, C. Straus, V. Attali, A. Duguet, and J.-P. Derenne Cervical magnetic stimulation as a method to discriminate between diaphragm and rib cage muscle fatigue J Appl Physiol, May 1, 1998; 84(5): 1692 - 1700. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Proc. Am. Thorac. Soc. | Am. J. Respir. Cell Mol. Biol. |