Am. J. Respir. Crit. Care Med.,
Volume 162, Number 6, December 2000, 2109-2112
Tidal Volumes For Ventilated Infants Should Be
Determined with a Pneumotachometer Placed
at the Endotracheal Tube
MICHAEL L.
CANNON,
JACK
CORNELL,
DONNA S.
TRIPP-HAMEL,
MICHAEL A.
GENTILE,
CHRISTOPHER L.
HUBBLE,
JON N.
MELIONES,
and
IRA M.
CHEIFETZ
Divisions of Pediatric Critical Care Medicine and Pediatric Respiratory Care, Duke Children's Hospital, Duke University Medical Center,
Durham, North Carolina
Many ventilators measure expired tidal volume (VT) without compensation either for the compliance of the ventilator circuit or for
variations in the circuit setup. We hypothesized that the exhaled
VT measured with a conventional ventilator at the expiratory valve
would differ significantly from the exhaled VT measured with a
pneumotachometer placed at the endotracheal tube. To investigate this we studied 98 infants and children requiring conventional ventilation. We used linear regression analysis to compare
the VT obtained with the pneumotachometer with the ventilator-measured volume. An additional comparison was made between
the pneumotachometer volume and a calculated effective VT. For
infant circuits (n = 70), our analysis revealed a poor correlation
between the expiratory VT measured with the pneumotachometer
and the ventilator-measured volume (r2 = 0.54). Similarly, the expiratory VT measured with the pneumotachometer did not correlate with the calculated effective volume (r2 = 0.58). For pediatric
circuits (n = 28), there was improved correlation between the expiratory VT measured with the pneumotachometer and both the
ventilator-measured volume and the calculated effective VT (r2 = 0.84 and r2 = 0.85, respectively). The data demonstrate a significant discrepancy between expiratory VT measured at a ventilator
and that measured with a pneumotachometer placed at the endotracheal tube in infants. Correcting for the compliance of the
ventilator circuit by calculating the effective VT did not alter this
discrepancy. In conventionally ventilated infants, exhaled VT
should be determined with a pneumotachometer placed at the airway.
This article has been cited by other articles:

|
 |

|
 |
 
P. R. Bachiller, J. M. McDonough, and J. M. Feldman
Do New Anesthesia Ventilators Deliver Small Tidal Volumes Accurately During Volume-Controlled Ventilation?
Anesth. Analg.,
May 1, 2008;
106(5):
1392 - 1400.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. Keszler
Volume-targeted Ventilation
NeoReviews,
May 1, 2006;
7(5):
e250 - e257.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
R. Farre, S. J. Lloyd-Owen, N. Ambrosino, G. Donaldson, J. Escarrabill, B. Fauroux, D. Robert, B. Schoenhofer, A. Simonds, and J. A. Wedzicha
Quality control of equipment in home mechanical ventilation: a European survey
Eur. Respir. J.,
July 1, 2005;
26(1):
86 - 94.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
A. G. Randolph, K. L. Meert, M. E. O'Neil, J. H. Hanson, P. M. Luckett, J. H. Arnold, R. G. Gedeit, P. N. Cox, J. S. Roberts, S. T. Venkataraman, et al.
The Feasibility of Conducting Clinical Trials in Infants and Children with Acute Respiratory Failure
Am. J. Respir. Crit. Care Med.,
May 15, 2003;
167(10):
1334 - 1340.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
S. E. Courtney, D. J. Durand, J. M. Asselin, M. L. Hudak, J. L. Aschner, C. T. Shoemaker, and the Neonatal Ventilation Study Group
High-Frequency Oscillatory Ventilation versus Conventional Mechanical Ventilation for Very-Low-Birth-Weight Infants
N. Engl. J. Med.,
August 29, 2002;
347(9):
643 - 652.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. J. TOBIN
Pediatrics, Surfactant, and Cystic Fibrosis in AJRCCM 2000
Am. J. Respir. Crit. Care Med.,
November 1, 2001;
164(9):
1581 - 1594.
[Full Text]
[PDF]
|
 |
|
Copyright © 2000 American Thoracic Society
|
|
|