Am. J. Respir. Crit. Care Med., Vol 153, No. 3, Mar 1996, 997-1004.
Evaluation of a knowledge-based system providing ventilatory management and decision for extubation
M Dojat, A Harf, D Touchard, M Laforest, F Lemaire and L Brochard
Institut National de la Sante et de la Recherche Medicale, Departement de Physiologie, Creteil, France.
We evaluated whether a knowledge-based system (KBS) connected to a
ventilator in pressure support mode could correctly predict the ability of
patients to tolerate total withdrawal from ventilatory support. The KBS was
designed to continuously adapt ventilatory assistance to the needs of the
patient, to manage a strategy of gradually decreasing ventilatory
assistance, and to indicate when the patient was able to breathe without
assistance. Thirty-eight patients for whom weaning was being considered
were evaluated using a conventional battery of parameters, including
weaning criteria, tolerance of a T-piece trial, and outcome 48h after
permanent withdrawal of ventilation. The results of this evaluation were
compared with the suggestions made by the KBS at the end of a period of
KBS-driven mechanical ventilation inserted in the conventional weaning
procedure. The positive predictive value of the KBS was 89%, versus 77% for
the conventional procedure and 81% for the rapid shallow breathing index
alone. The KBS correctly predicted the course of five patients who
tolerated a T-piece trial but required ventilation within 48 h. We conclude
that our KBS ensured appropriate patient management during the weaning
period and improved our ability to predict responses to weaning.
This article has been cited by other articles:

|
 |

|
 |
 
J-M. Boles, J. Bion, A. Connors, M. Herridge, B. Marsh, C. Melot, R. Pearl, H. Silverman, M. Stanchina, A. Vieillard-Baron, et al.
Weaning from mechanical ventilation
Eur. Respir. J.,
May 1, 2007;
29(5):
1033 - 1056.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
F. Lellouche, J. Mancebo, P. Jolliet, J. Roeseler, F. Schortgen, M. Dojat, B. Cabello, L. Bouadma, P. Rodriguez, S. Maggiore, et al.
A Multicenter Randomized Trial of Computer-driven Protocolized Weaning from Mechanical Ventilation
Am. J. Respir. Crit. Care Med.,
October 15, 2006;
174(8):
894 - 900.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. H. T. Bates and M. P. Young
Applying Fuzzy Logic to Medical Decision Making in the Intensive Care Unit
Am. J. Respir. Crit. Care Med.,
April 1, 2003;
167(7):
948 - 952.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
R. Twibell, D. Siela, and M. Mahmoodi
Subjective Perceptions and Physiological Variables During Weaning From Mechanical Ventilation
Am. J. Crit. Care.,
March 1, 2003;
12(2):
101 - 112.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
ATS/ERS Statement on Respiratory Muscle Testing
Am. J. Respir. Crit. Care Med.,
August 15, 2002;
166(4):
518 - 624.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
E. Henneman, K. Dracup, T. Ganz, O. Molayeme, and C. B. Cooper
Using a Collaborative Weaning Plan to Decrease Duration of Mechanical Ventilation and Length of Stay in the Intensive Care Unit for Patients Receiving Long-Term Ventilation
Am. J. Crit. Care.,
March 1, 2002;
11(2):
132 - 140.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. F. El-Khatib, G. W. Jamaleddine, A. R. Khoury, and M. Y. Obeid
Effect of Continuous Positive Airway Pressure on the Rapid Shallow Breathing Index in Patients Following Cardiac Surgery
Chest,
February 1, 2002;
121(2):
475 - 479.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
N. R. MacIntyre
Evidence-Based Guidelines for Weaning and Discontinuing Ventilatory Support : A Collective Task Force Facilitated by the American College of Chest Physicians; the American Association for Respiratory Care; and the American College of Critical Care Medicine
Chest,
December 1, 2001;
120(6_suppl):
375S - 396S.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. Meade, G. Guyatt, D. Cook, L. Griffith, T. Sinuff, C. Kergl, J. Mancebo, A. Esteban, and S. Epstein
Predicting Success in Weaning From Mechanical Ventilation
Chest,
December 1, 2001;
120(6_suppl):
400S - 424S.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. DOJAT, A. HARF, D. TOUCHARD, F. LEMAIRE, and L. BROCHARD
Clinical Evaluation of a Computer-controlled Pressure Support Mode
Am. J. Respir. Crit. Care Med.,
April 1, 2000;
161(4):
1161 - 1166.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
A. D. Brook and M. H. Kollef
An Outcomes-Based Approach to Ventilatory Management: Review of Two Examples: Brook AD, Kollef MH An outcomes-based approach to ventilatory management review of two examples J Intensive Care Med 1999,14 262-274
J Intensive Care Med,
November 1, 1999;
14(6):
262 - 274.
[PDF]
|
 |
|

|
 |

|
 |
 
T. NEMOTO, G. E. HATZAKIS, C. W. THORPE, R. OLIVENSTEIN, S. DIAL, and J. H. T. BATES
Automatic Control of Pressure Support Mechanical Ventilation Using Fuzzy Logic
Am. J. Respir. Crit. Care Med.,
August 1, 1999;
160(2):
550 - 556.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
E. W. Ely, A. M. Baker, D. P. Dunagan, H. L. Burke, A. C. Smith, P. T. Kelly, M. M. Johnson, R. W. Browder, D. L. Bowton, and E. F. Haponik
Effect on the Duration of Mechanical Ventilation of Identifying Patients Capable of Breathing Spontaneously
N. Engl. J. Med.,
December 19, 1996;
335(25):
1864 - 1869.
[Abstract]
[Full Text]
[PDF]
|
 |
|
Copyright © 1996 American Thoracic Society
|
|
|