American Journal of Respiratory and Critical Care Medicine Vol 166. pp. 469-473, (2002)
© 2002 American Thoracic Society
Response of Automatic Continuous Positive Airway Pressure Devices to Different Sleep Breathing Patterns
A Bench Study
Ramon Farré,
Josep M. Montserrat,
Jordi Rigau,
Xavier Trepat,
Paula Pinto and
Daniel Navajas
Unitat de Biofísica i Bioenginyeria, Facultat de Medicina, Universitat de Barcelona; Servei de Pneumologia i Al lèrgia Respiratòria, Hospital Clinic Provincial; and Institut d'Investigacions Biomèdiques August Pi Sunyer, Barcelona, Spain
Correspondence and requests for reprints should be addressed to Dr. Ramon Farré, Unitat de Biofisica i Bioenginyeria, Facultat de Medicina Casanova, 143 E-08036, Barcelona, Spain. E-mail: farre{at}medicina.ub.es
Evaluating the usefulness of automatic continuous positive airway pressure (CPAP) in treating the sleep apneahypopnea syndrome (SAHS) is not easy because the algorithms for automatic CPAP implemented in the devices available are not well known and are probably dependent on the device. In addition, at present it is not possible to test the behavior of automatic CPAP devices in response to well-defined breathing patterns. Our aim was to implement a bench test to characterize the responses of automatic CPAP devices by subjecting them to breathing patterns of patients with SAHS. To this end, a variety of typical breathing patterns (normal, apneas, hypopneas, flow limitation, snoring) previously recorded in patients with SAHS during sleep were reproduced by a breathing waveform generator. Five commercially available automatic CPAP devices were tested. The responses of the devices to apneas, hypopneas, flow limitation, and snoring were considerably different. In some devices, the response was modified by air leaks similar to the ones found in patients. Consequently, the effectiveness of automatic CPAP assessed in clinical tests performed by using particular devices has no general validity. Testing automatic CPAP devices in a bench study is a useful first step in evaluating the performance of this new type of device in adjusting nasal pressure for each patient.
Key Words: automatic CPAP devices sleep apnea hypopnea flow limitation self-regulation
This article has been cited by other articles:

|
 |

|
 |
 
R. Farre and D. Navajas
Quality control: a necessary, but sometimes overlooked, tool for improving respiratory medicine
Eur. Respir. J.,
April 1, 2009;
33(4):
722 - 723.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
D. Pevernagie, J. F. Masa, J. C. Meurice, R. Farre, O. Marrone, and D. Rodenstein
Treatment of obstructive sleep-disordered breathing with positive airway pressure systems
Eur. Respir. Rev.,
December 1, 2007;
16(106):
125 - 131.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
R. K. Kakkar and R. B. Berry
Positive Airway Pressure Treatment for Obstructive Sleep Apnea
Chest,
September 1, 2007;
132(3):
1057 - 1072.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
F. Lofaso, G. Desmarais, K. Leroux, V. Zalc, R. Fodil, D. Isabey, and B. Louis
Bench evaluation of flow limitation detection by automated continuous positive airway pressure devices.
Chest,
August 1, 2006;
130(2):
343 - 349.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. Rigau, J. M. Montserrat, H. Wohrle, D. Plattner, M. Schwaibold, D. Navajas, and R. Farre
Bench model to simulate upper airway obstruction for analyzing automatic continuous positive airway pressure devices.
Chest,
August 1, 2006;
130(2):
350 - 361.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
G. M. Nolan, S. Ryan, T. M. O'Connor, and W. T. McNicholas
Comparison of three auto-adjusting positive pressure devices in patients with sleep apnoea
Eur. Respir. J.,
July 1, 2006;
28(1):
159 - 164.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. F. Masa, A. Jimenez, J. Duran, F. Capote, C. Monasterio, M. Mayos, J. Teran, L. Hernandez, F. Barbe, A. Maimo, et al.
Alternative Methods of Titrating Continuous Positive Airway Pressure: A Large Multicenter Study
Am. J. Respir. Crit. Care Med.,
December 1, 2004;
170(11):
1218 - 1224.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
F. Abdenbi, B. Chambille, and P. Escourrou
Bench testing of auto-adjusting positive airway pressure devices
Eur. Respir. J.,
October 1, 2004;
24(4):
649 - 658.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
D. A. Pevernagie, P. M. Proot, K. B. Hertegonne, M. C. Neyens, K. P. Hoornaert, and R. A. Pauwels
Efficacy of Flow- vs Impedance-Guided Autoadjustable Continuous Positive Airway Pressure: A Randomized Cross-over Trial
Chest,
July 1, 2004;
126(1):
25 - 30.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
O. Senn, T. Brack, F. Matthews, E. W. Russi, and K. E. Bloch
Randomized Short-term Trial of Two AutoCPAP Devices versus Fixed Continuous Positive Airway Pressure for the Treatment of Sleep Apnea
Am. J. Respir. Crit. Care Med.,
December 15, 2003;
168(12):
1506 - 1511.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. J. Tobin
Sleep-Disordered Breathing, Control of Breathing, Respiratory Muscles, and Pulmonary Function Testing in AJRCCM 2002
Am. J. Respir. Crit. Care Med.,
February 1, 2003;
167(3):
306 - 318.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
C. A. Massie, N. McArdle, R. W. Hart, W. W. Schmidt-Nowara, A. Lankford, D. W. Hudgel, N. Gordon, and N. J. Douglas
Comparison between Automatic and Fixed Positive Airway Pressure Therapy in the Home
Am. J. Respir. Crit. Care Med.,
January 1, 2003;
167(1):
20 - 23.
[Abstract]
[Full Text]
[PDF]
|
 |
|
Copyright © 2002 American Thoracic Society
|
|
|