Am. J. Respir. Crit. Care Med.,
Volume 160, Number 2, August 1999, 415-420
Analysis of Observer Variability in Measurement of
Pulmonary Artery Occlusion Pressures
TAMIM
AL-KHARRAT,
STUART
ZARICH,
YAW
AMOATENG-ADJEPONG,
and
CONSTANTINE A.
MANTHOUS
Divisions of Pulmonary and Critical Care and Cardiology, Bridgeport Hospital, Bridgeport; and Yale University School
of Medicine, New Haven, Connecticut
The goal of this study was to determine inter- and intraobserver variability in measurement of pulmonary artery occlusion pressure (Ppao), comparing values recorded by critical care nurses and those
measured by physician specialists. Critical care nurses (CCNs) obtained contiguous pulmonary artery
and occlusion pressure paper tracings, up to twice a day, between June 1997 and March 1998. All
tracings were interpreted on two separate occasions, in blinded fashion, by our Chiefs of Critical Care
(CCMD) and Cardiology (CARD). Their values of Ppao were compared with those that had been recorded by CCNs. One hundred and forty-seven measurements of Ppao were performed on 40 patients with a mean age of 62.5 ± 2.2 yr and a mean APACHE II score of 21.5 ± 0.8. Either or both physician readers found 34 tracings as not satisfactory for Ppao interpretation. Intraobserver agreement
of Ppao measurements, determined by correlation coefficients, was 0.91 for the CCMD and 0.87 for
the CARD. Correlation coefficients for interobserver comparisons were 0.83 for CCMD-CARD, 0.66 for CARD-CCN, and 0.67 for CCMD-CCN. Clinically significant differences were observed between
CCMD-CARD (range of differences,
11 to 12 mm Hg), CARD-CCN (
13 to 15 mm Hg), and CCMD-
CCN (
11 to 15 mm Hg). When Ppao readings were categorized as low (< 5 mm Hg), normal (5-15
mm Hg), and high (> 15 mm Hg),
values were 0.57 for CARD-CCMD, 0.51 for CARD-CCN, and 0.41 for CCMD-CCN comparisons. Interobserver variability was not explained by positive pressure ventilation or by the presence of (> 4 mm Hg) ventricular waves. The absolute values of interobserver differences in tracings with respiratory phasic variations (RPV)
8 mm Hg were significantly greater
than for tracings with variations < 8 mm Hg (p < 0.05, except CCMD-CCN, p = 0.10). Intraobserver
differences also tended to be higher for tracings with RPV
8 mm Hg (p = 0.06 and 0.05). When selected tracings were presented to 23 CCNs and 18 physicians, variability of Ppao interpretation was
twice as great for tracings with large RPVs as compared with those with minimal RPVs. These data
suggest that observer variability of Ppao interpretation is of potential clinical importance and that
the degree of variability is associated with the magnitude of respiratory phasic variation of intrathoracic pressures. Although this could represent a local aberration, this study highlights a factor (respiratory phasic variation of Ppao) responsible for significantly increased intra- and interobserver variabilities.